Object. This study was conducted to assess the in vivo safety and accuracy of percutaneous lumbar pedicle screw placement using the owl's-eye view of the pedicle axis and a new guidance technology system that facilitates orientation of the C-arm into the appropriate fluoroscopic view and the pedicle cannulation tool in the corresponding trajectory. Methods. A total of 326 percutaneous pedicle screws were placed from L-3 to S-1 in 85 consecutive adult patients. Placement was performed using simple coaxial imaging of the pedicle with the owl's-eye fluoroscopic view. Neuro Vision, a new guidance system using accelerometer technology, helped align the C-arm trajectory into the owl's-eye view and the cannulation tool in the same trajectory. Postoperative fine-cut CT scans were acquired to assess screw position. Medical records were reviewed for complications. Results. Five of 326 screws breached a pedicle cortex all breaches were less than 2 mm for an accuracy rate of 98.47%. Five screws violated an adjacent facet joint. All were at the S-1 superior facet and included in a fusion. No screw violated an adjacent mobile facet or disc space. There were no cases of new or worsening neurological symptoms or deficits for an overall clinical accuracy of 100%. Conclusions. The owl's-eye technique of coaxial pedicle imaging with the C-arm fluoroscopy, facilitated by Neuro Vision, is a safe and accurate means by which to place percutaneous pedicle screws for degenerative conditions of the lumbar spine. This is the largest series reported to use the oblique or owl's-eye projection for percutaneous pedicle screw insertion. The accuracy of percutaneous screw insertion with this technique meets or exceeds that of other reported clinical series or techniques. (DOI: 10.3171/2010.4.SPINE09580)
引用
收藏
页码:509 / 515
页数:7
相关论文
共 50 条
  • [21] Accuracy and safety of percutaneous pedicle screw placement using the K-wireless minimally invasive spine percutaneous pedicle screw system in Japan: A randomized active controlled study
    Ohmori, Kazuo
    Terayama, Sei
    Ono, Koichiro
    Sakamoto, Miyuki
    Horikoshi, Yukie
    NORTH AMERICAN SPINE SOCIETY JOURNAL, 2022, 10
  • [22] Assessment of Pedicle Screw Placement Accuracy, Procedure Time, and Radiation Exposure Using a Miniature Robotic Guidance System
    Lieberman, Isador H.
    Hardenbrook, Mitchell A.
    Wang, Jeffrey C.
    Guyer, Richard D.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (05): : 241 - 248
  • [23] True accuracy of percutaneous pedicle screw placement in thoracic and lumbar spinal fixation with a CT-based navigation system: Intraoperative and postoperative assessment of 763 percutaneous pedicle screws
    Ouchida, Jun
    Kanemura, Tokumi
    Satake, Kotaro
    Nakashima, Hiroaki
    Segi, Naoki
    Suzuki, Kanae
    Imagama, Shiro
    JOURNAL OF CLINICAL NEUROSCIENCE, 2020, 79 : 1 - 6
  • [24] A Retrospective Analysis of Pedicle Screw Placement Accuracy Using the ExcelsiusGPS Robotic Guidance System: Case Series COMMENTS
    Villeneuve, Lance M.
    Smith, Zachary A.
    OPERATIVE NEUROSURGERY, 2023, 24 (03) : 247 - 247
  • [25] Accuracy of Pedicle Screw Placement Using Intraoperative CT-Guided Navigation and Conventional Fluoroscopy for Lumbar Spondylosis
    Alqurashi, Ashwaq
    Alomar, Soha A.
    Bakhaidar, Mohamad
    Alfiky, Mohammed
    Baeesa, Saleh S.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (08)
  • [26] A Novel Free-hand Technique of Pedicle Screw Placement in the Lumbar Spine: Accuracy Evaluation and Preliminary Clinical Results
    Jiang, Weiqi
    Leng, Ao
    Meng, Lingzhi
    Long, Zuoyao
    Long, Yu
    Wang, Qi
    ORTHOPAEDIC SURGERY, 2023, 15 (09) : 2260 - 2266
  • [27] Accuracy Assessment of Percutaneous Pedicle Screw Placement Using Cone Beam Computed Tomography with Metal Artifact Reduction
    Charles, Yann Philippe
    Al Ansari, Rawan
    Collinet, Arnaud
    De Marini, Pierre
    Schwartz, Jean
    Nachabe, Rami
    Schafer, Dirk
    Brendel, Bernhard
    Gangi, Afshin
    Cazzato, Roberto Luigi
    SENSORS, 2022, 22 (12)
  • [28] Comparison of Cervical Pedicle Screw Placement Accuracy With Robotic Guidance System Versus Image Guidance System Using Propensity Score Matching
    Yamamoto, Yuki
    Fujishiro, Takashi
    Hirai, Hiromichi
    Hayama, Sachio
    Nakaya, Yoshiharu
    Usami, Yoshitada
    Neo, Masashi
    CLINICAL SPINE SURGERY, 2024, 37 (10): : E424 - E432
  • [29] Minimally Invasive Transforaminal Lumbar Interbody Fusion Using Augmented Reality Surgical Navigation for Percutaneous Pedicle Screw Placement
    Charles, Yann P.
    Cazzato, Roberto L.
    Nachabe, Rami
    Chatterjea, Anindita
    Steib, Jean-Paul
    Gangi, Afshin
    CLINICAL SPINE SURGERY, 2021, 34 (07): : E415 - E424
  • [30] Clinical Assessment of Percutaneous Lumbar Pedicle Screw Placement Using the O-Arm Multidimensional Surgical Imaging System
    Houten, John K.
    Nasser, Rani
    Baxi, Nrupen
    NEUROSURGERY, 2012, 70 (04) : 990 - 995