3D transcranial ultrasound as a novel intra-operative imaging technique for DBS surgery: a feasibility study

被引:8
|
作者
Ahmadi, Seyed-Ahmad [1 ]
Milletari, Fausto [2 ]
Navab, Nassir [2 ]
Schuberth, Madeleine [1 ]
Plate, Annika [1 ]
Boetzel, Kai [1 ]
机构
[1] LMU Klinikum Grosshadern, Dept Neurol, Munich, Germany
[2] Tech Univ Munich, CAMP, D-80290 Munich, Germany
关键词
New intra-operative sensors; Ultrasound; Magnetic resonance imaging; Neurosurgery; Deep brain stimulation; DEEP BRAIN-STIMULATION; GLOBUS-PALLIDUS INTERNUS; SUBSTANTIA-NIGRA; REGISTRATION; NEUROSURGERY; PLACEMENT; DIAGNOSIS; DISEASE; RISK;
D O I
10.1007/s11548-015-1191-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-operative image guidance during deep brain stimulation (DBS) surgery is usually avoided due to cost and overhead of intra-operative MRI and CT acquisitions. Recently, there has been interest in the community towards the usage of non-invasive transcranial ultrasound (TCUS) through the preauricular bone window. In this work, we investigate, for the first time, the feasibility of using 3D-TCUS for imaging of already implanted DBS electrodes. As a first step towards this goal, we report imaging methods and electrode localisation errors outside of the operating room on eight previously operated DBS patients. We evaluate the feasibility of using 3D-TCUS by registering volumes to pre-operative T1-MRI. US-MRI registration is achieved through a two-step point-based approach. First, a rough surface scan of the subjects' skin surface in 3D-TCUS space is registered to a segmented skin-surface point cloud from MRI. Next, we perform a refinement using rigid registration of multiple pairs of manually marked anatomical landmarks. We validate against post-operative CT scans which are also registered to pre-operative MRI. Qualitative results are given in form of 3D reconstruction examples at 2.5 and 3.5 MHz TCUS image frequency, overlaid on pre-operative T1-MRI and post-operative CT. Quantitative evaluation is performed by reporting the accuracy of electrode tip localisation at 2.5 and 3.5 MHz after our US-MRI approach. As a baseline, we also report RMSE errors for pairs of anatomical landmarks in pre-operative MRI and 3D-TCUS. Multiple image examples show the appearance and quality of 3D-TCUS scans, depending on the bone window. Overall accuracy of anatomic point pairs lies on the order of 3.2 mm, using our registration approach. Compared to this baseline, electrode tip localisation in 3D-TCUS has a mean accuracy on the order of 4.8 mm and a precision on the order of 2.3 mm. While insufficient at first glance, we argue why these results are promising nonetheless. Our work motivates further future work in improved TCUS scanning, advanced TCUS-MRI registration and computer-aided electrode detection in 3D-TCUS.
引用
收藏
页码:891 / 900
页数:10
相关论文
共 50 条
  • [31] Towards intra-operative 3D nuclear imaging: Reconstruction of 3D radioactive distributions using tracked gamma probes
    Wendler, Thomas
    Hartl, Alexander
    Lasser, Tobias
    Traub, Joerg
    Daghighian, Farhad
    Ziegler, Sibylle I.
    Navab, Nassir
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION- MICCAI 2007, PT 2, PROCEEDINGS, 2007, 4792 : 909 - +
  • [32] Towards Intra-operative Monitoring of Ablation Using Tracked 3D Ultrasound Elastography and Internal Palpation
    Foroughi, Pezhman
    Burgner, Jessica
    Choti, Michael A.
    Webster, Robert J., III
    Hager, Gregory D.
    Boctor, Emad M.
    MEDICAL IMAGING 2012: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY, 2012, 8320
  • [33] Feasibility of intra-operative image guidance in burn excision surgery with multispectral imaging and deep learning
    Yu, Shuai
    Dwight, Jason
    Siska, Robert C.
    Burkart, Heather
    Quan, Peiran
    Yi, Faliu
    Du, Shimeng
    Daoud, Yahya
    Plant, Kevin
    Criscitiello, Anthony
    Molnar, Joseph
    Thatcher, Jeffrey E.
    BURNS, 2024, 50 (01) : 115 - 122
  • [34] Bayesian estimation of intra-operative deformation for image-guided surgery using 3-D ultrasound
    King, AP
    Blackall, JM
    Penney, GP
    Edwards, PJ
    Hill, DLG
    Hawkes, DJ
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2000, 2000, 1935 : 588 - 597
  • [35] Intra-operative transanal near infrared imaging of colorectal anastomotic perfusion: a feasibility study
    Sherwinter, D. A.
    Gallagher, J.
    Donkar, T.
    COLORECTAL DISEASE, 2013, 15 (01) : 91 - 96
  • [36] Refraction Correction in 3D Transcranial Ultrasound Imaging
    Lindsey, Brooks D.
    Smith, Stephen W.
    ULTRASONIC IMAGING, 2014, 36 (01) : 35 - 54
  • [37] Intra-Operative 3D Pose Estimation of Fractured Bone Segments for Image Guided Orthopedic Surgery
    Gamage, P.
    Xie, S. Q.
    Delmas, P.
    Xu, P.
    Mukherjee, S.
    2008 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND BIOMIMETICS, VOLS 1-4, 2009, : 288 - +
  • [38] 3D intra-operative ultrasound and MR image guidance: pursuing an ultrasound-based management of brainshift to enhance neuronavigation
    Marco Riva
    Christoph Hennersperger
    Fausto Milletari
    Amin Katouzian
    Federico Pessina
    Benjamin Gutierrez-Becker
    Antonella Castellano
    Nassir Navab
    Lorenzo Bello
    International Journal of Computer Assisted Radiology and Surgery, 2017, 12 : 1711 - 1725
  • [39] 3D intra-operative ultrasound and MR image guidance: pursuing an ultrasound-based management of brainshift to enhance neuronavigation
    Riva, Marco
    Hennersperger, Christoph
    Milletari, Fausto
    Katouzian, Amin
    Pessina, Federico
    Gutierrez-Becker, Benjamin
    Castellano, Antonella
    Navab, Nassir
    Bello, Lorenzo
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2017, 12 (10) : 1711 - 1725
  • [40] Intra-operative visualisation of 3D temperature maps and 3D navigation during tissue cryoablation
    Samset, E
    Mala, T
    Aurdal, L
    Balasingham, I
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2005, 29 (06) : 499 - 505