Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

被引:72
|
作者
Wen, Ning [1 ]
Lu, Siming [1 ]
Kim, Jinkoo [1 ]
Qin, Yujiao [1 ]
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Liu, Chang [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 West Grand Blvd, Detroit, MI 48202 USA
来源
RADIATION ONCOLOGY | 2016年 / 11卷
关键词
Stereotactic radiosurgery; Stereotactic body radiation therapy; Gafchromic films; Quality assurance; IONIZATION-CHAMBER ARRAYS; RADIOCHROMIC FILM; PHOTON BEAMS; SCANNERS; VERIFICATION;
D O I
10.1186/s13014-016-0709-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic (TM) (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Method and materials: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. Results: The scanner response varied within 1 % for the field sizes less than 5 x 5 cm(2), and up to 5 % for the field sizes of 10 x 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (similar to 10 %) in the first two hours and became smooth plateaued afterwards, similar to 3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3 similar to 15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was -0.64 +/- 2.02 for all plans, corresponding to a 95 % confidence interval of (-1.09, -0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 +/- 4.2. Conclusion: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance
    Fuss, Martina
    Sturtewagen, Eva
    De Wagter, Carlos
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14): : 4211 - 4225
  • [42] Sum signal dosimetry: a novel approach for high dose patient specific quality assurance with Gafchromic EBT3
    Cusumano, D.
    Fumagalli, M. L.
    Ghielmetti, F.
    Rossi, L.
    Grossi, G.
    Lanzarotti, R.
    Fariselli, L.
    De Martin, E.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S270 - S271
  • [43] Predicting Photon Dosimetry of EBT3 Gafchromic Films to Improve Accurate Dosimetry Using Machine Learning Method
    Lin, Y-S
    MEDICAL PHYSICS, 2018, 45 (06) : E311 - E311
  • [44] Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic (R) EBT3 film
    Iqbal, Khalid
    Iqbal, Muhammad Mazhar
    Akram, Muhammad
    Altaf, Saima
    Buzdar, Saeed Ahmad
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (01) : 85 - 95
  • [45] Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system
    Aldelaijan, Saad
    Devic, Slobodan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 49 : 112 - 118
  • [46] Characterisation of EPSONV700 flatbed scanner for EBT3 Gafchromic film dosimetry
    Nelson, V.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S737 - S738
  • [47] CALIBRATION OF GAFCHROMIC EBT3 FILM FOR DOSIMETRY OF SCANNING PROTON PENCIL BEAM (PBS)
    Krzempek, D.
    Mianowska, G.
    Bassler, N.
    Stolarczyk, L.
    Kopec, R.
    Sas-Korczynska, B.
    Olko, P.
    RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 324 - 328
  • [48] Skin dose in radiotherapy: results of in vivo measurements with gafchromic EBT3 films
    Giuliano, A.
    Ravaglia, V.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S784 - S784
  • [49] Selection of gamma analysis acceptance criteria in IMRT QA using Gafchromic EBT3 film dosimetry
    Khan, Muhammad Isa
    Shakil, Muhammad
    Tahir, Muhammad Bilal
    Rafique, Muhammad
    Iqbal, Tahir
    Zahoor, Aliza
    Rehman, Jalil Ur
    Iqbal, Khalid
    Chow, James C. L.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (02) : 127 - 131
  • [50] In vivo Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry
    Baba, Misba Hamid
    Singh, Benoy Kumar
    Wani, Shaqul Qamar
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (04) : 362 - 366