Correction of detector nonlinearity in Fourier transform spectroscopy with a low-temperature blackbody

被引:14
|
作者
Fiedler, L
Newman, S
Bakan, S
机构
[1] Met Off, Exeter EX1 3PB, Devon, England
[2] European Org Exploitat Meteorol Satellites, D-64295 Darmstadt, Germany
[3] Max Planck Inst Meteorol, D-20146 Hamburg, Germany
关键词
D O I
10.1364/AO.44.005332
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The nonlinearity of a mercury cadmium telluride photoconductive detector, an integral part of a modified commercial interferometer used for airborne research, has been analyzed and evaluated against a number of correction schemes. A high-quality blackbody with accurate temperature control has been used as a stable and well-characterized radiation source. The detector nonlinearity was established as a function of scene temperature between 194 and 263 K. Second- and third-order corrections to the measured interferogram have been tested by analyzing the measured signal both within and outside the spectral response region of the detector. A combined correction scheme is proposed that best represents the real nonlinear response of the detector.
引用
收藏
页码:5332 / 5340
页数:9
相关论文
共 50 条