Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures

被引:65
|
作者
Nowotarski, Peter J. [1 ]
Ervin, Bain [1 ]
Weatherby, Brian [2 ,6 ]
Pettit, Jonathan [3 ]
Goulet, Ron [4 ]
Norris, Brent [5 ]
机构
[1] Univ Tennessee, Coll Med Chattanooga, Dept Orthopaed, Chattanooga, TN 37403 USA
[2] Univ Texas Hlth Sci Ctr, Coll Med Houston, Dept Orthopaed, Houston, TX 77025 USA
[3] Middle Tennessee Bone & Joint, Columbia, TN 38401 USA
[4] Univ Tennessee, Coll Engn & Comp Sci, EMCS, Dept 2452, Chattanooga, TN 37403 USA
[5] Orthopaed Trauma Serv Oklahoma, Tulsa, OK 74114 USA
[6] Foot & Ankle Steadman Hawkins Clin Carolinas, Greenville, SC 29615 USA
关键词
Hip; Plate; Biomechanical; Locking; Cyclic; FIXATION; FEMUR; AGES;
D O I
10.1016/j.injury.2011.09.012
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 1358 AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm (1), respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm (1). The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods. (C) 2011 Elsevier Ltd. All rights reserved.