Monte Carlo calculations of electron beam output factors for a medical linear accelerator

被引:71
|
作者
Kapur, A [1 ]
Ma, CM [1 ]
Mok, EC [1 ]
Findley, DO [1 ]
Boyer, AL [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiat Oncol, Div Radiat Phys, Stanford, CA 94305 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1998年 / 43卷 / 12期
关键词
D O I
10.1088/0031-9155/43/12/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the application of the Monte Carlo technique to the calculation and analysis of output factors for electron beams used in radiotherapy. The code EGS4/BEAM was used to obtain phase-space files for 6, 12 and 20 MeV clinical electron beams from a scattering-foil linac (Varian Clinac 2100C) for a clinically representative range of applicator and square or rectangular insert combinations. The source-to-surface distance used was 100 cm. The field sizes ranged from 1 x 1 cm(2) to 20 x 20 cm(2). These phase-space files were analysed to study the intrinsic beam characteristics and used as sour ce input for relative dose and output factor computations in homogeneous water phantoms using the code EGS4/DOSXYZ. The calculated relative central-axis depth-dose and transverse dose profiles at various depths of clinical interest agreed with the corresponding measured dose profiles to within 2% of the maximum dose. Calculated output factors for the fields studied agreed with measured output factors to about 2%. This demonstrated that for the Varian Clinac 2100C linear accelerator, electron beam dose calculations in homogeneous water phantoms can be performed accurately at the 2% level using Monte Carlo simulations.
引用
收藏
页码:3479 / 3494
页数:16
相关论文
共 50 条
  • [21] Monte Carlo calculations of correction factors for plastic phantoms in clinical photon and electron beam dosimetry
    Araki, Fujio
    Hanyu, Yuji
    Fukuoka, Miyoko
    Matsumoto, Kenji
    Okumura, Masahiko
    Oguchi, Hiroshi
    MEDICAL PHYSICS, 2009, 36 (07) : 2992 - 3001
  • [22] Monte Carlo calculations of electron beam quality conversion factors for several ion chamber types
    Muir, B. R.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2014, 41 (11)
  • [23] Fast Monte Carlo calculation of linear accelerator head scatter factors
    MacPherson, M
    Badeen, A
    Wilkins, D
    Gerig, L
    MEDICAL PHYSICS, 2002, 29 (06) : 1203 - 1203
  • [24] Modeling electron transport and Monte Carlo calculations for electron beam dosimetry (WIP)
    Vonk, DT
    Rez, P
    Sapareto, S
    MEDICAL PHYSICS, 2003, 30 (06) : 1513 - 1513
  • [25] A PARAMETERIZATION OF THE ELECTRON-BEAM OUTPUT FACTORS OF A 25-MEV LINEAR-ACCELERATOR
    MCPARLAND, BJ
    MEDICAL PHYSICS, 1987, 14 (04) : 665 - 669
  • [26] Electron beam therapy at extended SSDs: an analysis of output correction factors for a Mitsubishi linear accelerator
    Rajasekar, D
    Datta, NR
    Das, KJM
    Ayyagari, S
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (18): : 3301 - 3311
  • [27] DETERMINATION OF ELECTRON-BEAM OUTPUT FACTORS FOR A 20-MEV LINEAR-ACCELERATOR
    MILLS, MD
    HOGSTROM, KR
    FIELDS, RS
    MEDICAL PHYSICS, 1985, 12 (04) : 473 - 476
  • [28] COMPARISON OF THREE-DIMENSIONAL MONTE CARLO SIMULATION AND THE PENCIL BEAM ALGORITHM FOR AN ELECTRON BEAM FROM A LINEAR ACCELERATOR.
    Manfredotti, C.
    Nastasi, U.
    Ragona, R.
    Anglesio, S.
    Nuclear instruments and methods in physics research, 1985, A255 (1-2): : 355 - 359
  • [29] Electron beam commissioning for a siemens PRIMUS accelerator for Monte Carlo treatment planning
    Qin, L
    Li, J
    Deng, J
    Paskalev, K
    Ding, M
    Fourkal, E
    Ma, C
    MEDICAL PHYSICS, 2003, 30 (06) : 1516 - 1516
  • [30] Accuracy of electron beam Monte Carlo calculations in radiotherapy treatment planning
    Arends, M
    Ausma, A
    van 't Veld, A
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S43 - S43