Water Vapor Interference Correction in a Non Dispersive Infrared Multi-Gas Analyzer

被引:16
|
作者
Sun You-Wen [1 ]
Liu Wen-Qing [1 ]
Zeng Yi [1 ]
Wang Shi-Mei [1 ]
Huang Shu-Hua [1 ]
Xie Pin-Hua [1 ]
Xiao-Man, Yu [1 ]
机构
[1] Chinese Acad Sci, Key Lab Environm Opt & Technol, Anhui Inst Opt & Fine Mech, Hefei 230031, Peoples R China
基金
中国国家自然科学基金;
关键词
SYSTEM; DESIGN;
D O I
10.1088/0256-307X/28/7/073302
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We demonstrate an effective method to eliminate the interfering effect of water vapor in a non-dispersive infrared multi-gas analyzer. The response coefficients of water vapor at each filter channel are measured from the humidity of the ambient air. Based on the proposed method, the water vapor interference is corrected with the measured response coefficients. By deducting the absorbance of each filter channel related to water vapor, the measuring precision of the analyzer is improved significantly and the concentration retrieval correlation accuracy of each target gas is more than 99%.
引用
收藏
页数:4
相关论文
共 50 条
  • [21] The development system of portable of CO2 non-dispersive infrared detector gas analyzer
    Yeh, T
    TECHNICAL DIGEST OF THE SEVENTH INTERNATIONAL MEETING ON CHEMICAL SENSORS, 1998, : 675 - 685
  • [22] Limitations of gas filter correlation: A case study on carbon monoxide non-dispersive infrared analyzer
    Dinh, Trieu-Vuong
    Ahn, Ji-Won
    Choi, In-Young
    Song, Kyu-Yong
    Chung, Chang-Ho
    Kim, Jo-Chun
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 243 : 684 - 689
  • [23] Mathematical model of dispersive infrared gas analyzer based on pyroelectric detector
    Zhang, YH
    Liu, JH
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2004, 24 (03) : 376 - 378
  • [24] CONTINUOUS FLOW INFRARED GAS ANALYZER EMPLOYING INTERFERENCE FILTERS
    PARSONS, JL
    IRLAND, MJ
    BRYAN, FR
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1956, 46 (03) : 164 - 166
  • [25] Integrated-path multi-gas sensor using near-infrared diode lasers: An alternative to vehicle-driven point gas analyzer
    Ashik, A. S.
    Rodrigo, Peter John
    Larsen, Henning E.
    Vechi, Nathalia Thygsen
    Kissas, Konstantinos
    Fredenslund, Anders Michael
    Onster, Jacob G. M.
    Scheutz, Charlotte
    Pedersen, Christian
    SENSORS AND ACTUATORS B-CHEMICAL, 2024, 412
  • [26] PRESSURE EFFECT OF NON-DISPERSIVE IR GAS ANALYZER
    YONEZAWA, T
    NIKI, E
    BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 1977, 50 (02) : 341 - 344
  • [27] Research on the Method of Interference Correction for Nondispersive Infrared Multi-Component Gas Analysis
    Sun You-wen
    Liu Wen-qing
    Wang Shi-mei
    Huang Shu-hua
    Yu Xiao-man
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31 (10) : 2719 - 2724
  • [28] Modeling of Non-Dispersive Infrared Gas Sensors
    Dietachmayr, F. W.
    Zagar, B. G.
    SENSORS AND ELECTRONIC INSTRUMENTATION ADVANCES (SEIA), 2016, : 95 - 99
  • [29] MATHEMATIC MODELING OF NON DISPERSIVE INFRARED GAS SENSORS
    Dietachmayr, Florian W.
    Zagar, Bernhard C.
    5. TAGUNG INNOVATION MESSTECHNIK, 2017, : 47 - 52
  • [30] A NEW AOFT-BASED SOLID-STATE INSITU MULTI-GAS ANALYZER FOR STACK EMISSIONS
    NELSON, RL