Study of the IMRT interplay effect using a 4DCT Monte Carlo dose calculation

被引:10
|
作者
Jensen, Michael D. [1 ]
Abdellatif, Ady [2 ]
Chen, Jeff [1 ,2 ,3 ]
Wong, Eugene [1 ,2 ,3 ,4 ]
机构
[1] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
[2] London Hlth Sci Ctr, London Reg Canc Program, Dept Phys & Engn, London, ON, Canada
[3] Univ Western Ontario, Dept Oncol, London, ON, Canada
[4] Univ Western Ontario, Dept Phys & Astron, London, ON, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 08期
关键词
DEFORMABLE IMAGE REGISTRATION; INTRA-FRACTION MOTION; QUALITY-ASSURANCE; VERIFICATION; MANAGEMENT; DOSIMETRY; DELIVERY; BEAM;
D O I
10.1088/0031-9155/57/8/N89
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Respiratory motion may lead to dose errors when treating thoracic and abdominal tumours with radiotherapy. The interplay between complex multileaf collimator patterns and patient respiratory motion could result in unintuitive dose changes. We have developed a treatment reconstruction simulation computer code that accounts for interplay effects by combining multileaf collimator controller log files, respiratory trace log files, 4DCT images and a Monte Carlo dose calculator. Two three-dimensional (3D) IMRT step-and-shoot plans, a concave target and integrated boost were delivered to a 1D rigid motion phantom. Three sets of experiments were performed with 100%, 50% and 25% duty cycle gating. The log files were collected, and five simulation types were performed on each data set: continuous isocentre shift, discrete isocentre shift, 4DCT, 4DCT delivery average and 4DCT plan average. Analysis was performed using 3D gamma analysis with passing criteria of 2%, 2 mm. The simulation framework was able to demonstrate that a single fraction of the integrated boost plan was more sensitive to interplay effects than the concave target. Gating was shown to reduce the interplay effects. We have developed a 4DCT Monte Carlo simulation method that accounts for IMRT interplay effects with respiratory motion by utilizing delivery log files.
引用
收藏
页码:N89 / N99
页数:11
相关论文
共 50 条
  • [31] Investigation of the Interplay Effect Between MLC and Lung Tumor Motions Using 4DCT and RPM Profile Data
    Li, H.
    Zhong, H.
    Kim, J.
    Nurushev, T.
    Chetty, I.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [32] A Monte Carlo based MU calculation tool for IMRT
    Qin, L
    Li, J
    Price, R
    Chen, L
    McNeeley, S
    Ding, M
    Fourkal, E
    Ma, C
    MEDICAL PHYSICS, 2002, 29 (06) : 1264 - 1264
  • [33] Simulation of dosimetry impact of 4DCT uncertainty in 4D dose calculation for lung SBRT
    Liu, Gang
    Hu, Fala
    Ding, Xuanfeng
    Li, Xiaoqiang
    Shao, Qihong
    Wang, Yuenan
    Yang, Jing
    Quan, Hong
    RADIATION ONCOLOGY, 2019, 14 (1)
  • [34] Simulation of dosimetry impact of 4DCT uncertainty in 4D dose calculation for lung SBRT
    Gang Liu
    Fala Hu
    Xuanfeng Ding
    Xiaoqiang Li
    Qihong Shao
    Yuenan Wang
    Jing Yang
    Hong Quan
    Radiation Oncology, 14
  • [35] Monte Carlo dose calculation of multiple beamlets for IMRT by using a discriminating particle transport and dose tallying approach on the base of MCNP 4B code
    He, T
    Van den Heuvel, F
    MEDICAL PHYSICS, 2002, 29 (06) : 1231 - 1231
  • [36] Dose Estimation in Lung Brachytherapy Using Voxelized Phantom 4DCT and MCNP
    Hadad, K.
    Zaker, N.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [37] Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculation
    Tyagi, Neelam
    Moran, Jean M.
    Litzenberg, Dale W.
    Bielajew, Alex F.
    Fraass, Benedick A.
    Chetty, Indrin J.
    MEDICAL PHYSICS, 2007, 34 (02) : 651 - 663
  • [38] A Monte-Carlo-based system for dose calculation and dosimetry QA of photon IMRT (WIP)
    Vassiliev, O
    Liu, HH
    Dong, L
    Antolak, J
    Cho, S
    Mohan, R
    MEDICAL PHYSICS, 2003, 30 (06) : 1453 - 1453
  • [39] Monte Carlo dose calculations for dynamic IMRT treatments
    Keall, PJ
    Siebers, JV
    Arnfield, M
    Kim, JO
    Mohan, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04): : 929 - 941
  • [40] A study of the interplay effect in radiation therapy using a Monte-Carlo model
    Leste, Jeremy
    Medjahed, Imene
    Chauvin, Maxime
    Younes, Tony
    Vieillevigne, Laure
    Ferrand, Regis
    Franceries, Xavier
    Bardies, Manuel
    Simon, Luc
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 87 : 73 - 82