Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model

被引:15
|
作者
Struffert, Tobias [1 ]
Ott, Sabine [1 ]
Adamek, Edyta [1 ]
Schwarz, Marc [1 ]
Engelhorn, Tobias [1 ]
Kloska, Stephan [1 ]
Deuerling-Zheng, Yu [2 ]
Doerfler, Arnd [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Neuroradiol, D-91054 Erlangen, Germany
[2] Siemens AG, Healthcare Sector, D-91301 Forchheim, Germany
关键词
Angiography; Flat detector computed tomography; Animal model; Multislice computed tomography; Intracranial stent; SMALL-VESSEL STENTS; ANEURYSM MODEL; COIL EMBOLIZATION; RESTENOSIS; VISUALIZATION; ANGIOPLASTY; HEMORRHAGE; STENOSES;
D O I
10.1007/s00330-011-2093-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Careful follow up is necessary after intracranial stenting because in-stent restenosis (ISR) or residual stenosis (RS) is not rare. A minimally invasive follow-up imaging technique is desirable. The objective was to compare the visualisation of stents in Flat Detector-CT Angiography (FD-CTA) after intravenous contrast medium injection (i.v.) with Multi Detector Computed Tomography Angiography (MD-CTA) and Digital Subtracted Angiography (DSA) in an animal model. Stents were implanted in the carotid artery of 12 rabbits. In 6 a residual stenosis (RS) was surgically created. Imaging was performed using FD-CTA, MD-CTA and DSA. Measurements of the inner and outer diameter and cross-section area of the stents were performed. Stenosis grade was calculated. In subjective evaluation FD-CTA was superior to MD-CTA. FD-CTA was more accurate compared with DSA than MD-CTA. Cross-sectional area of the stent lumen was significantly larger (p < 0.05) in FD-CTA in comparison to MD-CTA. Accurate evaluation of stenosis was impossible in MD-CTA. There was no statistically significant difference in the stenosis grade of DSA and FD-CTA. Our results show that visualisation of stent and stenosis using intravenous FD-CTA compares favourably with DSA and may replace DSA in the follow-up of patients treated with intracranial stents.
引用
收藏
页码:1779 / 1787
页数:9
相关论文
共 50 条
  • [1] Flat-detector computed tomography in the assessment of intracranial stents: comparison with multi detector CT and conventional angiography in a new animal model
    Tobias Struffert
    Sabine Ott
    Edyta Adamek
    Marc Schwarz
    Tobias Engelhorn
    Stephan Kloska
    Yu Deuerling-Zheng
    Arnd Doerfler
    European Radiology, 2011, 21 : 1779 - 1787
  • [2] Flat-detector computed tomography with intravenous contrast material application in experimental aneurysms: Comparison with multislice CT and conventional angiography
    Struffert, Tobias
    Doelken, Marc
    Adamek, Edyta
    Schwarz, Marc
    Engelhorn, Tobias
    Kloska, Stephan
    Ott, Sabine
    Doerfler, Arnd
    ACTA RADIOLOGICA, 2010, 51 (04) : 431 - 437
  • [3] Flat-detector computed tomography (FD-CT)
    Kalender, Willi A.
    Kyriakou, Yiannis
    EUROPEAN RADIOLOGY, 2007, 17 (11) : 2767 - 2779
  • [4] Flat-detector computed tomography in diagnostic and interventional neuroradiology
    Struffert, T.
    Doerfler, A.
    RADIOLOGE, 2009, 49 (09): : 820 - 829
  • [5] The Feasibility of Intravenous Flat-Detector CT (IV FDCT) Angiography for Intracranial Arterial Stenosis
    Sheen, Seung
    Jeon, Jin Sue
    Kim, Jeong Eun
    Bang, Jae Seung
    Son, Young-Je
    STROKE, 2013, 44 (02)
  • [6] Detection of wall and neck calcification of unruptured intracranial aneurysms with flat-detector computed tomography
    Kizilkilic, Osman
    Huseynov, Eldeniz
    Kandemirli, Sedat G.
    Kocer, Naci
    Islak, Civan
    INTERVENTIONAL NEURORADIOLOGY, 2016, 22 (03) : 293 - 298
  • [7] Intravenous Flat-Detector Computed Tomography Angiography for High-Grade Carotid Stenosis
    Jeon, Jin Sue
    Sheen, Seung Hun
    Kim, Heung Cheol
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (02) : 242 - 246
  • [8] Time-Resolved Perfusion Imaging at the Angiography Suite Preclinical Comparison of a New Flat-Detector Application to Computed Tomography Perfusion
    Juergens, Julian H. W.
    Schulz, Nadine
    Wybranski, Christian
    Seidensticker, Max
    Streit, Sebastian
    Brauner, Jan
    Wohlgemuth, Walter A.
    Deuerling-Zheng, Yu
    RickeMd, Jens
    Dudeck, Oliver
    INVESTIGATIVE RADIOLOGY, 2015, 50 (02) : 108 - 113
  • [9] Assessment of cavopulmonary connections by advanced imaging: value of flat-detector computed tomography
    Gloeckler, Martin
    Koch, Andreas
    Halbfass, Julia
    Greim, Verena
    Rueffer, Andre
    Cesnjevar, Robert
    Achenbach, Stephan
    Dittrich, Sven
    CARDIOLOGY IN THE YOUNG, 2013, 23 (01) : 18 - 26
  • [10] Flat-detector computed tomography evaluation in an experimental animal aneurysm model after endovascular treatment: A pilot study
    Ott, Sabine
    Goelitz, Philipp
    Adamek, Edyta
    Royalty, Kevin
    Doerfler, Arnd
    Struffert, Tobias
    INTERVENTIONAL NEURORADIOLOGY, 2015, 21 (04) : 490 - 498