A national dosimetric audit of IMRT

被引:53
|
作者
Budgell, Geoff [1 ]
Berresford, Joe [1 ]
Trainer, Michael [1 ]
Bradshaw, Ellie [1 ]
Sharpe, Peter [2 ]
Williams, Peter [1 ]
机构
[1] Christie NHS Fdn Trust, Manchester M20 4BX, Lancs, England
[2] Natl Phys Lab, Middlesex, England
关键词
IMRT; Audit; Dosimetry; Alanine; Film; Verification; UK RADIOTHERAPY CENTERS; QUANTITATIVE-EVALUATION; DOSE DISTRIBUTIONS; CLINICAL-TRIALS; NECK-CANCER; PHANTOM; BEAMS; HEAD;
D O I
10.1016/j.radonc.2011.03.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: A dosimetric audit of IMRT has been carried out within the UK between June 2009 and March 2010 in order to provide an independent check of safe implementation and to identify problems in the modelling and delivery of IMRT. Methods and materials: A mail based audit involving film and alanine dosimeters was utilized. Measurements were made for each individual field in an IMRT plan isocentrically in a flat water-equivalent phantom at a depth of 5 cm. The films and alanine dosimeters were processed and analysed centrally; additional ion chamber measurements were made by each participating centre. Results: 57 of 62 centres participated, with a total of 78 plans submitted. For the film measurements, all 176 fields from the less complex IMRT plans (including prostate and breast plans) achieved over 95% pixels passing a gamma criterion of 3%/3 mm within the 20% isodose. For the more complex IMRT plans (mainly head and neck) 8/245 fields (3.3%) achieved less than 95% pixels passing a 4%/4 mm gamma criterion. Of the alanine measurements, 4/78 (5.1%) of the measurements differed by >5% from the dose predicted by the treatment planning system. Three of these were large deviations of -77.1%, -29.1% and 14.1% respectively. Excluding the three measurements outside 10%, the mean difference was 0.05% with a standard deviation of 1.5%. The out of tolerance results have been subjected to further investigations. Conclusions: A dosimetric audit has been successfully carried out of IMRT implementation by over 90% of UK radiotherapy departments. The audit shows that modelling and delivery of IMRT is accurate, suggesting that the implementation of IMRT has been carried out safely. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 99 (2011) 246-252
引用
收藏
页码:246 / 252
页数:7
相关论文
共 50 条
  • [21] Results of the IAEA supported national end-to-end audit of the IMRT technique in Poland
    Wesolowska, Paulina
    Slusarczyk-Kacprzyk, Wioletta
    Fillmann, Marta
    Kazantsev, Pavel
    Bulski, Wojciech
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 116
  • [22] A Dosimetric Comparison of VMAT and IMRT in Sinonasal Cancers
    Gupta, S.
    Laviraj, M. A.
    Haresh, K. P.
    Sharma, D. N.
    Pandey, D.
    Prabhakar, R.
    Rajendran, M.
    Julka, P. K.
    Rath, G. K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 88 (02): : 493 - 493
  • [23] DOSIMETRIC AND RADIOBIOLOGIC EVALUATION OF BIOLOGICALLY ADAPTED IMRT
    Rodal, J.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S74 - S74
  • [24] Supraclavicular IMRT in breast cancer: A dosimetric study
    Gunn, G. B.
    Endres, E. J.
    Hatch, S. S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S522 - S522
  • [25] Dosimetric implications of intentional skin dose with IMRT
    Gulliford, S
    James, H
    MacKenzi, C
    Le Vay, J
    Sherwin, E
    Smith, S
    Perry, R
    Poynter, A
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S108 - S108
  • [26] Dosimetric verification needs for conformal therapy and IMRT
    Fraass, B
    Moran, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1358 - 1358
  • [27] Unbiased Metrics of Dosimetric Variation in IMRT Planning
    Nohadani, O.
    Srivastava, S.
    Medawar, C.
    Das, I. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S758 - S758
  • [28] Statistical process control for IMRT dosimetric verification
    Breen, Stephen L.
    Moseley, Douglas J.
    Zhang, Beibei
    Sharpe, Michael B.
    MEDICAL PHYSICS, 2008, 35 (10) : 4417 - 4425
  • [29] Dosimetric adaptive IMRT driven by fiducial points
    Crijns, Wouter
    Van Herck, Hans
    Defraene, Gilles
    Van den Bergh, Laura
    Slagmolen, Pieter
    Haustermans, Karin
    Maes, Frederik
    Van den Heuvel, Frank
    MEDICAL PHYSICS, 2014, 41 (06)
  • [30] A method for dosimetry audit of IMRT and VMAT
    Adolfsson, E.
    Gustafsson, H.
    Lund, E.
    Olsson, S.
    Tedgren, A. Carlsson
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S285 - S285