Two-Point Frequency Shift Method for Shear Wave Attenuation Measurement

被引:28
|
作者
Kijanka, Piotr [1 ,2 ]
Urban, Matthew W. [1 ,3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] AGH Univ Sci & Technol, Dept Robot & Mechatron, PL-30059 Krakow, Poland
[3] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
Attenuation; Attenuation measurement; Frequency measurement; Shape; Biomedical measurement; Elastography; Phantoms; Acoustic radiation force (ARF); attenuation; attenuation measuring ultrasound shear wave elastography (AMUSE); frequency shift (FS); full-width at half-maximum (FWHM); phantom; shear wave elastography (SWE); soft tissue; ultrasound; viscoelastic; VISCOELASTIC PROPERTIES; TISSUE ELASTICITY; RADIATION FORCE; ELASTOGRAPHY; PROPAGATION; VISCOSITY; QUANTIFICATION; STEATOSIS; FIBROSIS;
D O I
10.1109/TUFFC.2019.2945620
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound shear wave elastography (SWE) is an increasingly used noninvasive modality for quantitative evaluation of tissue mechanical properties. SWE typically uses an acoustic radiation force to produce laterally propagating shear waves that are tracked in the spatial and temporal domains, in order to obtain the wave velocity. One of the ways to study the viscoelasticity is through studying the shear wave phase velocity dispersion curves. Shear wave attenuation can be also characterized in viscoelastic tissues with methods that use multiple lateral data samples. In this article, we present an alternative method for measuring the shear wave attenuation without using a rheological model two-point frequency shift (2P-FS). The technique uses information related to the amplitude spectra FS of shear waves measured at only two lateral locations. The theoretical basis for the 2P-FS is derived and validated. We examined how the first signal position and the distance between the two locations affect the shear wave attenuation estimation in the 2P-FS method. We tested this new method on digital phantom data created using the local interaction simulation approach (LISA) in viscoelastic media. Moreover, we tested data acquired from custom-made tissue-mimicking viscoelastic phantom experiments and ex vivo porcine liver measurements. We compared results from the 2P-FS method with the other two techniques used for assessing a shear wave attenuation: the FS-based method and the attenuation-measuring ultrasound shear wave elastography (AMUSE) technique. In addition, we evaluated the 2P-FS algorithm with different levels of added white Gaussian noise to the shear wave particle velocity using numerical phantoms. Tests conducted showed that the 2P-FS method gives robust results based on only two measurements and can be used to measure attenuation of viscoelastic soft tissues.
引用
收藏
页码:483 / 496
页数:14
相关论文
共 50 条
  • [11] Two-point microrheology and diffusing wave spectroscopy
    Li, Qi
    Dennis, Kimberly A.
    Lee, Yu-Fan
    Furst, Eric M.
    JOURNAL OF RHEOLOGY, 2023, 67 (06) : 1107 - 1118
  • [12] Two-point method uncertainty during control and measurement of cylindrical element diameters
    Glukhov, V. I.
    Shalay, V. V.
    Radev, H.
    METROLOGY, STANDARDIZATION, QUALITY: THEORY AND PRACTICE, (MSQ-2017), 2018, 998
  • [13] TWO-POINT MEASUREMENT OF HIGH-FREQUENCY DENSITY FLUCTUATIONS BY REFLECTOMETRY IN GAMMA 10
    Sato, T.
    Ikezoe, R.
    Ichimura, M.
    Hirata, M.
    Yokoyama, T.
    Ugajin, Y.
    Iimura, T.
    Saito, Y.
    Annaka, Y.
    Shiratani, A.
    Shima, Y.
    Kohagura, J.
    Yoshikawa, M.
    Imai, T.
    FUSION SCIENCE AND TECHNOLOGY, 2013, 63 (1T) : 328 - 330
  • [14] Novel Two-Point Interpolation DFT Method for Frequency Estimation of Sine-Wave Within a Short Time Window
    Song, Jian
    Zhang, Junhao
    Mingotti, Alessandro
    Peretto, Lorenzo
    Wen, He
    2022 IEEE 12TH INTERNATIONAL WORKSHOP ON APPLIED MEASUREMENTS FOR POWER SYSTEMS (AMPS), 2022,
  • [15] A new method for attenuation coefficient measurement in the liver - Comparison with the spectral shift central frequency method
    Fujii, Y
    Taniguchi, N
    Itoh, K
    Shigeta, K
    Wang, Y
    Tsao, JW
    Kumasaki, K
    Itoh, T
    JOURNAL OF ULTRASOUND IN MEDICINE, 2002, 21 (07) : 783 - 788
  • [16] Two-point, timesaving method for measurement of gastric emptying with diagnostic accuracy comparable to that of the conventional method
    Halkar, RK
    Paszkowski, AL
    Jones, ME
    Galt, JR
    Goldfarb, LR
    Hunter, JG
    Taylor, AT
    RADIOLOGY, 1999, 213 (02) : 599 - 602
  • [17] Ultrasound shear wave attenuation in nonalcoholic fatty liver disease: performance of the revisited frequency-shift method with preclinical and human datasets
    Yazdani, Ladan
    Rafati, Iman
    Olivie, Damien
    Giard, Jeanne-Marie
    Sebastiani, Giada
    Nguyen, Bich N.
    Tang, An
    Cloutier, Guy
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2023, 153 (03):
  • [18] Measurement of frequency-dependent shear wave attenuation coefficients using an oblique incidence pulse-echo ultrasonic method
    Hou, Hongyang
    Zhang, Guangdong
    Song, Yongfeng
    Li, Xiongbing
    Zhang, Shuzeng
    Kundu, Tribikram
    MEASUREMENT, 2024, 228
  • [19] Liquid distinction by capacitive measurement using non-contact simultaneous two-point measurement method
    Soejima, T
    Shida, K
    SICE 2002: PROCEEDINGS OF THE 41ST SICE ANNUAL CONFERENCE, VOLS 1-5, 2002, : 586 - 587
  • [20] Two-point coherence of wave packets in turbulent jets
    Jaunet, V.
    Jordan, P.
    Cavalieri, A. V. G.
    PHYSICAL REVIEW FLUIDS, 2017, 2 (02):