Influence of the number and position of stripped screws on plate-screw construct biomechanical properties

被引:1
|
作者
Garces, Gerardo L. [1 ]
Yanez, Alejandro [2 ]
Cuadrado, Alberto [2 ]
Martel, Oscar [2 ]
机构
[1] Univ Las Palmas Gran Canaria, Orthopaed Surg, Dept Med & Surg Sci, Las Palmas Gran Canaria, Spain
[2] Univ Las Palmas Gran Canaria, Dept Mech Engn, Las Palmas Gran Canaria, Spain
关键词
screw stripping; stiffness; plate-screw construct; DYNAMIC COMPRESSION PLATE; ANKLE FRACTURE FIXATION; OSTEOPOROTIC BONE; LOCKING ELEMENTS; STRAIN; STRENGTH; MODEL; STABILITY; STIFFNESS; SYSTEM;
D O I
10.1016/S0020-1383(17)30795-7
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Screw stripping is a common situation in fracture fixation, particularly in osteopenic bone treatment. Surgeons' perception of screw stripping is relatively poor and the real number of loose screws in every plate-screw construct is unknown. The biomechanical and clinical implications of the different possible screw-stripping situations are also unidentified. In this study, construct stiffness in different scenarios of stripped screws is investigated. Method: A bone surrogate comminuted osteoporotic fracture was fixed with four screws in both sides of the fracture gap in 75 specimens. In four groups, one or two screws closest or distal to the gap were over-tightened and left in place in one part of the construct and the remaining screws were tightened with 0.3 N m torque (four groups). In the fifth group (control), all the screws were tightened with 0.3 N m torque. Construct stiffness was tested in terms of compression, bending, and torsion for 1000 cycles. Results: When one or two screws closest to the gap were stripped, stiffness only decreased by, respectively, 5.7% or 7.6% under compression and 4.7% or 6.7% under bending; however, stiffness in torsion was 15.1% or 32%, respectively, lower than the initial stiffness. When a screw distal to the gap was stripped, the stiffness decreased by 28% under bending and 10% under compression; no change was noted under torsion. When two screws distal to the gap were stripped, the stiffness decreased by 11% in compression, collapsed under bending, and decreased by 8% under torsion. Conclusions: Position and number of stripped screws affect the biomechanical properties of a construct in different ways, depending on the acting forces. (c) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:S54 / S59
页数:6
相关论文
共 50 条
  • [21] Intramedullary Threaded Nail Fixation Versus Plate and Screw Construct in Metacarpal Neck Fractures: A Biomechanical Study
    Wallace, Doyle R.
    Shiver, A. Luke
    Pulliam, Steven K.
    Byrd, Bruce M.
    McGee-Lawrence, Meghan E.
    Snoddy, Mark C.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2023, 31 (11) : E516 - E522
  • [22] Biomechanical investigation of the type and configuration of screws used in high tibial osteotomy with titanium locking plate and screw fixation
    Chen, Yen-Nien
    Chang, Chih-Wei
    Li, Chun-Ting
    Chen, Chih-Hsien
    Chung, Chi-Rung
    Chang, Chih-Han
    Peng, Yao-Te
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)
  • [23] Biomechanical investigation of the type and configuration of screws used in high tibial osteotomy with titanium locking plate and screw fixation
    Yen-Nien Chen
    Chih-Wei Chang
    Chun-Ting Li
    Chih-Hsien Chen
    Chi-Rung Chung
    Chih-Han Chang
    Yao-Te Peng
    Journal of Orthopaedic Surgery and Research, 14
  • [24] Fixation properties of a biodegradable "free-form" osteosynthesis plate with screws with cut-off screw heads: Biomechanical evaluation over 26 weeks
    Vaananen, Petteri
    Nurmi, Janne T.
    Lappalainen, Reijo
    Jank, Siegfried
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 107 (04): : 462 - 468
  • [25] Comparative biomechanical analysis of two techniques of radiolunate plate-fusion: Oblique screw vs. parallel screws
    Gaulke, Ralph
    Oszwald, Markus
    Liodakis, Emmanouil
    Stueber, Volker
    Bachmann, Soeren
    Krettek, Christian
    TECHNOLOGY AND HEALTH CARE, 2012, 20 (03) : 225 - 232
  • [26] Biomechanical analysis of distal femur fracture fixation: Fixed-angle screw-plate construct versus condylar blade plate
    Higgins, Thomas F.
    Pittman, Gavin
    Hines, Jerod
    Bachus, Kent N.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (01) : 43 - 46
  • [27] The influence of distal screw length on the primary stability of volar plate osteosynthesis—a biomechanical study
    Sebastian F. Baumbach
    Alexander Synek
    Hannes Traxler
    Wolf Mutschler
    Dieter Pahr
    Yan Chevalier
    Journal of Orthopaedic Surgery and Research, 10
  • [28] BIOMECHANICAL ANALYSIS OF BONE-MINERAL DENSITY, INSERTION TECHNIQUE, SCREW TORQUE, AND HOLDING STRENGTH OF ANTERIOR CERVICAL PLATE SCREWS
    RYKEN, TC
    CLAUSEN, JD
    TRAYNELIS, VC
    GOEL, VK
    JOURNAL OF NEUROSURGERY, 1995, 83 (02) : 324 - 329
  • [29] The effect of locked screw angulation on the biomechanical properties of the SPS Free-Block plate
    Merino, M. K. A.
    Rahal, S. C.
    Ribeiro, C. R.
    Padovani, C. R.
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2013, 26 (02) : 117 - 122
  • [30] Biomechanical properties of anterior thoracolumbar multisegmental fixation - An analysis of construct stiffness and screw-rod strain
    Oda, I
    Cunningham, BW
    Lee, GA
    Abumi, K
    Kaneda, K
    McAfee, PC
    SPINE, 2000, 25 (18) : 2303 - 2311