BIOMECHANICAL ANALYSIS OF BONE-MINERAL DENSITY, INSERTION TECHNIQUE, SCREW TORQUE, AND HOLDING STRENGTH OF ANTERIOR CERVICAL PLATE SCREWS

被引:68
|
作者
RYKEN, TC
CLAUSEN, JD
TRAYNELIS, VC
GOEL, VK
机构
[1] UNIV IOWA,DIV NEUROSURG,IOWA CITY,IA
[2] UNIV IOWA,DEPT BIOMED ENGN,IOWA CITY,IA 52242
关键词
BIOMECHANICS; BONE MINERAL DENSITY; CERVICAL SPINE; TORQUE INSTRUMENTATION;
D O I
10.3171/jns.1995.83.2.0324
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The bone mineral density (BMD) of 99 cadaveric cervical vertebral bodies (C3-7) was determined using dual x-ray absorptiometry. The vertebral bodies were randomly assigned to receive either a unicortical (51 bodies) or bicortical (48 bodies) Caspar cervical plating screw. The initial insertion torque was measured using a digital electronic torque wrench, and the force required to withdraw the screw from the vertebral body was determined. The mean BMD for the total group of 99 was 0.787 +/- 0.154 g/cm(2), the mean insertion torque was 0.367 +/- 0.243 newton-meters, and the mean pullout force was 210.4 +/- 158.1 newtons. A significant correlation was noted between BMD and torque (p < 0.0001, r = 0.42), BMD and pullout force (p < 0.0001, r = 0.54), and torque and pullout force (p < 0.0001, r = 0.88). Although the BMD of the unicortical and bicortical groups was equivalent (p = 0.92), the insertion torque and pullout force differed significantly (p = 0.02 and p = 0.008, respectively) for the unicortical and bicortical groups. A holding index for each screw and insertion technique was defined as the product of the BMD and insertion torque. The calculated holding index and resultant pullout force were significantly correlated for both techniques of screw insertion (r = 0.92), and a significant difference in holding index was observed with unicortical versus bicortical screw placement (p = 0.04). The determination of BMD and measurement of insertion torque to create a unique holding index provides an assessment of bone-screw interaction and holding strength of the screw, both of which impact on the resultant stability of cervical instrumentation. As the number of cervical plating systems increases, the determination of a holding index for various screws and insertion techniques may assist in the comparison of cervical instrumentation.
引用
收藏
页码:324 / 329
页数:6
相关论文
共 34 条
  • [1] STRENGTH OF ANTERIOR VERTEBRAL SCREW FIXATION IN RELATIONSHIP TO BONE-MINERAL DENSITY
    LIM, TH
    AN, HS
    EVANICH, C
    HASANOGLU, KY
    MCGRADY, L
    WILSON, CR
    JOURNAL OF SPINAL DISORDERS, 1995, 8 (02): : 121 - 125
  • [2] Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine
    Reitman, CA
    Nguyen, L
    Fogel, GR
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2004, 17 (04): : 306 - 311
  • [3] Divergent Configuration Improves Insertion Torque and Pullout Strength of Anterior Cervical Screws
    Xavier, Fred
    Saha, Subrata
    2016 32ND SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE (SBEC), 2016, : 98 - 98
  • [4] Effect of bone mineral density and amorphous diamond coatings on insertion torque of bone screws
    Koistinen, A
    Santavirta, SS
    Kröger, H
    Lappalainen, R
    BIOMATERIALS, 2005, 26 (28) : 5687 - 5694
  • [5] Biomechanical comparison of plate and screw fixation in anterior pelvic ring fractures with low bone mineral density
    Acklin, Yves P.
    Zderic, Ivan
    Buschbaum, Jan
    Varga, Peter
    Inzana, Jason A.
    Grechenig, Stephan
    Richards, R. Geoff
    Gueorguiev, Boyko
    Schmitz, Paul
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (07): : 1456 - 1460
  • [6] Does cancellous screw insertion torque depend on bone mineral density and/or microarchitecture?
    Ab-Lazid, Rosidah
    Perilli, Egon
    Ryan, Melissa K.
    Costi, John J.
    Reynolds, Karen J.
    JOURNAL OF BIOMECHANICS, 2014, 47 (02) : 347 - 353
  • [7] Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength
    Weiser, Lukas
    Huber, Gerd
    Sellenschloh, Kay
    Viezens, Lennart
    Pueschel, Klaus
    Morlock, Michael M.
    Lehmann, Wolfgang
    EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2891 - 2897
  • [8] Insufficient stability of pedicle screws in osteoporotic vertebrae: biomechanical correlation of bone mineral density and pedicle screw fixation strength
    Lukas Weiser
    Gerd Huber
    Kay Sellenschloh
    Lennart Viezens
    Klaus Püschel
    Michael M. Morlock
    Wolfgang Lehmann
    European Spine Journal, 2017, 26 : 2891 - 2897
  • [9] Anterior cervical plate fixation: A biomechanical study to evaluate the effects of plate design, endplate preparation, and bone mineral density
    Dvorak, MF
    Pitzen, T
    Zhu, QG
    Gordon, JD
    Fisher, CG
    Oxland, TR
    SPINE, 2005, 30 (03) : 294 - 301
  • [10] Interference screw fixation strength of a quadrupled hamstring tendon graft is directly related to bone mineral density and insertion torque
    Brand, JC
    Pienkowski, D
    Steenlage, E
    Hamilton, D
    Johnson, DL
    Caborn, DNM
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (05): : 705 - 710