Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films

被引:72
|
作者
Wen, Ning [1 ]
Lu, Siming [1 ]
Kim, Jinkoo [1 ]
Qin, Yujiao [1 ]
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Liu, Chang [1 ]
Chetty, Indrin J. [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 West Grand Blvd, Detroit, MI 48202 USA
来源
RADIATION ONCOLOGY | 2016年 / 11卷
关键词
Stereotactic radiosurgery; Stereotactic body radiation therapy; Gafchromic films; Quality assurance; IONIZATION-CHAMBER ARRAYS; RADIOCHROMIC FILM; PHOTON BEAMS; SCANNERS; VERIFICATION;
D O I
10.1186/s13014-016-0709-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study is to evaluate the dosimetric uncertainty associated with Gafchromic (TM) (EBT3) films and establish a practical and efficient film dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Method and materials: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference ion chamber array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to 4 days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 83 SRS/SBRT treatment plans, and analysis of the measurements and calculations are presented here. Results: The scanner response varied within 1 % for the field sizes less than 5 x 5 cm(2), and up to 5 % for the field sizes of 10 x 10 cm(2) for all color channels. The scanner correction method was able to remove visually evident, irregular detector responses for larger field sizes. The dose response of the film changed rapidly (similar to 10 %) in the first two hours and became smooth plateaued afterwards, similar to 3 % change between 2 and 24 h. The uncertainties were approximately 1.5, 1.7 and 4.8 % over the dose range of 3 similar to 15 Gy for the red, green and blue channels. The green channel showed very high sensitivity and low uncertainty in the dose range between 10 and 15 Gy, which is suitable for SRS/SBRT commissioning and PSQA. The difference between the calculated dose and measured dose of ion chamber measurement at isocenter was -0.64 +/- 2.02 for all plans, corresponding to a 95 % confidence interval of (-1.09, -0.26). The percentage of points passing the 3 %/1 mm gamma criteria in absolute dose, averaged over all tests was 95.0 +/- 4.2. Conclusion: We have developed the EBT3 films based dosimetry protocol to obtain absolute dose values. The overall uncertainty has been established to be 1.5 % for SRS and SBRT PSQA.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry
    Rebecca Moylan
    Trent Aland
    Tanya Kairn
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 331 - 337
  • [32] Inter-fractional entrance dose monitoring as quality assurance using Gafchromic EBT3 film
    Moon, Sun Young
    Jo, Yunhui
    Seo, Jaehyeon
    Shin, Dongho
    Yoon, Myonggeun
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2022, 18 (04) : 1152 - 1158
  • [33] A diffuse reflectance spectrophotometer for radiation dosimetry of EBT3 GAFchromic films
    Piliero, Maria A.
    Pupillo, Francesco
    Presilla, Stefano
    Radiation Measurements, 2022, 154
  • [34] External Photon Beam Dosimetry for Linac QA Using Gafchromic EBT3 Film
    Khan, I.
    Tahir, B.
    Rafique, M.
    Iqbal, T.
    Zulfiqar, S.
    Zahoor, A.
    Rehman, J.
    Iqbal, K.
    Chow, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E309 - E309
  • [35] A diffuse reflectance spectrophotometer for radiation dosimetry of EBT3 GAFchromic films
    Piliero, Maria A.
    Pupillo, Francesco
    Presilla, Stefano
    RADIATION MEASUREMENTS, 2022, 154
  • [36] Dosimetry evaluation of 192Ir Flexisource using gafchromic EBT3 film
    Oustous, Aziz
    Airouss, Zouhair
    EL Attaoui, Yacine
    Oustous, Yassine
    Khalis, Mohammed
    Sebihi, Rajaa
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2024, 23
  • [37] Feasibility of Establishing a National Postal Dosimetry System Using EBT3 Gafchromic Film
    Abdel-Rahman, W.
    Mohammed, I.
    Bamajboor, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3431 - 3431
  • [38] Effect of verification imaging on in vivo dosimetry results using Gafchromic EBT3 film
    Aland, Trent
    Moylan, Rebecca
    Kairn, Tanya
    Trapp, Jamie
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (11): : 1461 - 1465
  • [39] Gafchromic EBT3 film for absolute dosimetry in proton therapy based on averaging of beam quality
    Resch, A.
    Heyes, P.
    Fuchs, H.
    Georg, D.
    Hugo, P.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S482 - S483
  • [40] Linearization of EBT3 film dose response and virtual film dosimetry for SBRT quality assurance
    Cai, M.
    Archibald-Heeren, B.
    Wang, Y.
    Metcalfe, P.
    MICRO-MINI & NANO-DOSIMETRY & INNOVATIVE TECHNOLOGIES IN RADIATION THERAPY (MMND&ITRO2016), 2017, 777