Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction

被引:93
|
作者
Manzke, R [1 ]
Grass, M
Nielsen, T
Shechter, G
Hawkes, D
机构
[1] Philips Res Labs, Sector Tech Syst, Hamburg, Germany
[2] KCL, Imaging Sci Grp, London, England
[3] Philips Med Syst, Haifa, Israel
关键词
cardiac cone beam computed tomography; temporal resolution; optimization of approximate 3D reconstruction;
D O I
10.1118/1.1624756
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cone beam computed tomography scanners in combination with heart rate adaptive reconstruction schemes have the potential to enable cardiac volumetric computed tomography (CT) imaging for a larger number of patients and applications. In this publication, an adaptive scheme for the automatic and patient-specific reconstruction optimization is introduced to improve the temporal resolution and image quality. The optimization method permits the automatic determination of the required amount of gated helical cone beam projection data for the reconstruction volume. It furthermore allows one to optimize subvolume reconstruction yielding an increased temporal resolution. In addition, methods for the assessment of the temporal resolution are given which enable a quantitative documentation of the reconstruction improvements. Results are presented for patient data sets acquired in low pitch helical mode using a 16-slice cone beam CT system with parallel ECG recording. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3072 / 3080
页数:9
相关论文
共 50 条
  • [31] Cardiac cone-beam CT volume reconstruction using ART
    Nielsen, T
    Manzke, R
    Proksa, R
    Grass, M
    MEDICAL PHYSICS, 2005, 32 (04) : 851 - 860
  • [32] Helical CT Reconstruction with Large Cone Angle
    Zamyatin, Alexander A.
    Katsevich, Alexander
    Silver, Michael D.
    Nakanishi, Satoru
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2264 - 2267
  • [33] Helical cone beam CT with an asymmetrical detector
    Zamyatin, AA
    Taguchi, K
    Silver, MD
    MEDICAL PHYSICS, 2005, 32 (10) : 3117 - 3127
  • [34] An exact reconstruction algorithm for triple-source helical cone-beam CT
    Zhao, Jun
    Jiang, Ming
    Zhuang, Tiange
    Wang, Ge
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2006, 14 (03) : 191 - 206
  • [35] Three-term exact FBP reconstruction in cone-beam helical CT
    Zou, Y
    Pan, XC
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2735 - 2738
  • [36] Backprojection-filtration reconstruction for helical cone-beam CT with curved detectors
    Zuo, Nianming
    Xia, Dan
    Sidky, Emil Y.
    Yu, Lifeng
    Zou, Yu
    Pan, Xiaochuan
    Tang, Tianzi
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2306 - 2309
  • [37] An improved exact FBP algorithm for image reconstruction in cone-beam helical CT
    Ma, Jianhua
    Chen, Wufan
    2006 INTERNATIONAL CONFERENCE ON COMPUTATIONAL INTELLIGENCE AND SECURITY, PTS 1 AND 2, PROCEEDINGS, 2006, : 1635 - 1640
  • [38] Exact image reconstruction in reverse helical cone-beam CT for radiation therapy
    Cho, S.
    Xia, D.
    Pelizzari, C.
    Pan, X.
    MEDICAL PHYSICS, 2007, 34 (06) : 2633 - 2634
  • [39] Optimization-Based 3D Variable Resolution Image Reconstruction in Cone-Beam CT
    Zhang, Zheng
    Sidky, Emil Y.
    Pan, Xiaochuan
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [40] A projection-driven pre-correction technique for iterative reconstruction of helical cone-beam cardiac CT images
    Do, Synho
    Liang, Zhuangli
    Karl, William Clem
    Brady, Thomas
    Pien, Homer
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913