Diagnostic reference levels (DRLs) for routine X-ray examinations in Lorestan province, Iran

被引:0
|
作者
Gholami, M. [1 ]
Maziar, A. [2 ]
Khosravi, H. R. [3 ]
Ebrahimzadeh, F. [4 ]
Mayahi, S. [1 ]
机构
[1] Lorestan Univ Med Sci, Fac Allied Med Sci, Dept Med Phys & Radiat Sci, Khorramabad, Iran
[2] Isfahan Univ Med Sci, Fac Med, Dept Med Phys & Med Engn, Esfahan, Iran
[3] Atom Energy Org Iran, Tehran, Iran
[4] Lorestan Univ Med Sci, Fac Hlth & Nutr, Dept Biostat, Khorramabad, Iran
来源
关键词
Diagnostic radiology; ESAK; patient dose; Lorestan province; quality control; DOSE MEASUREMENTS; ENTRANCE SURFACE; EXPOSURES; BRAZIL; CHEST; UK;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: In diagnostic radiology there are two reasons for measuring or estimating radiation doses to patients. Firstly measurements provide a means for setting and checking standards of good practice as an aid to the optimization of patient protection. Secondly estimates of the absorbed dose to tissue and organs in the patients. Material and Methods: A total of 2382 patients were studied to calculate the Entrance Surface Air Kerma (ESAK) following seven radiographic examinations including: chest (PA, Lat), lumbar spines (AP, Lat), pelvis (AP), abdomen (AP), skull (PA, Lat), thoracic spine (AP, Lat) and cervical spines (AP, Lat). The ESAKs values were measured according to x- ray tube output, optimized exposure parameters and body thickness (tp) for each technique. Results: The parameters such as, 1st quartile, mean, median, 3rd quartile, minimum, maximum and standard deviation of each ESAK values are reported and compared to NRPB guide levels. The results showed that the ESAKs values in the lumbar spines and chest X-ray examinations were 30% above the guide levels. However, for the pelvis (AP), skull (PA) and abdomen (AP) examinations, these values were below than those reported by the NRPB. Conclusion: Periodic quality control and monitoring the technical performance of radiographers might effectively improve the image quality and eventually reducing the dose received by patients.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [41] Establishing national diagnostic reference levels (DRLs) for computed tomography in Egypt
    Salama, Dina Husseiny
    Vassileva, Jenia
    Mahdaly, Gamal
    Shawki, Mona
    Salama, Ahmad
    Gilley, Debbie
    Rehani, Madan Mohan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 39 : 16 - 24
  • [42] Estimation of entrance surface air kerma due to diagnostic X-ray examinations of adult patients in Uttarakhand, India and establishment of local diagnostic reference levels
    Uniyal, Satish C.
    Chaturvedi, Vineet
    Sharma, Sunil D.
    Raghuvanshi, Shailendra
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (03) : 687 - 694
  • [43] Estimation of entrance surface air kerma due to diagnostic X-ray examinations of adult patients in Uttarakhand, India and establishment of local diagnostic reference levels
    Satish C. Uniyal
    Vineet Chaturvedi
    Sunil D. Sharma
    Shailendra Raghuvanshi
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 687 - 694
  • [44] The Establishment of Institutional Diagnostic Reference Levels (DRLs) in the Cipto Mangunkusumo Hospital
    Amalia, T.
    Zulkarnaien, B.
    Anam, C.
    Nurcahyo, K.
    Tussyadiah, H.
    Pradana, D. E.
    ATOM INDONESIA, 2022, 48 (02) : 159 - 167
  • [45] New diagnostic Reference Values for X-Ray
    不详
    RADIOLOGIE, 2023, 63 (03): : 222 - 222
  • [46] Establishment of CT diagnostic reference levels (DRLs) for a Singapore healthcare cluster
    Arlany, L.
    Toh, H. G.
    Nazir, B.
    Ng, J. J.
    Tay, Y. H.
    Tay, Y. S.
    Poon, C. J. X.
    Lee, L. N.
    Fum, W. K. S.
    Lee, E. Y. T.
    Mariah, S., I
    Fortier, M., V
    Tham, W. P.
    Chew, L. L.
    Chong, K. H. E.
    Chong, L. R.
    RADIOGRAPHY, 2023, 29 (01) : 184 - 189
  • [47] DIAGNOSTIC REFERNCE LEVELS AND PATIENT DOSES FOR ADULT PATIENTS IN CONVENTIONAL X-RAY EXAMINATIONS IN LATVIA
    Katlapa, Agnese
    Zalcmanis, Emils
    MEDICAL PHYSICS IN THE BALTIC STATES, 2019, : 31 - 33
  • [48] The design and construct a website for collection and report diagnostic reference levels (DRLs) in diagnostic radiography
    Chaychid, Narongsak
    Bunta, Kaiwit
    Awikunprasert, Panatsada
    Sangsuwan, Kamonwan
    Wuttanatum, Chaleomchok
    Duksukkeaw, Natharin
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2025, 56 (01)
  • [49] DOSES TO PATIENTS FROM ROUTINE DIAGNOSTIC-X-RAY EXAMINATIONS IN ENGLAND
    SHRIMPTON, PC
    WALL, BF
    JONES, DG
    FISHER, ES
    HILLIER, MC
    KENDALL, GM
    HARRISON, IRM
    BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (704): : 749 - 758
  • [50] Lung Resection: How useful are routine postoperative X-ray Examinations?
    Lorenz, Judith