Spectral matching consideration in the design of a novel x-ray image intensifier

被引:3
|
作者
Yu, Chunyu [1 ]
Chang, Benkang
Wei, Dianxiu
机构
[1] Peking Univ, Shenzhen Grad Sch, Informat Engn Inst, Shenzhen 518055, Guangdong, Peoples R China
[2] Nanjing Univ Sci & Technol, Sch Elect Engn & Optoelect Technol, Nanjing 210094, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2006年 / 77卷 / 12期
关键词
D O I
10.1063/1.2400022
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this article, a novel x-ray image intensifier is introduced. It is mainly composed of an x-ray intensifying screen and a low-light-level (L-3) image intensifier. In order to obtain a bright enough image, the spectral compatibility among three different combinations of the x-ray intensifying screen and the photocathode is analyzed. The comparison indicates that a (Zn,Cd)S:Ag screen is superior to either a CaWO4 screen or a Gd2O2S:Tb screen when combined with the Super S-25 photocathode and spectral compatibility analysis is a useful guide when designing an optoelectronic imaging device. (c) 2006 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 50 条
  • [31] Component Level Modular Design of a Solid State X-ray Image Intensifier for an MxN Array
    Huang, Ying
    Qu, Bin
    Sharma, Prateek
    Kuhls-Gilcrist, Andrew
    Wang, Weiyuan
    Titus, Albert H.
    Cartwright, Alexander N.
    Bednarek, Daniel R.
    Rudin, Stephen
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2714 - 2717
  • [32] The use of a Hamamatsu X-ray image intensifier with a cooled CCD as a solution X-ray scattering detector
    Fujisawa, T
    Inoko, Y
    Yagi, N
    JOURNAL OF SYNCHROTRON RADIATION, 1999, 6 : 1106 - 1114
  • [33] AN X-RAY IMAGE INTENSIFIER SYSTEM FOR PRECISION WAVELENGTH DISPERSIVE-X-RAY SPECTROSCOPY
    HAILEY, CJ
    LUPTON, JH
    SIEGMUND, OHW
    STEWART, RE
    ZIOCK, KP
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (08): : 2121 - 2126
  • [34] An X-ray image intensifier for microsecond time-resolved experiments
    Yagi, N.
    Aoyama, K.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [35] Measurement of the effective size of the input phosphor of an x-ray image intensifier
    Feldman, A
    Cox, R
    Rosser, J
    MEDICAL PHYSICS, 1997, 24 (10) : 1641 - 1642
  • [36] X-ray image intensifier assisted placement of a thoracic epidural catheter
    Johnson, T
    Haslett, R
    ANAESTHESIA, 2000, 55 (08) : 822 - 823
  • [37] Flat X-ray image intensifier system for real time fluoroscopy
    Onihashi, H
    Aida, H
    Nou, K
    Noji, T
    Murakoshi, Y
    Ito, K
    Saito, K
    Koma, J
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 509 - 515
  • [38] X-RAY IMAGE INTENSIFIER WITH THIN METAL INPUT WINDOW.
    Sano, Testu
    Harao, Norio
    Denshi Tokyo/Electron Tokyo, 1980, (19): : 99 - 102
  • [39] X-RAY TUBE AND IMAGE INTENSIFIER FOR PRESENT-DAY APPLICATION
    SWINBOURNE, R
    AUSTRALASIAN RADIOLOGY, 1978, 22 (03): : 204 - 214
  • [40] The Next Generation Solid State X-Ray Image Intensifier (SSXII)
    Kuhls-Gilcrist, A.
    Bednarek, D.
    Rudin, S.
    MEDICAL PHYSICS, 2010, 37 (06)