Low-temperature blackbodies and facility for calibration of them

被引:2
|
作者
Morozova, SP [1 ]
Morozov, PA [1 ]
Sapritsky, VI [1 ]
Lisiansky, BE [1 ]
Dovgilov, NL [1 ]
Liang, YX [1 ]
Li, YM [1 ]
Yue, WL [1 ]
机构
[1] All Russian Res Inst Opt & Phys Measurements, VNIIOFI, Moscow 119361, Russia
来源
OPTICAL DESIGN AND TESTING | 2002年 / 4927卷
关键词
blackbody; thermal radiation reference; calibration;
D O I
10.1117/12.465976
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The calibration of IR sensors, thermal imagers, instrumentation for remote earth sensing, signature recognition, and low background spacebome radiometers requires development of low- and near-ambient-temperature reference standard sources capable of operating in vacuum at low or medium background conditions. The paper contains the characteristics of blackbodies (BB), developed at the All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI). A novel Medium Background Facility (MBF) intended for calibrating IR sources has been developed at VNIIOFI. The MBF used for the calibration under vacuum conditions (10(-3) Pa) and medium background temperatures includes a Ga Fixed-Point Blackbody (29.76degreesC), Variable Temperature Blackbody (-60degreesC similar to + 80degreesC) and Filter Radiometer. A brief description of the operating principles and specifications of the MBF is presented. The results of a comparison of the Variable Temperature Blackbody and Ga Fixed Point Blackbody in the Medium Background Facility are given. The measurement uncertainties radiance temperature associated with spectral bandwidth are discussed.
引用
收藏
页码:118 / 124
页数:7
相关论文
共 50 条
  • [21] Measuring Systems for Thermometer Calibration in Low-Temperature Range
    Szmyrka-Grzebyk, A.
    Lipinski, L.
    Manuszkiewicz, H.
    Kowal, A.
    Grykalowska, A.
    Jancewicz, D.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (11-12) : 2466 - 2476
  • [22] Comparison of Blackbody Sources for Low-Temperature IR Calibration
    Ljungblad, S.
    Holmsten, M.
    Josefson, L. E.
    Klason, P.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2015, 36 (12) : 3310 - 3319
  • [23] Low-temperature radiator for verification and calibration of radiation thermometers
    A. V. Zuev
    V. A. Rozhkov
    Measurement Techniques, 2013, 56 : 805 - 808
  • [24] Comparison of Blackbody Sources for Low-Temperature IR Calibration
    S. Ljungblad
    M. Holmsten
    L. E. Josefson
    P. Klason
    International Journal of Thermophysics, 2015, 36 : 3310 - 3319
  • [25] A tunable filter comparator for the spectral calibration of near-ambient temperature blackbodies
    Khromchenko, V. B.
    Mekhontsev, S. N.
    Hanssen, L. M.
    INFRARED SPACEBORNE REMOTE SENSING AND INSTRUMENTATION XV, 2007, 6678
  • [26] Calibration and application of radiometer for low-temperature infrared radiation source
    Shi, Qingfeng
    Gao, Yanze
    Li, Zhuo
    Shi, Rui
    AOPC 2021: INFRARED DEVICE AND INFRARED TECHNOLOGY, 2021, 12061
  • [27] SUBMILLIMETER DETECTOR CALIBRATION WITH A LOW-TEMPERATURE REFERENCE FOR SPACE APPLICATIONS
    NOLT, IG
    RADOSTITZ, JV
    KITTEL, P
    DONNELLY, RJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (06): : 700 - 702
  • [28] The low-temperature energy calibration system for the CUORE bolometer array
    Sangiorgio, S.
    Ejzak, L. M.
    Heeger, K. M.
    Maruyama, R. H.
    Nucciotti, A.
    Olcese, M.
    Wise, T. S.
    Woodcraft, A. L.
    LOW TEMPERATURE DETECTORS LTD 13, 2009, 1185 : 677 - +
  • [29] COMPUTER CALCULATION OF CALIBRATION CURVES OF LOW-TEMPERATURE RESISTANCE THERMOMETERS
    KOPYLOV, VN
    MEZHOVDE.LP
    PRIBORY I TEKHNIKA EKSPERIMENTA, 1974, (02): : 252 - 253
  • [30] Calibration of low-temperature ac susceptometers with a copper cylinder standard
    Chen, D. -X.
    Skumryev, V.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (02):