Diagnostic reference levels (DRLs) for routine X-ray examinations in Lorestan province, Iran

被引:0
|
作者
Gholami, M. [1 ]
Maziar, A. [2 ]
Khosravi, H. R. [3 ]
Ebrahimzadeh, F. [4 ]
Mayahi, S. [1 ]
机构
[1] Lorestan Univ Med Sci, Fac Allied Med Sci, Dept Med Phys & Radiat Sci, Khorramabad, Iran
[2] Isfahan Univ Med Sci, Fac Med, Dept Med Phys & Med Engn, Esfahan, Iran
[3] Atom Energy Org Iran, Tehran, Iran
[4] Lorestan Univ Med Sci, Fac Hlth & Nutr, Dept Biostat, Khorramabad, Iran
来源
关键词
Diagnostic radiology; ESAK; patient dose; Lorestan province; quality control; DOSE MEASUREMENTS; ENTRANCE SURFACE; EXPOSURES; BRAZIL; CHEST; UK;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: In diagnostic radiology there are two reasons for measuring or estimating radiation doses to patients. Firstly measurements provide a means for setting and checking standards of good practice as an aid to the optimization of patient protection. Secondly estimates of the absorbed dose to tissue and organs in the patients. Material and Methods: A total of 2382 patients were studied to calculate the Entrance Surface Air Kerma (ESAK) following seven radiographic examinations including: chest (PA, Lat), lumbar spines (AP, Lat), pelvis (AP), abdomen (AP), skull (PA, Lat), thoracic spine (AP, Lat) and cervical spines (AP, Lat). The ESAKs values were measured according to x- ray tube output, optimized exposure parameters and body thickness (tp) for each technique. Results: The parameters such as, 1st quartile, mean, median, 3rd quartile, minimum, maximum and standard deviation of each ESAK values are reported and compared to NRPB guide levels. The results showed that the ESAKs values in the lumbar spines and chest X-ray examinations were 30% above the guide levels. However, for the pelvis (AP), skull (PA) and abdomen (AP) examinations, these values were below than those reported by the NRPB. Conclusion: Periodic quality control and monitoring the technical performance of radiographers might effectively improve the image quality and eventually reducing the dose received by patients.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [31] ROUTINE CALIBRATION OF DIAGNOSTIC X-RAY MACHINES
    BULL, SM
    BURLIN, TE
    BRITISH JOURNAL OF RADIOLOGY, 1962, 35 (419): : 796 - 797
  • [32] Estimation of adult patient doses for common diagnostic X-ray examinations in Wad-madani, Sudan: derivation of local diagnostic reference levels
    I. I. Suliman
    T. S. Mohammedzein
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 425 - 429
  • [33] Estimation of adult patient doses for common diagnostic X-ray examinations in Wad-madani, Sudan: derivation of local diagnostic reference levels
    Suliman, I. I.
    Mohammedzein, T. S.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 425 - 429
  • [34] Radiation dose measurement for patients undergoing common spine medical x-ray examinations and proposed local diagnostic reference levels
    Rasuli, Behrouz
    Ghorbani, Mohammad
    Juybari, Raheleh Tabari
    RADIATION MEASUREMENTS, 2016, 87 : 29 - 34
  • [35] Towards the establishment of national diagnostic reference levels for chest x-ray examinations in Sri Lanka: a multi-centric study
    Welarathna, Sachith
    Velautham, Sivakumar
    Sarasanandarajah, Sivananthan
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (04)
  • [36] PATIENT DOSES IN COMMON DIAGNOSTIC X-RAY EXAMINATIONS
    Metaxas, Vasileios, I
    Messaris, Gerasimos A.
    Lekatou, Aristea N.
    Petsas, Theodore G.
    Panayiotakis, George S.
    RADIATION PROTECTION DOSIMETRY, 2019, 184 (01) : 12 - 27
  • [37] Diagnostic reference levels for routine computed tomography angiography examinations: a systematic review
    Alhailiy, Ali Bahny
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2024, 55 (01):
  • [38] Routine Diagnostic X-ray Examinations and Increased Frequency of Chromosome Translocations among US Radiologic Technologists
    Sigurdson, Alice J.
    Bhatti, Parveen
    Preston, Dale L.
    Doody, Michele Morin
    Karnpa, Diane
    Alexander, Bruce H.
    Petibone, Dayton
    Yong, Lee C.
    Edwards, Alan A.
    Ron, Elaine
    Tucker, James D.
    CANCER RESEARCH, 2008, 68 (21) : 8825 - 8831
  • [39] Diagnostic reference levels for X-ray investigations in pain management units in Belgium
    Lichtherte, S.
    de Waroux, B. le Polain
    RADIATION PROTECTION DOSIMETRY, 2012, 150 (04) : 520 - 522
  • [40] Management of patient doses and diagnostic reference levels in X-ray radiography in Lithuania
    Ziliukas, J.
    Krynke, L.
    Urboniene, A.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 313 - 316