Application of Dual-Energy Spectral Computed Tomography in Bone Mineral Density Measurement: Phantom and Clinical Research

被引:6
|
作者
Wang, Mingyue [1 ]
Wu, Yan [1 ]
Zhou, Yue [1 ]
Dong, Junqiang [1 ]
Hu, Shenshen [1 ]
Hou, Ping [1 ]
Gao, Jianbo [1 ]
机构
[1] Zhengzhou Univ, Affiliated Hosp 1, Dept Radiol, Zhengzhou, Peoples R China
关键词
dual-energy spectral computed tomography; quantitative computed tomography; bone mineral density; European spine phantom; spine; MATERIAL DECOMPOSITION; TRABECULAR BONE; CT; OSTEOPOROSIS; RECONSTRUCTION; ACCURACY; SPINE; FAT;
D O I
10.2147/IJGM.S381857
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Early detection and treatment of osteoporosis through bone mineral density (BMD) measurement could aid in the prevention of osteoporosis-related fractures. We aimed to assess the parameter of dual-energy spectral CT (DesCT) consistency with BMD determination using quantitative computed tomography (QCT), thereby establishing a basis for further DesCT application for BMD determination. Methods: We subjected the European spine phantom, which contains three vertebral bodies (V1, V2, and V3), to DesCT with different radiation doses. The basis material pairs were hydroxyapatite (water), calcium (water), and hydroxyapatite (fat). Additionally, the medical records of 152 patients who underwent QCT and DesCT for chest scans in a two-month period were reviewed to measure BMD values. Results: No significant differences were found in the basis pair values of the V1, V2, or V3 vertebrae under different radiation doses in the phantom; in particular, the hydroxyapatite (water), hydroxyapatite (fat), relative error values of V1, V2, and V3 under different radiation doses were not significantly different (all p > 0.05). For patients, the hydroxyapatite (water), hydroxyapatite (fat), and hydroxyapatite (average) values measured by DesCT had a significant correlation with BMD measured by QCT Among 242 vertebrae (152 T12 and 90 L1 vertebrae), there was no significant difference between the BMD measured by QCT and the HAP (average) measured by DesCT (p = 0.071). The interclass correlation coefficient (ICC) value was 0.925 between the HAP (average) and HAP (average) with DesCT and BMD measured by QCT (p < 0.001). Bland-Altman diagram indicated that both measurements were in good agreement. Discussion: We showed that BMD values measured by DesCT were stable and repeatable under different radiation doses. DesCT and QCT measurements of human BMD were highly correlated. Thus, DesCT-based BMD assessment of the spine in a clinical setting could be considered feasible.
引用
收藏
页码:6887 / 6896
页数:10
相关论文
共 50 条
  • [41] Improving the Image Quality in Computed Tomographic Pulmonary Angiography With Dual-Energy Subtraction: A New Application of Spectral Computed Tomography
    Bai, Aiguo
    Sun, Yibo
    Qi, Lin
    Yang, Yanli
    Hua, Yanqing
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (05) : 718 - 724
  • [42] Radiomics nomogram based on dual-energy spectral CT imaging to diagnose low bone mineral density
    Qianqian Yao
    Mengke Liu
    Kemei Yuan
    Yue Xin
    Xiaoqian Qiu
    Xiuzhu Zheng
    Changqin Li
    Shaofeng Duan
    Jian Qin
    BMC Musculoskeletal Disorders, 23
  • [43] Comparison of Iodine Density Measurement Among Dual-Energy Computed Tomography Scanners From 3 Vendors
    Kim, Hyungjin
    Goo, Jin Mo
    Kang, Chang Kyung
    Chae, Kum Ju
    Park, Chang Min
    INVESTIGATIVE RADIOLOGY, 2018, 53 (06) : 321 - 327
  • [44] Absolute Measurement of Effective Atomic Number and Electron Density using Dual-Energy Computed Tomography images
    Kim, Dae-Hong
    Kim, Hee-Joung
    Lee, Chang-Lae
    Cho, Hyo-Min
    Park, Hye-Suk
    Lee, Seung-Wan
    Choi, Yu-Na
    Kim, Ye-Seul
    Park, Su-Jin
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [45] Dual-Energy Computed Tomography Dose Reduction, Series Reduction, and Contrast Load Reduction in Dual-Energy Computed Tomography
    Parakh, Anushri
    Macri, Francesco
    Sahani, Dushyant
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2018, 56 (04) : 601 - +
  • [46] Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model
    James D. Koonce
    Rozemarijn Vliegenthart
    U. Joseph Schoepf
    Bernhard Schmidt
    Amy E. Wahlquist
    Paul J. Nietert
    Gorka Bastarrika
    Thomas G. Flohr
    Felix G. Meinel
    European Radiology, 2014, 24 : 512 - 518
  • [47] Dual-Energy Computed Tomography for Stone Type Assessment: A Pilot Study of Dual-Energy Computed Tomography with Five Indices
    Rudenko, Vadim
    Serova, Natalia
    Kapanadze, Lida
    Taratkin, Mark
    Okhunov, Zhamshid
    Leonard, Stephen P.
    Ritter, Manuel
    Kriegmair, Maximilian
    Snurnitsyna, Olesya
    Kozlov, Vasiliy
    Laukhtina, Ekaterina
    Arshiev, Magomed
    Aleksandrova, Ksenia
    Salomon, Georg
    Enikeev, Dmitry
    Glybochko, Petr
    JOURNAL OF ENDOUROLOGY, 2020, 34 (09) : 893 - 899
  • [48] Accuracy of dual-energy computed tomography for the measurement of iodine concentration using cardiac CT protocols: validation in a phantom model
    Koonce, James D.
    Vliegenthart, Rozemarijn
    Schoepf, U. Joseph
    Schmidt, Bernhard
    Wahlquist, Amy E.
    Nietert, Paul J.
    Bastarrika, Gorka
    Flohr, Thomas G.
    Meinel, Felix G.
    EUROPEAN RADIOLOGY, 2014, 24 (02) : 512 - 518
  • [49] Image Enhancement by Spectral-Error Correction for Dual-Energy Computed Tomography
    Park, Kyung-Kook
    Oh, Chang-Hyun
    Akay, Metin
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 8491 - 8494
  • [50] Dual-Energy, Spectral and Photon Counting Computed Tomography for Evaluation of the Gastrointestinal Tract
    Nehra, Avinash K.
    Dane, Bari
    Yeh, Benjamin M.
    Fletcher, Joel G.
    Leng, Shuai
    Mileto, Achille
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 1031 - 1049