Detection of Respiratory Phases to Estimate Breathing Pattern Parameters using Wearable Bioimpendace

被引:3
|
作者
Blanco-Almazan, Dolores [1 ,2 ,3 ]
Groenendaal, Willemijn [4 ]
Catthoor, Francky [5 ]
Jane, Raimon [1 ,2 ,3 ]
机构
[1] Inst Bioengn Catalonia IBEC BIST, Barcelona 08028, Spain
[2] Univ Politecn Cataluny, Dept Automat Control, ESAII, Barcelona 08034, Spain
[3] Biomed Res Network Ctr Bioengn Biomat & Nanomed C, Barcelona 08028, Spain
[4] Netherlands Holst Ctr, NL-5656 AE Eindhoven, Netherlands
[5] Katholieke Univ Leuven, B-3001 Leuven, Belgium
关键词
BIOIMPEDANCE; ALGORITHMS; SIGNALS; COPD;
D O I
10.1109/EMBC46164.2021.9630811
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many studies have focused on novel noninvasive techniques to monitor respiratory rate such as bioimpedance. We propose an algorithm to detect respiratory phases using wearable bioimpedance to compute time parameters like respiratory rate, inspiratory and expiratory times, and duty cycle. The proposed algorithm was compared with two other algorithms from literature designed to estimate the respiratory rate using physiological signals like bioimpedance. We acquired bioimpedance and airflow from 50 chronic obstructive pulmonary disease (COPD) patients during an inspiratory loading protocol. We compared performance of the algorithms by computing accuracy and mean average percentage error (MAPE) between the bioimpedance parameters and the reference parameters from airflow. We found similar performance for the three algorithms in terms of accuracy (>0.96) and respiratory time and rate errors (<3.42 %). However, the proposed algorithm showed lower MAPE in duty cycle (10.18 %), inspiratory time (10.65 %) and expiratory time (8.61 %). Furthermore, only the proposed algorithm kept the statistical differences in duty cycle between COPD severity levels that were observed using airflow. Accordingly, we suggest bioimpedance to monitor breathing pattern parameters in home situations. Clinical relevance- This study exhibits the suitability of wearable thoracic bioimpedance to detect respiratory phases and to compute accurate breathing pattern parameters.
引用
收藏
页码:5508 / 5511
页数:4
相关论文
共 50 条
  • [31] Synchronized moving aperture radiation therapy (SMART): improvement of breathing pattern reproducibility using respiratory coaching
    Neicu, T
    Berbeco, R
    Wolfgang, J
    Jiang, SB
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (03): : 617 - 636
  • [32] EFFECTS OF DIFFERENTIAL BREATHING TIME RATIOS ON HEMODYNAMIC AND RESPIRATORY PARAMETERS USING CONTROLLED ARTIFICIAL-VENTILATION
    SCHUSTER, S
    WEILEMANN, LS
    KELBEL, C
    SCHINZEL, H
    MEYER, J
    SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 1987, 117 (12) : 455 - 455
  • [33] Respiratory Event Detection During Sleep Using Electrocardiogram and Respiratory Related Signals: Using Polysomnogram and Patch-Type Wearable Device Data
    Yeo, Minsoo
    Byun, Hoonsuk
    Lee, Jiyeon
    Byun, Jungick
    Rhee, Hak Young
    Shin, Wonchul
    Yoon, Heenam
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2022, 26 (02) : 550 - 560
  • [34] Algorithms to Estimate PaCO2 and pH Using Noninvasive Parameters for Children with Hypoxemic Respiratory Failure
    Khemani, Robinder G.
    Celikkaya, E. Busra
    Shelton, Christian R.
    Kale, Dave
    Ross, Patrick A.
    Wetzel, Randall C.
    Newth, Christopher J. L.
    RESPIRATORY CARE, 2014, 59 (08) : 1248 - 1257
  • [35] SLEEP STAGE PREDICTION AND SLEEP DISORDERED BREATHING DETECTION USING RAW ACTIGRAPHY AND PHOTOPLETHYSMOGRAPHY FROM WEARABLE CONSUMER DEVICE
    Olsen, M.
    Sorensen, H.
    Jennum, P.
    Mignot, E.
    SLEEP, 2020, 43 : A461 - A462
  • [36] An Accelerometer-Based Wearable Patch for Robust Respiratory Rate and Wheeze Detection Using Deep Learning
    Sang, Brian
    Wen, Haoran
    Junek, Gregory
    Neveu, Wendy
    Di Francesco, Lorenzo
    Ayazi, Farrokh
    Biosensors, 2024, 14 (03):
  • [37] An Accelerometer-Based Wearable Patch for Robust Respiratory Rate and Wheeze Detection Using Deep Learning
    Sang, Brian
    Wen, Haoran
    Junek, Gregory
    Neveu, Wendy
    Di Francesco, Lorenzo
    Ayazi, Farrokh
    BIOSENSORS-BASEL, 2024, 14 (03):
  • [38] Detection and classification of sleep-disordered breathing using acoustic respiratory input impedance and nasal pressure
    Steltner, H
    Staats, R
    Vogel, M
    Virchow, C
    Matthys, H
    Guttmann, J
    Timmer, J
    MEDICAL DATA ANALYSIS, PROCEEDINGS, 2000, 1933 : 61 - 66
  • [39] Using Wearable ECG/PPG Sensors for Driver Drowsiness Detection Based on Distinguishable Pattern of Recurrence Plots
    Lee, Hyeonjeong
    Lee, Jaewon
    Shin, Miyoung
    ELECTRONICS, 2019, 8 (02)
  • [40] Changes of Emphysema Parameters over the Respiratory Cycle During Free Breathing: Preliminary Results Using Respiratory Gated 4D-CT
    Ley-Zaporozhan, Julia
    Ley, Sebastian
    Mews, Jurgen
    Weinheimer, Oliver
    Kandel, Sonja
    Rogalla, Patrik
    COPD-JOURNAL OF CHRONIC OBSTRUCTIVE PULMONARY DISEASE, 2017, 14 (06) : 597 - 602