Radioactive source strength effect on gamma radiation monitoring with a NaI (Tl) scintillation detector

被引:0
|
作者
Yeboah, E. [1 ]
Yakovlev, G. A. [2 ]
Smirnov, S. V. [3 ]
Yakovleva, V. S. [1 ]
机构
[1] Natl Res Tomsk Polytech Univ, 30 Lenina Ave, Tomsk 634050, Russia
[2] Tomsk State Univ, 36 Lenina Ave, Tomsk 634050, Russia
[3] Russian Acad Sci, Inst Monitoring Climat & Ecol Syst, Siberian Branch, 10-3 Acad Ave, Tomsk 634055, Russia
关键词
Detector alignment and calibration methods (lasers; sources; particle-beams); Gamma detectors (scintillators; CZT; HPGe; HgI etc); Radiation monitoring; DOSE-RATE; TEMPERATURE; NAI(TL); LABR3(CE);
D O I
10.1088/1748-0221/17/05/P05041
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The accurate monitoring of gamma radiation doses in the environment has become essential due to its effect on human health. In this study, the temperature dependence of NaI (Tl) scintillation detectors based on the daily/annually measured dose rate in the environment as well as the importance of selecting the appropriate radioactive source strength when calibrating NaI (Tl) detectors for gamma radiation monitoring were investigated. The temperature correction coefficients discovered during the calibration of the BDKG-03 detector by Atomtex were verified, and whether the time duration of the measurement interval has an effect on readings of gamma radiation was also investigated in the city of Tomsk. NaI (Tl) scintillation detectors were used to monitor the gamma background radiation in the environment. The detectors were calibrated with both high and low radioactive sources to obtain a temperature correction coefficient in order to stabilize the influence of temperature change on the detector at different time intervals. This was used to study the correlation between the daily and annual dose rates of low gamma background radiation. The results showed that there was a shift in the spectrum of the daily and annual dose rates calculated using the algorithm obtained when the detector was calibrated with a highly radioactive source to the position of the constant coefficient for low-level dose. However, the ones obtained when the detector was calibrated with a low radioactive source and that of the constant coefficient for a low-level dose overlapped each other. This demonstrated that the type of radioactive source used in detector calibration during manufacture has an effect on the temperature correction coefficient, which in turn has an effect on the accuracy of the ambient dose rate when used to monitor gamma radiation. The duration of the time interval for measurement was found to be very important since it has an effect on dose rate readings.
引用
收藏
页数:16
相关论文
共 50 条
  • [21] EFFECT OF IMPURITIES ON SCINTILLATION - OPTICAL AND THERMOLUMINESCENT PROPERTIES OF NAI(TL)
    SABHARWAL, SC
    KATHURIA, SP
    GHOSH, B
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 255 (03): : 501 - 506
  • [22] CMOS-coupled NaI scintillation detector for gamma decay measurements
    Bergeson, Scott D.
    Ware, Michael J.
    Hawk, Jeremy
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (03):
  • [23] An improved method for liquid density measurement using NaI(Tl) detector and low-strength source
    Hoang Duc Tam
    Nguyen Thi My Le
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2018, 317 (01) : 161 - 168
  • [24] A NAI(TL) SCINTILLATION SPECTROMETER STUDY OF PROTON GAMMA-RAY COINCIDENCES
    ALLEN, RC
    MAY, JE
    RALL, W
    PHYSICAL REVIEW, 1951, 84 (06): : 1203 - 1206
  • [25] An improved method for liquid density measurement using NaI(Tl) detector and low-strength source
    Hoang Duc Tam
    Nguyen Thi My Le
    Journal of Radioanalytical and Nuclear Chemistry, 2018, 317 : 161 - 168
  • [26] Effects of gamma irradiation on some chemicals using an NaI (Tl) detector
    Bhosale, R. R.
    Gaikwad, D. K.
    Pawar, P. P.
    Rode, M. N.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2016, 171 (5-6): : 398 - 407
  • [27] Gamma-Ray Source Positioning Using Array NaI(Tl) Detectors in the Radiation Portal Monitors
    Qibiao Wang
    Chao Deng
    Jianbo Yang
    Rui Shi
    Songbai Zhang
    Xianguo Tuo
    Journal of the Korean Physical Society, 2019, 75 : 196 - 201
  • [28] Longitudinal monitoring of whole body counter NaI(Tl) detector efficiency
    Shypailo, R. J.
    APPLIED RADIATION AND ISOTOPES, 2017, 125 : 74 - 79
  • [29] Gamma-Ray Source Positioning Using Array NaI(Tl) Detectors in the Radiation Portal Monitors
    Wang, Qibiao
    Deng, Chao
    Yang, Jianbo
    Shi, Rui
    Zhang, Songbai
    Tuo, Xianguo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2019, 75 (03) : 196 - 201
  • [30] THE EFFECT OF THERMAL-TREATMENT ON SCINTILLATION AND THERMOLUMINESCENT OUTPUT OF NAI(TL)
    SABHARWAL, SC
    GHOSH, B
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 243 (01): : 225 - 226