Radioactive source strength effect on gamma radiation monitoring with a NaI (Tl) scintillation detector

被引:0
|
作者
Yeboah, E. [1 ]
Yakovlev, G. A. [2 ]
Smirnov, S. V. [3 ]
Yakovleva, V. S. [1 ]
机构
[1] Natl Res Tomsk Polytech Univ, 30 Lenina Ave, Tomsk 634050, Russia
[2] Tomsk State Univ, 36 Lenina Ave, Tomsk 634050, Russia
[3] Russian Acad Sci, Inst Monitoring Climat & Ecol Syst, Siberian Branch, 10-3 Acad Ave, Tomsk 634055, Russia
关键词
Detector alignment and calibration methods (lasers; sources; particle-beams); Gamma detectors (scintillators; CZT; HPGe; HgI etc); Radiation monitoring; DOSE-RATE; TEMPERATURE; NAI(TL); LABR3(CE);
D O I
10.1088/1748-0221/17/05/P05041
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The accurate monitoring of gamma radiation doses in the environment has become essential due to its effect on human health. In this study, the temperature dependence of NaI (Tl) scintillation detectors based on the daily/annually measured dose rate in the environment as well as the importance of selecting the appropriate radioactive source strength when calibrating NaI (Tl) detectors for gamma radiation monitoring were investigated. The temperature correction coefficients discovered during the calibration of the BDKG-03 detector by Atomtex were verified, and whether the time duration of the measurement interval has an effect on readings of gamma radiation was also investigated in the city of Tomsk. NaI (Tl) scintillation detectors were used to monitor the gamma background radiation in the environment. The detectors were calibrated with both high and low radioactive sources to obtain a temperature correction coefficient in order to stabilize the influence of temperature change on the detector at different time intervals. This was used to study the correlation between the daily and annual dose rates of low gamma background radiation. The results showed that there was a shift in the spectrum of the daily and annual dose rates calculated using the algorithm obtained when the detector was calibrated with a highly radioactive source to the position of the constant coefficient for low-level dose. However, the ones obtained when the detector was calibrated with a low radioactive source and that of the constant coefficient for a low-level dose overlapped each other. This demonstrated that the type of radioactive source used in detector calibration during manufacture has an effect on the temperature correction coefficient, which in turn has an effect on the accuracy of the ambient dose rate when used to monitor gamma radiation. The duration of the time interval for measurement was found to be very important since it has an effect on dose rate readings.
引用
收藏
页数:16
相关论文
共 50 条
  • [41] Standardization of a NaI(Tl) scintillation detector and its use on environments for Geology teaching in a Brazilian university
    Eder Queiroz Barbosa
    Daniel Marcos Bonotto
    Gabrielle Roveratti
    Environmental Earth Sciences, 2018, 77
  • [42] Calculation of spectrum to dose conversion factors for a NaI(Tl) scintillation detector using the response matrix
    Jun, JS
    Yi, CY
    Chai, HS
    Cho, H
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 1995, 28 (06) : 716 - 726
  • [43] THE SCINTILLATION EFFICIENCY OF SODIUM AND IODINE RECOILS IN A NAI(TL) DETECTOR FOR DARK-MATTER SEARCHES
    SPOONER, NJC
    DAVIES, GJ
    DAVIES, JD
    PYLE, GJ
    BUCKNELL, TD
    SQUIER, GTA
    LEWIN, JD
    SMITH, PF
    PHYSICS LETTERS B, 1994, 321 (1-2) : 156 - 160
  • [44] Semi-empirical method for determining the density of liquids using a NaI(Tl) scintillation detector
    Huynh Dinh Chuong
    Nguyen Thi My Le
    Hoang Duc Tam
    APPLIED RADIATION AND ISOTOPES, 2019, 152 : 109 - 114
  • [45] RESPONSE OF 7.6 CM DIAMETER AND 7.6 CM LENGTH NAI(TL) SCINTILLATION DETECTOR TO NEUTRONS
    URABE, I
    JOURNAL OF THE ATOMIC ENERGY SOCIETY OF JAPAN, 1988, 30 (01): : 64 - 70
  • [46] Standardization of a NaI(Tl) scintillation detector and its use on environments for Geology teaching in a Brazilian university
    Barbosa, Eder Queiroz
    Bonotto, Daniel Marcos
    Roveratti, Gabrielle
    ENVIRONMENTAL EARTH SCIENCES, 2018, 77 (03)
  • [47] Efficiency measurement of NaI(Tl) detector for a long tube volume source of 6.13MeV gamma-ray
    Shi, ZR
    Xu, K
    Ding, SY
    Wang, SQ
    Guo, ZR
    Zuo, LZ
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2005, 29 (02): : 124 - 129
  • [48] Temporal Fluctuations in Indoor Background Gamma Radiation Using NaI(Tl)
    Smith, Travis
    Cao, Sana
    Kearfott, Kimberlee J.
    HEALTH PHYSICS, 2018, 114 (03): : 360 - 372
  • [49] Calculation of efficiency and resolution of a hexagonal NaI(Tl) detector as a function of source position
    Hamzawy, A.
    Badawi, M. S.
    Ruskov, I. N.
    Grozdanov, D. N.
    Badawi, M. I.
    Thabet, A. A.
    Kopatch, Yu. N.
    Fedorov, N. A.
    Salem, B. A.
    JOURNAL OF INSTRUMENTATION, 2024, 19 (07):
  • [50] Neutron/gamma scintillation detector for status monitoring of accelerator-driven neutron source IREN
    Nuruyev, S.
    Berikov, D.
    Akbarov, R.
    Ahmadov, G.
    Ahmadov, F.
    Sadigov, A.
    Holik, M.
    Naghiyev, J.
    Madadzada, A.
    Udovichenko, K.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2024, 56 (05) : 1667 - 1671