A novel valveless rotary pump for cardiac assist

被引:0
|
作者
Rosenfeldt, FL [1 ]
Ostberg, BN [1 ]
Tcherkas, DD [1 ]
PastorizaPinol, JV [1 ]
Southwell, JE [1 ]
机构
[1] ALFRED HOSP,PRAHRAN,VIC 3181,AUSTRALIA
关键词
rotary pump; pulsatile; nonpulsatile;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe the design and preliminary testing of a novel pump designed for cardiac assist. It is a valveless rotary pump that can generate pulsatile or nonpulsatile output. The basis of the pump design is a cylindrical cavity swept by a vane sliding in a rotating disc. During rotation, rollers at the ends of the vane are in constant contact with the internal wall of the cavity so that a fluid-containing space is swept around the cavity from the inflow to the outflow ports. Bench tests showed that the output was relatively independent of the outflow resistance at pressures up to 100 mm Hg. In tests in dogs, the pimp performed satisfactorily as a left ventricular assist device and as a pump for total cardiopulmonary bypass. We believe that this simple pump design has potential as a left ventricular assist device or total artificial heart and merits further investigation.
引用
收藏
页码:420 / 425
页数:6
相关论文
共 50 条
  • [1] INVITRO EVALUATION OF A VALVELESS CARDIAC-ASSIST PUMP
    EINAV, S
    ELAD, D
    AVIDOR, JM
    STOLERO, D
    HAGIGAT, A
    BARAK, J
    VIDNE, BA
    ARTIFICIAL ORGANS, 1989, 13 (05) : 480 - 483
  • [2] AN IMPLANTABLE VALVELESS HEART ASSIST PUMP
    TURNER, MC
    PIERCE, WS
    METZ, HD
    GOODMAN, L
    JOURNAL OF BASIC ENGINEERING, 1969, 91 (02): : 284 - &
  • [3] AN IMPLANTABLE VALVELESS HEART ASSIST PUMP
    TURNER, MC
    PIERCE, WS
    METZ, HD
    GOODMAN, L
    MECHANICAL ENGINEERING, 1969, 91 (04) : 59 - &
  • [4] DISCUSSION OF AN IMPLANTABLE VALVELESS HEART ASSIST PUMP
    ROGERS, A
    TURNER, MC
    JOURNAL OF BASIC ENGINEERING, 1969, 91 (02): : 288 - &
  • [5] A novel, valveless ventricular assist device:: The FishTail pump.: First experimental in vivo studies
    Feier, H
    Mekkaoui, C
    Drevet, JB
    Sérée, Y
    Richomme, C
    Rolland, PH
    Mesana, TG
    ARTIFICIAL ORGANS, 2002, 26 (12) : 1026 - 1031
  • [6] Numerical simulation of carotid hemodynamics in patients with rotary blood pump cardiac assist
    Schima, H
    Lackner, B
    Prosi, M
    Perktold, K
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (02): : 152 - 160
  • [7] Numerical simulation of carotid hemodynamics in patients with rotary blood pump cardiac assist
    Schima, H. (heinrich.schima@akh-wien.ac.at), 1600, Wichtig Editore s.r.l. (26):
  • [8] Cora valveless pulsatile rotary pump: New design and control
    Monties, JRE
    Trinkl, J
    Mesana, T
    Havlik, PJ
    Demunck, JLY
    ANNALS OF THORACIC SURGERY, 1996, 61 (01): : 463 - 468
  • [9] Flow Field of a Novel Implantable Valveless Counterpulsation Heart Assist Device
    A. Berthe
    S. Gärtlein
    Ch. Lederer
    U. Kertzscher
    K. Affeld
    L. Goubergrits
    Annals of Biomedical Engineering, 2012, 40 : 1982 - 1995
  • [10] Development of a novel centrifugal pump: Magnetic rotary pump
    Naganuma, S
    Yambe, T
    Sonobe, T
    Kobayashi, S
    Nitta, S
    ARTIFICIAL ORGANS, 1997, 21 (07) : 746 - 750