Interferometric measurement of the axial length of the eye

被引:0
|
作者
Hård, S [1 ]
Fuxin, PO [1 ]
Gerhardsson, D [1 ]
Torstensson, PA [1 ]
Nilsson, T [1 ]
Sjöstrand, J [1 ]
机构
[1] Chalmers Univ Technol, S-41296 Gothenburg, Sweden
关键词
interferometry; eye length measurement; axial length; bifocal lens;
D O I
10.1117/12.339166
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A fast and accurate fiberoptic-based interferometer for non-contact measurements of the axial length of the eye was constructed and tested. A novelty of the interferometer design is the use of a diffractive, bifocal lens, which simultaneously focuses on the cornea and on the retina. A superluminescent light emitting diode with a coherence length of 30 mu m is used as light source. The axial length of the eye is measured by rapidly translating a reference mirror a distance covering the axial length. The estimated accuracy of a measurement is about 100 mu m, a number which compares well with those of conventional ultrasonic-based measurements.