Radioactive source strength effect on gamma radiation monitoring with a NaI (Tl) scintillation detector

被引:0
|
作者
Yeboah, E. [1 ]
Yakovlev, G. A. [2 ]
Smirnov, S. V. [3 ]
Yakovleva, V. S. [1 ]
机构
[1] Natl Res Tomsk Polytech Univ, 30 Lenina Ave, Tomsk 634050, Russia
[2] Tomsk State Univ, 36 Lenina Ave, Tomsk 634050, Russia
[3] Russian Acad Sci, Inst Monitoring Climat & Ecol Syst, Siberian Branch, 10-3 Acad Ave, Tomsk 634055, Russia
关键词
Detector alignment and calibration methods (lasers; sources; particle-beams); Gamma detectors (scintillators; CZT; HPGe; HgI etc); Radiation monitoring; DOSE-RATE; TEMPERATURE; NAI(TL); LABR3(CE);
D O I
10.1088/1748-0221/17/05/P05041
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The accurate monitoring of gamma radiation doses in the environment has become essential due to its effect on human health. In this study, the temperature dependence of NaI (Tl) scintillation detectors based on the daily/annually measured dose rate in the environment as well as the importance of selecting the appropriate radioactive source strength when calibrating NaI (Tl) detectors for gamma radiation monitoring were investigated. The temperature correction coefficients discovered during the calibration of the BDKG-03 detector by Atomtex were verified, and whether the time duration of the measurement interval has an effect on readings of gamma radiation was also investigated in the city of Tomsk. NaI (Tl) scintillation detectors were used to monitor the gamma background radiation in the environment. The detectors were calibrated with both high and low radioactive sources to obtain a temperature correction coefficient in order to stabilize the influence of temperature change on the detector at different time intervals. This was used to study the correlation between the daily and annual dose rates of low gamma background radiation. The results showed that there was a shift in the spectrum of the daily and annual dose rates calculated using the algorithm obtained when the detector was calibrated with a highly radioactive source to the position of the constant coefficient for low-level dose. However, the ones obtained when the detector was calibrated with a low radioactive source and that of the constant coefficient for a low-level dose overlapped each other. This demonstrated that the type of radioactive source used in detector calibration during manufacture has an effect on the temperature correction coefficient, which in turn has an effect on the accuracy of the ambient dose rate when used to monitor gamma radiation. The duration of the time interval for measurement was found to be very important since it has an effect on dose rate readings.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Hybrid analytical method for calibrating a standard NaI(Tl) gamma-ray scintillation detector using a lateral hexagonal radioactive source
    Badawi, Mohamed S.
    Hamzawy, A.
    Thabet, Abouzeid A.
    AIP ADVANCES, 2022, 12 (06)
  • [2] GAMMA SPECTRUM STABILIZATION FOR ENVIRONMENTAL RADIATION MONITORING STATIONS USING NAI(TL) DETECTOR
    Dinh Tien Hung
    Cao Van Hiep
    Pham Dinh Khang
    Nguyen Xuan Hai
    Nguyen Ngoc Anh
    Dinh, D. A.
    Nhan Hao, T., V
    Pham, V
    RADIATION PROTECTION DOSIMETRY, 2020, 189 (01) : 48 - 55
  • [3] Calibration of cylindrical NaI(Tl) gamma-ray detector intended for truncated conical radioactive source
    Badawi, Mohamed S.
    Thabet, Abouzeid A.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2022, 54 (04) : 1421 - 1430
  • [4] THE RESOLVING CAPABILITY OF THE NAI(TL) SCINTILLATION DETECTOR
    MATHIESON, AML
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1982, 15 (FEB) : 98 - 99
  • [5] Inspection of radioactive nuclei of foodstuffs in Saudi Arabia using a NaI(tl) scintillation detector
    EL-Araby, Entesar H.
    Shabaan, Doaa H.
    COGENT FOOD & AGRICULTURE, 2024, 10 (01):
  • [6] APPLICATION OF NAI(TL) SCINTILLATION DETECTOR TO MEASUREMENT OF TRITIUM CONCENTRATION
    NISHIZAWA, K
    ENDO, Y
    SHINAGAWA, M
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 1977, 14 (10) : 705 - 709
  • [7] A NaI(Tl)-based radioactive noble-gas monitoring system used for radiation monitoring
    Li, HuiBin
    Lei, WeiGuo
    Feng, TianCheng
    Huang, DingWei
    Tian, ZiNing
    APPLIED RADIATION AND ISOTOPES, 2020, 163
  • [8] Experimental response function of NaI(Tl) scintillation detector for gamma photons and tomographic measurements for defect detection
    Sharma, Amandeep
    Singh, Karamjit
    Singh, Bhajan
    Sandhu, B. S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2011, 269 (03): : 247 - 256
  • [9] Measurement of ambient dose equivalent using a NaI(Tl) scintillation detector
    Yi, CY
    Jun, JS
    Chai, HS
    Oh, JJ
    Yun, JY
    RADIATION PROTECTION DOSIMETRY, 1997, 74 (04) : 273 - 278
  • [10] Comparative study for intermediate crystal size of NaI(Tl) scintillation detector
    Singh, Inderjeet
    Singh, Bhajan
    Sandhu, B. S.
    Sabharwal, Arvind D.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (07):