Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results

被引:42
|
作者
Jeong, Seonji [1 ]
Kim, Se Hyung [1 ,2 ]
Hwang, Eui Jin [1 ]
Shin, Cheong-Il [1 ,2 ]
Han, Joon Koo [1 ,2 ]
Choi, Byung Ihn [1 ,2 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
[2] Seoul Natl Univ Hosp, Inst Radiat Med, Seoul 110744, South Korea
基金
新加坡国家研究基金会;
关键词
CT; iterative reduction algorithm; metal artifact; phantom; TUBE CURRENT CT; COMPUTED-TOMOGRAPHY; SUPPRESSION; RECONSTRUCTION; ABDOMEN;
D O I
10.2214/AJR.14.12745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate the usefulness of a metal artifact reduction (MAR) algorithm for orthopedic prostheses in phantom and clinical CT. MATERIALS AND METHODS. An agar phantom with two sets of spinal screws was scanned at various tube voltage (80-140 kVp) and tube current-time (34-1032 mAs) settings. The orthopedic MAR algorithm was combined with filtered back projection (FBP) or iterative reconstruction. The mean SDs in three ROIs were compared among four datasets (FBP, iterative reconstruction, FBP with orthopedic MAR, and iterative reconstruction with orthopedic MAR). For the clinical study, the mean SDs of three ROIs and 4-point scaled image quality in 52 patients with metallic orthopedic prostheses were compared between CT images acquired with and without orthopedic MAR. The presence and type of image quality improvement with orthopedic MAR and the presence of orthopedic MAR-related new artifacts were also analyzed. RESULTS. In the phantom study, the mean SD with orthopedic MAR was significantly lower than that without orthopedic MAR regardless of dose settings and reconstruction algorithms (FBP versus iterative reconstruction). The mean SD near the metallic prosthesis in 52 patients was significantly lower on CT images with orthopedic MAR (28.04 HU) than those without it (49.21 HU). Image quality regarding metallic artifact was significantly improved with orthopedic MAR (rating of 2.60 versus 1.04). Notable reduction of metallic artifacts and better depiction of abdominal organs were observed in 45 patients. Diagnostic benefit was achieved in six patients, but orthopedic MAR-related new artifacts were seen in 30 patients. CONCLUSION. Use of the orthopedic MAR algorithm significantly reduces metal artifacts in CT of both phantoms and patients and has potential for improving diagnostic performance in patients with severe metallic artifacts.
引用
收藏
页码:307 / 317
页数:11
相关论文
共 50 条
  • [21] Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy
    Axente, Marian
    Paidi, Ajay
    Von Eyben, Rie
    Zeng, Chuan
    Bani-Hashemi, Ali
    Krauss, Andreas
    Hristov, Dimitre
    MEDICAL PHYSICS, 2015, 42 (03) : 1170 - 1183
  • [22] Photon-counting detector computed tomography for metal artifact reduction: a comparative study of different artifact reduction techniques in patients with orthopedic implants
    Pallasch, Fabian Bernhard
    Rau, Alexander
    Reisert, Marco
    Rau, Stephan
    Diallo, Thierno
    Stein, Thomas
    Faby, Sebastian
    Bamberg, Fabian
    Weiss, Jakob
    RADIOLOGIA MEDICA, 2024, 129 (06): : 890 - 900
  • [23] A Constrained Optimization Algorithm for CT Metal Artifact Reduction
    Zhang, X.
    Wang, J.
    Xing, L.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [24] Optimization of Monoenergetic Extrapolations in Dual-Energy CT for Metal Artifact Reduction in Different Body Regions and Orthopedic Implants
    Horat, Lukas
    Hamie, Mustafa Q.
    Huber, Florian A.
    Guggenberger, Roman
    ACADEMIC RADIOLOGY, 2019, 26 (05) : E67 - E74
  • [25] Metal artifact reduction using iterative CBCT reconstruction algorithm for head and neck radiation therapy: A phantom and clinical study
    Washio, Hayate
    Ohira, Shingo
    Funama, Yoshinori
    Morimoto, Masahiro
    Wada, Kentaro
    Yagi, Masashi
    Shimamoto, Hiroaki
    Koike, Yuhei
    Ueda, Yoshihiro
    Karino, Tsukasa
    Inui, Shoki
    Nitta, Yuya
    Miyazaki, Masayoshi
    Teshima, Teruki
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 132
  • [26] Clinical Implementation of the Philips Orthopedic Metallic Artifact Reduction (OMAR) Algorithm for Proton Radiotherapy CT Simulations
    Miksys, N.
    Curley, C.
    Tokarz, R.
    MEDICAL PHYSICS, 2022, 49 (06) : E699 - E699
  • [27] Improving CT-Based PET Attenuation Correction in the Vicinity of Metal Implants by an Iterative Metal Artifact Reduction Algorithm of CT Data and Its Comparison to Dual-Energy-Based Strategies A Phantom Study
    Schabel, Christoph
    Gatidis, Sergios
    Bongers, Malte
    Huettig, Fabian
    Bier, Georg
    Kupferschlaeger, Juergen
    Bamberg, Fabian
    la Fougere, Christian
    Nikolaou, Konstantin
    Pfannenberg, Christina
    INVESTIGATIVE RADIOLOGY, 2017, 52 (01) : 61 - 65
  • [28] Image quality and metal artifact reduction in total hip arthroplasty CT: deep learning-based algorithm versus virtual monoenergetic imaging and orthopedic metal artifact reduction
    Selles, Mark
    Wellenberg, Ruud H. H.
    Slotman, Derk J.
    Nijholt, Ingrid M.
    van Osch, Jochen A. C.
    van Dijke, Kees F.
    Maas, Mario
    Boomsma, Martijn F.
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [29] Image quality and metal artifact reduction in total hip arthroplasty CT: deep learning-based algorithm versus virtual monoenergetic imaging and orthopedic metal artifact reduction
    Mark Selles
    Ruud H. H. Wellenberg
    Derk J. Slotman
    Ingrid M. Nijholt
    Jochen A. C. van Osch
    Kees F. van Dijke
    Mario Maas
    Martijn F. Boomsma
    European Radiology Experimental, 8
  • [30] Metal Artifact Reduction in Virtual Monoenergetic Spectral Dual-Energy CT of Patients With Metallic Orthopedic Implants in the Distal Radius
    Yoo, Hye Jin
    Hong, Sung Hwan
    Chung, Bo Mi
    Moon, Sun Jeong
    Choi, Ja-Young
    Chae, Hee Dong
    Chang, Min-Yung
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 211 (05) : 1083 - 1091