Use of population input functions for reduced scan duration whole-body Patlak 18F-FDG PET imaging

被引:18
|
作者
van Sluis, Joyce [1 ]
Yaqub, Maqsood [2 ]
Brouwers, Adrienne H. [1 ]
Dierckx, Rudi A. J. O. [1 ]
Noordzij, Walter [1 ]
Boellaard, Ronald [1 ,2 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, Hanzeplein 1, NL-9713 GZ Groningen, Netherlands
[2] Vrije Univ Amsterdam, Canc Ctr Amsterdam, Amsterdam UMC, Dept Radiol & Nucl Med, Amsterdam, Netherlands
关键词
Patlak; PET; CT; Dynamic imaging; Population input function; Scan time; CELL LUNG-CANCER; FDG-PET; THERAPY; RECOMMENDATIONS; QUANTIFICATION; REPEATABILITY; ANATOMY; TUMORS;
D O I
10.1186/s40658-021-00357-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Whole-body Patlak images can be obtained from an acquisition of first 6 min of dynamic imaging over the heart to obtain the arterial input function (IF), followed by multiple whole-body sweeps up to 60 min pi. The use of a population-averaged IF (PIF) could exclude the first dynamic scan and minimize whole-body sweeps to 30-60 min pi. Here, the effects of (incorrect) PIFs on the accuracy of the proposed Patlak method were assessed. In addition, the extent of mitigating these biases through rescaling of the PIF to image-derived IF values at 30-60 min pi was evaluated. Methods Using a representative IF and rate constants from the literature, various tumour time-activity curves (TACs) were simulated. Variations included multiplication of the IF with a positive and negative gradual linear bias over 60 min of 5, 10, 15, 20, and 25% (generating TACs using an IF different from the PIF); use of rate constants (K-1, k(3), and both K-1 and k(2)) multiplied by 2, 1.5, and 0.75; and addition of noise (mu = 0 and sigma = 5, 10 and 15%). Subsequent Patlak analysis using the original IF (representing the PIF) was used to obtain the influx constant (K-i) for the differently simulated TACs. Next, the PIF was scaled towards the (simulated) IF value using the 30-60-min pi time interval, simulating scaling of the PIF to image-derived values. Influence of variabilities in IF and rate constants, and rescaling the PIF on bias in K-i was evaluated. Results Percentage bias in K-i observed using simulated modified IFs varied from - 16 to 16% depending on the simulated amplitude and direction of the IF modifications. Subsequent scaling of the PIF reduced these K-i biases in most cases (287 out of 290) to < 5%. Conclusions Simulations suggest that scaling of a (possibly incorrect) PIF to IF values seen in whole-body dynamic imaging from 30 to 60 min pi can provide accurate Ki estimates. Consequently, dynamic Patlak imaging protocols may be performed for 30-60 min pi making whole-body Patlak imaging clinically feasible.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Combined early dynamic 18F-FDG PET/CT and conventional whole-body 18F-FDG PET/CT in hepatocellular carcinoma
    Yiqiu Zhang
    Yun Dong
    Wenjun Yu
    Shuguang Chen
    Haojun Yu
    Beilei Li
    Hongcheng Shi
    Abdominal Radiology, 2023, 48 : 3127 - 3134
  • [32] Improvement of motion artifacts using dynamic whole-body 18F-FDG PET/CT imaging
    Tomohito Kaji
    Kouji Osanai
    Atsushi Takahashi
    Atsushi Kinoshita
    Daiki Satoh
    Tomoaki Nakata
    Nagara Tamaki
    Japanese Journal of Radiology, 2024, 42 : 374 - 381
  • [33] Improvement of motion artifacts using dynamic whole-body 18F-FDG PET/CT imaging
    Kaji, Tomohito
    Osanai, Kouji
    Takahashi, Atsushi
    Kinoshita, Atsushi
    Satoh, Daiki
    Nakata, Tomoaki
    Tamaki, Nagara
    JAPANESE JOURNAL OF RADIOLOGY, 2023, 42 (4) : 374 - 381
  • [34] Iatrogenic artifacts on whole-body F-18 FDG PET imaging
    Bhargava, P
    Zhuang, H
    Kumar, R
    Charron, M
    Alavi, A
    CLINICAL NUCLEAR MEDICINE, 2004, 29 (07) : 429 - 439
  • [35] Prognostic value of whole-body dynamic 18F-FDG PET/CT Patlak in diffuse large B-cell lymphoma
    Yin, Jiankang
    Wang, Hui
    Zhu, Gan
    Chen, Ni
    Khan, Muhammad Imran
    Zhao, Ye
    HELIYON, 2023, 9 (09)
  • [36] Benign Colonic F-18-FDG Uptake on Whole-Body FDG-PET Scan
    Jang, Byung Ik
    CLINICAL ENDOSCOPY, 2012, 45 (02) : 109 - 110
  • [37] Evaluation of 18F-FDG PET/MRI, 18F-FDG PET/CT, MRI, and CT in whole-body staging of recurrent breast cancer
    Sawicki, Lino M.
    Grueneisen, Johannes
    Schaarschmidt, Benedikt M.
    Buchbender, Christian
    Nagarajah, James
    Umutlu, Lale
    Antoch, Gerald
    Kinner, Sonja
    EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (02) : 459 - 465
  • [38] Improved lesion detection in Whole-Body Parametric 18F-FDG PET/CT vs. Static 18F-FDG PET/CT
    Dias, Andre
    Pedersen, Mette
    Danielsen, Helle
    Munk, Ole
    Gormsen, Lars
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [39] Whole-body 18F-FDG PET identifies high-risk myeloma
    Durie, BGM
    Waxman, AD
    D'Agnolo, A
    Williams, CM
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (11) : 1457 - 1463
  • [40] Characterising polymyalgia rheumatica on whole-body 18F-FDG PET/CT: an atlas
    Owen, Claire E.
    Poon, Aurora M. T.
    Liu, Bonnia
    Liew, David F. L.
    Yap, Lee Pheng
    Yang, Victor
    Leung, Jessica L.
    McMaster, Christopher R.
    Scott, Andrew M.
    Buchanan, Russell R. C.
    RHEUMATOLOGY ADVANCES IN PRACTICE, 2023, 8 (01)