Re-treatment rates after treatment with the Pipeline Embolization Device alone versus Pipeline and coil embolization of cerebral aneurysms: a single-center experience

被引:44
|
作者
Park, Min S. [1 ]
Nanaszko, Michael [2 ]
Sanborn, Matthew R. [2 ]
Moon, Karam [2 ]
Albuquerque, Felipe C. [2 ]
McDougall, Cameron G. [2 ]
机构
[1] Univ Utah Hlth Care, Dept Neurosurg, Salt Lake City, UT USA
[2] St Josephs Hosp, Barrow Neurol Inst, Dept Neurosurg, Phoenix, AZ USA
关键词
aneurysms; coil embolization; flow diversion; Pipeline Embolization Device; re-treatment; vascular disorders; INTRACRANIAL ANEURYSMS; FLOW-DIVERSION; RECONSTRUCTION; COMPLICATIONS;
D O I
10.3171/2015.7.JNS15582
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE The optimal strategy for use of the Pipeline Embolization Device (PED, ev3 Neurovascular) has not been clearly defined. The authors examined re-treatment rates after treatment with PED alone versus PED and adjunctive coil embolization (PED/coil). METHODS The authors retrospectively examined cerebral aneurysms treated with the PED from May 2011 to March 2014. Overall, 133 patients (25 men, 108 women; mean age 60.4 years, range 23-85 years) were treated for 140 aneurysms (mean size 11.8 +/- 8.3 mm) requiring 224 PEDs (mean 1.7 PEDs per patient). Sixty-eight patients (13 men, 55 women) were treated with PED alone for 73 aneurysms (mean size 10.6 +/- 9.2 mm) and 65 patients (12 men, 53 women) were treated with PED/coil for 67 aneurysms (mean size 12.8 +/- 7.4 mm). RESULTS Eight aneurysms in 8 patients were re-treated in the PED-alone cohort versus only 1 aneurysm in 1 patient in the PED/coil cohort for re-treatment rates of 11.8% (8/68) and 1.5% (1/65), respectively (p = 0.03). Two patients in the PED-alone cohort were re-treated due to PED contraction, while the other 6 were re-treated for persistent filling of the aneurysms. The PED/coil patient experienced continued filling of a vertebrobasilar artery aneurysm. No aneurysms in either group ruptured after treatment. CONCLUSIONS Adjunctive coil embolization during flow diversion with the PED resulted in a significantly lower re treatment rate compared with PED alone, suggesting an added benefit with adjunctive coil embolization. This result may provide the basis for future evaluation with randomized, controlled trials.
引用
收藏
页码:137 / 144
页数:8
相关论文
共 50 条
  • [31] First Indian single center experience with pipeline embolization device for complex intracranial aneurysms
    Cherian, Mathew P.
    Yadav, Manish Kumar
    Mehta, Pankaj
    Vijayan, K.
    Arulselvan, V.
    Jayabalan, Suresh
    NEUROLOGY INDIA, 2014, 62 (06) : 618 - 624
  • [32] Pipeline Embolization Device for the Treatment of Unruptured Intracranial Dissecting Aneurysms
    Chen, Jigang
    Tao, Mushun
    Han, Jiangli
    Feng, Xin
    Peng, Fei
    Tong, Xin
    Niu, Hao
    Ma, Ning
    Liu, Aihua
    FRONTIERS IN NEUROLOGY, 2021, 12
  • [33] Safety of the Pipeline Embolization Device in Treatment of Posterior Circulation Aneurysms
    Phillips, T. J.
    Wenderoth, J. D.
    Phatouros, C. C.
    Rice, H.
    Singh, T. P.
    Devilliers, L.
    Wycoco, V.
    Meckel, S.
    McAuliffe, W.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (07) : 1225 - 1231
  • [34] A reappraisal of the Pipeline embolization device for the treatment of posterior circulation aneurysms
    Albuquerque, Felipe C.
    Park, Min S.
    Abla, Adib A.
    Crowley, R. Webster
    Ducruet, Andrew F.
    McDougall, Cameron G.
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (09) : 641 - 645
  • [35] Treatment of Posterior Circulation Aneurysms With the Pipeline Embolization Device COMMENT
    Serrone, Joe
    Ringer, Andrew
    NEUROSURGERY, 2013, 72 (06) : 889 - 889
  • [36] Treatment of Blister-Like Aneurysms With the Pipeline Embolization Device
    Chalouhi, Nohra
    Zanaty, Mario
    Tjoumakaris, Stavropoula
    Gonzalez, L. Fernando
    Hasan, David
    Kung, David
    Rosenwasser, Robert H.
    Jabbour, Pascal
    NEUROSURGERY, 2014, 74 (05) : 527 - 532
  • [37] The Role of the Pipeline Embolization Device for the Treatment of Dissecting Intracranial Aneurysms
    de Barros Faria, M.
    Nella Castro, R.
    Lundquist, J.
    Scrivano, E.
    Ceratto, R.
    Ferrario, A.
    Lylyk, P.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (11) : 2192 - 2195
  • [38] Treatment of Ruptured Intracranial Aneurysms With the Pipeline Embolization Device COMMENT
    Gemmete, Joseph J.
    NEUROSURGERY, 2015, 76 (02) : 172 - 172
  • [39] Predictors of Clinical Outcome After Treatment of Intracranial Aneurysms with the Pipeline Embolization Device
    Griffin, Andrew
    Reese, Vanessa
    Huseyinoglu, Zeynep
    Niedzwiecki, Donna
    Yang, Lexie
    Cutler, Andrew
    Gonzalez, L. Fernando
    Zomorodi, Ali
    Smith, Tony
    Hauck, Erik F.
    WORLD NEUROSURGERY, 2019, 130 : E666 - E671
  • [40] Pipeline Embolization Device for Recurrent Cerebral Aneurysms after Microsurgical Clipping
    Adeeb, Nimer
    Griessenauer, Christoph J.
    Moore, Justin
    Stapleton, Christopher J.
    Patel, Aman B.
    Gupta, Raghav
    Patel, Apar S.
    Thomas, Ajith J.
    Ogilvy, Christopher S.
    WORLD NEUROSURGERY, 2016, 93 : 341 - 345