Reproducibility and accuracy of measurements with a hand held autorefractor in children

被引:64
|
作者
Harvey, EM
Miller, JM
Wagner, LK
Dobson, V
机构
[1] UNIV ARIZONA, CTR OPT SCI, TUCSON, AZ 85719 USA
[2] UNIV ARIZONA, DEPT PSYCHOL, TUCSON, AZ 85719 USA
关键词
D O I
10.1136/bjo.81.11.941
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim-To determine reproducibility and accuracy of the Nikon Retinomax autorefractor when used with children who were made cycloplegic. Methods-Autorefraction and retinoscopy or subjectively refined retinoscopy (where, under the patient's direction, the refraction was varied until the best visual acuity was achieved) were performed on the right eye of 47 children, age 11-93 months. Autorefraction was performed using the Nikon Retinomax, which provides up to eight measured values of refractive error and one representative measurement of refractive error. Results-Autorefractor measurements were successfully obtained from 7/9 children age 3 years or younger, and from all older children. Vector methods were used to calculate differences. Retinomax reproducibility averaged 0.43 D. Unbiased Retinomax and retinoscopy measurements differed by an average of 0.82 D. Unbiased Retinomax and subjectively refined retinoscopy differed by an average of 1.03 D. Conclusions-Reproducibility of Retinomax measured values in children is comparable with reproducibility of retinoscopy, subjective refraction, and autorefraction measurements in adults. Agreement between Retinomax and retinoscopy and agreement between Retino-MATERIALS max and subjective refinement in children is comparable with agreement between autorefraction and subjective refraction in adults. The study indicates that the Retinomax is a useful instrument for measuring refractive errors in young children.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 50 条
  • [1] Accuracy of a hand-held autorefractor in a pediatric population
    Wagner, LK
    Miller, JM
    Wilson, MC
    Harvey, EM
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 3334 - 3334
  • [2] Repeatability and accuracy of autorefractor measurements in young children
    Huffman, SJ
    Schimmoeller, KL
    Schmidt, P
    Mitchell, G
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U994 - U994
  • [3] Accommodation in a hand-held autorefractor
    Chen, Ying-Ling
    Shi, Lei
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [4] Evaluation of the hand-held autorefractor Nikon Retinomax on adults and children
    Wesemann, W
    Dick, B
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1997, 211 (06) : 387 - 394
  • [5] Evaluation of a hand-held autorefractor in children younger than 6
    El-Defrawy, S
    Clarke, WN
    Belec, F
    Pham, B
    JOURNAL OF PEDIATRIC OPHTHALMOLOGY & STRABISMUS, 1998, 35 (02) : 107 - 109
  • [6] Reproducibility of the hand held Nidek autokeratometer in children
    Chandna, A
    Noonan, CP
    Rao, GP
    Pathmanathan, T
    Kaye, S
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 1418 - 1418
  • [7] Clinical Accuracy and Precision of a Hand-Held Shack Hartman Wavefront Sensor Autorefractor
    Brown, Nicolas
    Vaz, Tara C.
    Pikul, Olga
    Dowd, Robert
    Nally, Rebecca L.
    Klein, Sarah B.
    Yoon, Geunyoung
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [8] Examination of children for refractive errors in the kindergarten using a hand-held autorefractor
    Buchner, TF
    Schnorbus, U
    Grenzebach, UH
    Busse, H
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U336 - U336
  • [9] REPRODUCIBILITY OF CORNEAL ASTIGMATISM MEASUREMENTS WITH A HAND-HELD KERATOMETER IN PRESCHOOL-CHILDREN
    HARVEY, EM
    MILLER, JM
    DOBSON, V
    BRITISH JOURNAL OF OPHTHALMOLOGY, 1995, 79 (11) : 983 - 990
  • [10] Cycloplegic refraction in preschool children: comparisons between the hand-held autorefractor, table-mounted autorefractor and retinoscopy
    Prabakaran, S.
    Dirani, M.
    Chia, A.
    Gazzard, G.
    Fan, Q.
    Leo, S. -W.
    Ling, Y.
    Eong, K. -G. Au
    Wong, T. -Y.
    Saw, S. -M.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2009, 29 (04) : 422 - 426