Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats

被引:27
|
作者
Woderer, Stefanie
Henninger, Nadja
Garthe, Claus-Dieter
Kloetzer, Hans Martin
Hajnsek, Martin
Kamecke, Ulrike
Gretz, Norbert
Kraenzlin, Bettina
Pill, Johannes
机构
[1] Roche Diagnost GmbH, D-68305 Mannheim, Germany
[2] Heidelberg Univ, Univ Hosp Mannheim, Med Res Ctr, D-68167 Mannheim, Germany
[3] Roche Diagnost GmbH, A-8020 Graz, Austria
[4] Metecon GmbH, D-68165 Mannheim, Germany
关键词
glucose biosensor; continuous glucose monitoring; glucose oxidase; glucose in interstitial fluid; blood glucose; rat;
D O I
10.1016/j.aca.2006.08.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Glucose monitoring is of importance for success of complex therapeutic interventions in diabetic patients. Its impact on treatment and glycemic control is demonstrated in large clinical trials. Up to eight blood glucose measurements per day are recommended. Notwithstanding, a substantial number of diabetic patients cannot or will not monitor their blood glucose appropriately. Considerable progress in control of disturbed metabolism in diabetic patients can be expected by continuous glucose monitoring. The aim of the study was to evaluate the performance of a new amperometric glucose oxidase-based glucose sensor in vitro and in vivo after subcutaneous implantation into rats. For in vitro testing current output of sensors was measured by exposure to increasing and decreasing glucose concentrations up to 472 mg dL(-1) over a time period of 7 days. After subcutaneous implantation of sensors into interscapular region of male rats glucose in interstitial fluid was evaluated and compared to glucose in arterial blood up to 7 days. Hyper- and hypoglycaemia were induced by intravenous application of glucose and insulin, respectively. Current of each implanted sensor was converted into glucose concentration using the first blood glucose measurement only. A change of current with glucose of 0.35 nA mg(-1) dL(-1) indicates high sensitivity of the sensor in vitro. The response time (90% of steady state) was calculated by approximately 60 s. Test strips for blood glucose measurement as reference for sensor readings was found as an appropriate and rapidly available method in rats by comparison with established hexokinase method in an automated lab analyzer with limits of agreement of +32.8 and -25.7 mg dL(-1) in Bland-Altman analysis. In normo- and hypoglycaemic range sensor readings in interstitial fluid correlated well with blood glucose measurements whereas hyperglycaemia was not reflected by the sensor completely when blood glucose was changing rapidly. The data given characterize a sensor with high sensitivity, long term stability and short response time. A single calibration of the sensor is required only in measurement periods up to 7 days. The findings demonstrate that the sensor is a highly promising candidate for assessment in humans. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [1] Possible cause of glucose oxidase-based sensor failure during clinical use of continuous glucose monitoring system
    Pallayova, M
    Donicova, V
    Donic, V
    DIABETOLOGIA, 2005, 48 : A301 - A301
  • [2] Self-monitoring of blood glucose compared with continuous tissue glucose measurement
    Sternberg, F
    Hoss, U
    Salgado, M
    Pfeiffer, EF
    DIABETES CARE, 1997, 20 (07) : 1208 - 1209
  • [3] Continuous Glucose Monitoring by Real Time Sensor in Interstitial Fluid
    Honsova, Sarka
    Navratil, Tomas
    XXXIII MODERNI ELEKTROCHEMICKE METODY, 2013, : 63 - 67
  • [4] Glucose measurement in interstitial fluid by microdialysis for the calibration of minimally invasive blood glucose monitoring
    Li, Dachao
    Wang, Ridong
    Chong, Hao
    Liu, Yu
    Xu, Kexin
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS XI, 2013, 8572
  • [5] Measurement of glucose concentration in interstitial fluid - an alternative or a supplement to conventional blood glucose monitoring?
    Kublin, Oskar
    Stepien, Mariusz
    CLINICAL DIABETOLOGY, 2019, 8 (02): : 121 - 126
  • [6] CONTINUOUS MONITORING OF BLOOD-GLUCOSE USING GEL ENTRAPPED GLUCOSE OXIDASE
    ODRISCOLL, KF
    KAPOULAS, A
    ALBISSER, AM
    GANDER, R
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1975, 170 (AUG24): : 89 - 89
  • [7] A method for continuous glucose monitoring in interstitial fluid based on microdialysis
    Li, D. (dchli@tju.edu.cn), 1600, Tianjin University (46):
  • [8] Performance of glucose dehydrogenase- and glucose oxidase-based blood glucose meters at high altitude and low temperature
    Öberg, D
    Östenson, CG
    DIABETES CARE, 2005, 28 (05) : 1261 - 1261
  • [9] LOW NOISE MEASUREMENT OF PHOTOCURRENT FOR CONTINUOUS GLUCOSE MONITORING Low Noise Measurement System Enables Continuous Monitoring of Glucose in Subcutaneous Interstitial Fluid
    Cooley, Daniel W.
    Andersen, David R.
    BIODEVICES 2010: PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON BIOMEDICAL ELECTRONICS AND DEVICES, 2010, : 195 - 200
  • [10] Detection of Hypoglycemia with Continuous Interstitial and Traditional Blood Glucose Monitoring Using the FreeStyle Navigator Continuous Glucose Monitoring System
    McGarraugh, Geoffrey
    Bergenstal, Richard
    DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (03) : 145 - 150