Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures

被引:65
|
作者
Nowotarski, Peter J. [1 ]
Ervin, Bain [1 ]
Weatherby, Brian [2 ,6 ]
Pettit, Jonathan [3 ]
Goulet, Ron [4 ]
Norris, Brent [5 ]
机构
[1] Univ Tennessee, Coll Med Chattanooga, Dept Orthopaed, Chattanooga, TN 37403 USA
[2] Univ Texas Hlth Sci Ctr, Coll Med Houston, Dept Orthopaed, Houston, TX 77025 USA
[3] Middle Tennessee Bone & Joint, Columbia, TN 38401 USA
[4] Univ Tennessee, Coll Engn & Comp Sci, EMCS, Dept 2452, Chattanooga, TN 37403 USA
[5] Orthopaed Trauma Serv Oklahoma, Tulsa, OK 74114 USA
[6] Foot & Ankle Steadman Hawkins Clin Carolinas, Greenville, SC 29615 USA
关键词
Hip; Plate; Biomechanical; Locking; Cyclic; FIXATION; FEMUR; AGES;
D O I
10.1016/j.injury.2011.09.012
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 1358 AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm (1), respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm (1). The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:802 / 806
页数:5
相关论文
共 50 条
  • [41] ENDOPROSTHESES IN TREATMENT OF FEMORAL NECK FRACTURES
    SALVATI, EA
    ARTZ, T
    AGLIETTI, P
    ASNIS, SE
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1974, 5 (04) : 757 - 777
  • [42] TREATMENT OF FEMORAL NECK FRACTURES BY COMPRESSION
    CHARNLEY, J
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1959, 41 (01): : 212 - 213
  • [43] Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures
    Jichao Liu
    Zhengwei Li
    Jie Ding
    Bingzhe Huang
    Chengdong Piao
    BMC Musculoskeletal Disorders, 23
  • [44] Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures
    Liu, Jichao
    Li, Zhengwei
    Ding, Jie
    Huang, Bingzhe
    Piao, Chengdong
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [45] Biomechanical analysis of a new cannulated screw for unstable femoral neck fractures
    Chen, Zhigang
    Chen, Feiyang
    Xu, Xinbao
    Li, Xin
    Cui, Haidong
    Zhang, Wen
    Jiang, Dong
    Zhang, Feixiang
    Chen, Yinbing
    Zhou, Shiran
    Lyu, Shujun
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [46] Finite element analysis of a new plate for Pauwels type III femoral neck fractures
    Wang, Gang
    Tang, Yong
    Wu, Xuhua
    Yang, Huilin
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2020, 48 (02)
  • [47] Treatment of Pauwels type III femoral neck fracture with medial femoral neck support screw: a biomechanical and clinical study
    Zhichao Gao
    Mei Wang
    Baojie Shen
    Xiaodong Chu
    Di Ruan
    Scientific Reports, 11
  • [48] Treatment of Pauwels type III femoral neck fracture with medial femoral neck support screw: a biomechanical and clinical study
    Gao, Zhichao
    Wang, Mei
    Shen, Baojie
    Chu, Xiaodong
    Ruan, Di
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [49] Biomechanical comparison of femoral neck anti-rotation and support system versus femoral neck system for unstable pauwels III femoral neck fractures
    Wang, Taiyou
    Wang, Guangjian
    Zhu, Fukang
    Qiao, Bo
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01):
  • [50] Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis
    Shaolong Huang
    Yazhong Zhang
    Xu Zhang
    Chengqiang Zhou
    Wenbo Li
    Yunqing Wang
    Bin Wang
    Ziqiang Zhu
    BioMedical Engineering OnLine, 22