Clinical Evaluation of the New Amplatzer Duct Occluder II for Patent Arterial Duct Occlusion

被引:22
|
作者
Bhole, V. [1 ]
Miller, P. [1 ]
Mehta, C. [2 ]
Stumper, O. [1 ]
Reinhardt, Z. [1 ]
De Giovanni, J. V. [1 ]
机构
[1] Birmingham Childrens Hosp, Heart Unit, Birmingham B4 6NH, W Midlands, England
[2] Alder Hey Childrens Hosp, Dept Cardiol, Liverpool L12 2AP, Merseyside, England
关键词
device closure; Amplatzer ADO II; cardiac catheterization; patent arterial duct; TRANSCATHETER CLOSURE; EXPERIENCE; DEVICE; COILS; IMMEDIATE;
D O I
10.1002/ccd.22095
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Several devices such as coils and Amplatzer duct occluder (ADO) are used for catheter closure of patent arterial ducts (PDA). These carry a high success rate but residual shunts, suboptimal device orientation, and technical problems are encountered. The Amplatzer duct occluder II (ADO II) is designed to address these limitations. Objectives: To evaluate the technical features of the new ADO II device for PDA closure and document the immediate/early closure rate, complications and device behavior during implantation. Methods: Prospective, two center study from February 2008 to January 2009. Twenty-seven patients (18 females) received the ADO II The median age was 22 months (range: 7 months-68 years) and the median weight was 11.7 kg (range: 6.9-108). The median PDA diameter was 2.6 mm (range: 1-4.4). The approach was arterial in 13 and venous in 14 patients. Follow-up included echocardiography at 1 day and 1 month postimplantation. Results: All implantations were technically successful with immediate complete angiographic closure in 21 and trivial contrast flow in six patients. The median procedure time was 43 min (range: 15-82) and the fluoroscopy time was 6 min (range: 2.2-26.5). Echocardiography confirmed no residual shunts on the following day. There were no complications. Conclusion: The new ADO II is a versatile and very effective device for closure of PDAs of various shapes, lengths, and up to diameters of 5.5 mm. The disc articulations, high early closure rate, arterial or venous approach options, and small diameter delivery catheter are all beneficial features. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:762 / 769
页数:8
相关论文
共 50 条
  • [1] Multi-center trial for the occlusion of patent arterial duct by the Amplatzer Duct Occluder in China
    Leung, MP
    Chin, M
    Chau, O
    Han, L
    Li, Y
    Chau, AK
    CIRCULATION, 2000, 102 (18) : 587 - 587
  • [2] Transcatheter Closure of Patent Arterial Duct With Amplatzer Duct Occluder II Additional Sizes in Children
    Eleftherakis, Nikolaos G.
    Vekiou, Athina
    JOURNAL OF INVASIVE CARDIOLOGY, 2013, 25 (02): : 96 - 97
  • [3] Severe haemolysis after transcatheter closure of a patent arterial duct with the new Amplatzer™ duct occluder
    Godart, F
    Rodés, J
    Rey, C
    CARDIOLOGY IN THE YOUNG, 2000, 10 (03) : 265 - 267
  • [4] Closure of the patent ductus arteriosus with the Amplatzer Duct Occluder II: a clinical experience
    Karagoz, Tevfik
    Akin, Alper
    Ertugrul, Ilker
    Aykan, Hayrettin Hakan
    Alehan, Dursun
    Ozer, Sema
    Ozkutlu, Suheyla
    ACTA CARDIOLOGICA, 2012, 67 (06) : 675 - 680
  • [5] Early Clinical Experience With the New Amplatzer Ductal Occluder II for Closure of the Persistent Arterial Duct
    Forsey, Jonathan
    Kenny, Damien
    Morgan, Gareth
    Hayes, Alison
    Turner, Mark
    Tometzki, Andrew
    Martin, Robin
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 74 (04) : 615 - 623
  • [6] Transcatheter occlusion of patent ductus arteriosus using a new angled Amplatzer duct occluder: Initial clinical experience
    Masura, J
    Gavora, P
    Podnar, T
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (02) : 261 - 267
  • [7] Transcatheter Closure of Long Tubular Patent Arterial Ducts: The Amplatzer Duct Occluder II - A New and Valuable Tool
    Morgan, Gareth
    Tornetzki, Andrew J.
    Martin, Robin P.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 73 (04) : 576 - 580
  • [8] Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes
    Giuseppe Santoro
    Mario Giordano
    Gianpiero Gaio
    Maria Teresa Palladino
    Giovanbattista Capozzi
    Carola Iacono
    Maria Giovanna Russo
    Pediatric Cardiology, 2018, 39 : 627 - 632
  • [9] Initial Experience with the New Amplatzer® Duct Occluder II
    Dua, Jaspal
    Chessa, Massimo
    Piazza, Luciane
    Negura, Diana
    Micheletti, Angelo
    Bussadori, Claudio
    Butera, Gianfranco
    Carminati, Mario
    JOURNAL OF INVASIVE CARDIOLOGY, 2009, 21 (08): : 401 - 405
  • [10] Transcatheter closure of a patent arterial duct with the Amplatzer duct occluder in the area of a large thoracic aortic aneurysm
    Ghazi, Payam
    Haji-Zeinali, Ali-Mohammad
    CARDIOLOGY IN THE YOUNG, 2009, 19 (02) : 209 - 211