Automatic determination of left ventricular ejection fraction from gated blood-pool tomography

被引:0
|
作者
Vanhove, C [1 ]
Franken, PR [1 ]
Defrise, M [1 ]
Momen, A [1 ]
Everaert, H [1 ]
Bossuyt, A [1 ]
机构
[1] Free Univ Brussels, AZ VUB, Univ Hosp, Div Nucl Med, B-1090 Brussels, Belgium
关键词
blood-pool SPECT; automatic algorithm; left ventricular ejection fraction;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to develop and validate a new algorithm to automatically compute left ventricular ejection fraction (LVEF) from gated blood-pool tomography (GBPT). The results were compared with those of conventional planar radionuclide angiocardiography (PRNA). Methods: Fifty-three consecutive patients received an injection of 740 MBq Tc-99m-labeled human serum albumin. PRNA and GBPT were performed consecutively in a random sequence. PRNA served as the reference, and GBPT images were processed using a new edge detection algorithm. The algorithm is fast (<45 s), fully automatic, and works in three-dimensional space. The method includes identification of the Valve plane and the septum, The left ventricular cavity at end-diastole is delineated by segmentation using an iterative threshold technique. An optimal threshold is reached when the corresponding isocontour best fits the first derivative of the end-diastolic count distribution in three dimensions. This optimal threshold is then applied to delineate the left Ventricular cavity on the other time bins. The data are corrected for the partial-volume effect. Left ventricular Volumes are determined using a geometry-based method and are used to calculate the ejection fraction. Results: The success rate of the new algorithm was 94%, LVEFs calculated from GBPT agreed well with those calculated from PRNA (r = 0.78; GBPT = 0.94 PRNA + 6.33). The systematic error was 2.8%, and the random error was 8.8%. Excellent inter- and intraobserver reproducibility was found, with average differences of 1.1% +/- 4.6% and 1.1% +/- 5.0%, respectively, between the two measurements. Conclusion: This new algorithm provides a fast, automated, and objective method to calculate LVEF from GBPT.
引用
收藏
页码:401 / 407
页数:7
相关论文
共 50 条
  • [1] Right and left ventricular volume and ejection fraction by tomographic gated blood-pool scintigraphy
    Chin, BB
    Bloomgarden, DC
    Xia, WS
    Kim, HJ
    Fayad, ZA
    Ferrari, VA
    Berlin, JA
    Axel, L
    Alavi, A
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (06) : 942 - 948
  • [2] Automatic quantification of left ventricular ejection fraction from gated blood pool SPECT
    Serge D. Van Kriekinge
    Daniel S. Berman
    Guido Germano
    Journal of Nuclear Cardiology, 1999, 6 : 498 - 506
  • [3] Automatic quantification of left ventricular ejection fraction from gated blood pool SPECT
    Van Kriekinge, SD
    Berman, DS
    Germano, G
    JOURNAL OF NUCLEAR CARDIOLOGY, 1999, 6 (05) : 498 - 506
  • [4] Automatic quantification of left ventricular ejection fraction from gated blood pool SPECT
    vanKriekinge, SD
    Berman, DS
    Germano, G
    CIRCULATION, 1997, 96 (08) : 1725 - 1725
  • [5] Left ventricular ejection fraction: Comparison of results from planar and SPECT gated blood-pool studies
    Bartlett, ML
    Srinivasan, G
    Barker, WC
    Kitsiou, AN
    Dilsizian, V
    Bacharach, SL
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (11) : 1795 - 1799
  • [6] COMPARISON OF PROTOCOLS IN THE MEASUREMENT OF LEFT-VENTRICULAR EJECTION FRACTION IN GATED BLOOD-POOL SCINTIGRAPHY
    YOUNG, KC
    RAILTON, R
    BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (648): : 1091 - 1093
  • [7] Left ventricular ejection fraction and volumes from gated blood-pool SPECT: Comparison with planar gated blood-pool imaging and assessment of repeatability in patients with heart failure
    Wright, GA
    Thackray, S
    Howey, S
    Cleland, JG
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (04) : 494 - 498
  • [8] GATED BLOOD POOL TRANSAXIAL TOMOGRAPHY - LEFT-VENTRICULAR VOLUMES AND EJECTION FRACTION
    CORBETT, JR
    JANSEN, DE
    LEWIS, SE
    WOLFE, C
    NICOD, P
    REDISH, GR
    GABLIANI, G
    FILIPCHUK, N
    WILLERSON, JT
    CLINICAL RESEARCH, 1983, 31 (05): : A864 - A864
  • [9] GATED BLOOD POOL TRANSAXIAL TOMOGRAPHY - LEFT-VENTRICULAR VOLUMES AND EJECTION FRACTION
    CORBETT, JR
    JANSEN, DE
    LEWIS, SE
    WOLFE, C
    NICOD, P
    REDISH, GR
    GABLIANI, G
    FILIPCHUK, N
    WILLERSON, JT
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 3 (02) : 590 - 590
  • [10] Gated myocardial perfusion tomography versus gated blood pool tomography for the calculation of left ventricular volumes and ejection fraction
    Christian Vanhove
    Natasha Walgraeve
    Frank De Geeter
    Philippe R. Franken
    European Journal of Nuclear Medicine and Molecular Imaging, 2002, 29 : 735 - 741