Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results

被引:41
|
作者
Jeong, Seonji [1 ]
Kim, Se Hyung [1 ,2 ]
Hwang, Eui Jin [1 ]
Shin, Cheong-Il [1 ,2 ]
Han, Joon Koo [1 ,2 ]
Choi, Byung Ihn [1 ,2 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
[2] Seoul Natl Univ Hosp, Inst Radiat Med, Seoul 110744, South Korea
基金
新加坡国家研究基金会;
关键词
CT; iterative reduction algorithm; metal artifact; phantom; TUBE CURRENT CT; COMPUTED-TOMOGRAPHY; SUPPRESSION; RECONSTRUCTION; ABDOMEN;
D O I
10.2214/AJR.14.12745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate the usefulness of a metal artifact reduction (MAR) algorithm for orthopedic prostheses in phantom and clinical CT. MATERIALS AND METHODS. An agar phantom with two sets of spinal screws was scanned at various tube voltage (80-140 kVp) and tube current-time (34-1032 mAs) settings. The orthopedic MAR algorithm was combined with filtered back projection (FBP) or iterative reconstruction. The mean SDs in three ROIs were compared among four datasets (FBP, iterative reconstruction, FBP with orthopedic MAR, and iterative reconstruction with orthopedic MAR). For the clinical study, the mean SDs of three ROIs and 4-point scaled image quality in 52 patients with metallic orthopedic prostheses were compared between CT images acquired with and without orthopedic MAR. The presence and type of image quality improvement with orthopedic MAR and the presence of orthopedic MAR-related new artifacts were also analyzed. RESULTS. In the phantom study, the mean SD with orthopedic MAR was significantly lower than that without orthopedic MAR regardless of dose settings and reconstruction algorithms (FBP versus iterative reconstruction). The mean SD near the metallic prosthesis in 52 patients was significantly lower on CT images with orthopedic MAR (28.04 HU) than those without it (49.21 HU). Image quality regarding metallic artifact was significantly improved with orthopedic MAR (rating of 2.60 versus 1.04). Notable reduction of metallic artifacts and better depiction of abdominal organs were observed in 45 patients. Diagnostic benefit was achieved in six patients, but orthopedic MAR-related new artifacts were seen in 30 patients. CONCLUSION. Use of the orthopedic MAR algorithm significantly reduces metal artifacts in CT of both phantoms and patients and has potential for improving diagnostic performance in patients with severe metallic artifacts.
引用
收藏
页码:307 / 317
页数:11
相关论文
共 50 条
  • [31] Influence of CT metal artifact reduction on SPECT/CT quantification of bone scintigraphy - Retrospective study for selected types of metal implants
    Kugler, Amelie
    Kuwert, Torsten
    Ritt, Philipp
    Grings, Andreas
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2022, 61 (06): : 449 - 457
  • [32] Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results
    Toso, Seema
    Laurent, Meryle
    Lozeron, Elise Dupuis
    Brindel, Pauline
    Lacalamita, Marirosa Cristallo
    Hanquinet, Sylviane
    PEDIATRIC RADIOLOGY, 2018, 48 (13) : 1884 - 1890
  • [33] Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results
    Seema Toso
    Meryle Laurent
    Elise Dupuis Lozeron
    Pauline Brindel
    Marirosa Cristallo Lacalamita
    Sylviane Hanquinet
    Pediatric Radiology, 2018, 48 : 1884 - 1890
  • [34] Evaluation of a metal artifact reduction algorithm for radiotherapy CT scans
    Rechner, L.
    Kovacs, D.
    Bangsgaard, A.
    Berthelsen, A.
    Aznar, M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S442 - S443
  • [35] A CT metal artifact reduction algorithm based on sinogram surgery
    Jeon, Soomin
    Lee, Chang-Ock
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (03) : 413 - 434
  • [36] An Algorithm for Metal Streaking Artifact Reduction in Cone Beam CT
    Bazalova, M.
    Landry, G.
    Beaulieu, L.
    Reniers, B.
    Verhaegen, F.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [37] Comparison of Metal Artifact Reduction in Dual- and Single-Source CT: A Vertebral Phantom Study
    Jagoda, Philippe
    Schmitz, Daniel
    Wagenpfeil, Stefan
    Buecker, Arno
    Minko, Peter
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 211 (06) : 1298 - 1305
  • [38] Comparison of the Phantom Dose Distribution of Dual-Energy CT Scan and That of Single-Energy CT Scan with Orthopedic Metal Artifact Reduction Algorithm Using Volumetric Arc Therapy in Monaco
    Li, K.
    Leung, W.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [39] Dosimetric Impact of Metal Artifact Reduction in Treatment Planning: Phantom and Clinical Studies
    Chen, D.
    Chetty, I.
    Glide-Hurst, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [40] Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?
    Hildegard M. Wichtmann
    Kai R. Laukamp
    Sebastian Manneck
    Konrad Appelt
    Bram Stieltjes
    Daniel T. Boll
    Matthias R. Benz
    Markus M. Obmann
    Abdominal Radiology, 2023, 48 : 424 - 435