Amperometric glucose biosensors based on Prussian Blue- and polyaniline-glucose oxidase modified electrodes

被引:158
|
作者
Garjonyte, R [1 ]
Malinauskas, A [1 ]
机构
[1] Inst Chem, LT-2600 Vilnius, Lithuania
来源
BIOSENSORS & BIOELECTRONICS | 2000年 / 15卷 / 9-10期
关键词
biosensor; glucose oxidase; polyaniline; Prussian Blue;
D O I
10.1016/S0956-5663(00)00101-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The properties of glucose sensors fabricated by immobilization of glucose oxidase in a layer of electrochemically deposited polyaniline were investigated. Selective amperometric glucose sensors were prepared by immobilization of glucose oxidase on a Prussian Blue-modified platinum electrode in a layer of polyaniline during a one-step electropolymerization procedure from phosphate buffer. The influence of ascorbic acid and acetaminophen was completely eliminated due to impermeability of polyaniline to these substances. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:445 / 451
页数:7
相关论文
共 50 条
  • [21] Amperometric biosensors based on carbon paste electrodes modified with nanostructured mixed-valence manganese oxides and glucose oxidase
    Cui, Xiaoli
    Liu, Guodong
    Lin, Yuehe
    NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2005, 1 (02) : 130 - 135
  • [23] A paper-based amperometric glucose biosensor developed with Prussian Blue-modified screen-printed electrodes
    Sekar, Nadia Chandra
    Shaegh, Seyed Ali Mousavi
    Ng, Sum Huan
    Ge, Liya
    Tan, Swee Ngin
    SENSORS AND ACTUATORS B-CHEMICAL, 2014, 204 : 414 - 420
  • [24] An Amperometric Glucose Biosensor Composed of Prussian Blue, Nafion, and Glucose Oxidase Studied by Flow Injection Analysis
    Drew, Steven M.
    Belzer, Tristan
    JOURNAL OF CHEMICAL EDUCATION, 2023, 100 (02) : 760 - 766
  • [25] Glucose biosensing based on the highly efficient immobilization of glucose oxidase on a Prussian blue modified nanostructured Au surface
    Ahmadalinezhad, Asieh
    Kafi, A. K. M.
    Chen, Aicheng
    ELECTROCHEMISTRY COMMUNICATIONS, 2009, 11 (10) : 2048 - 2051
  • [26] Amperometric glucose microbiosensor based on a Prussian Blue modified carbon fiber electrode for physiological applications
    Salazar, P.
    O'Neill, R. D.
    Martin, M.
    Roche, R.
    Gonzalez-Mora, J. L.
    SENSORS AND ACTUATORS B-CHEMICAL, 2011, 152 (02) : 137 - 143
  • [27] Amperometric glucose biosensors based on glassy carbon and SWCNT-modified glassy carbon electrodes
    Carpani, Irene
    Scavetta, Erika
    Tonelli, Domenica
    ELECTROANALYSIS, 2008, 20 (01) : 84 - 90
  • [28] Galactose Oxidase/Prussian Blue Based Biosensors
    Charmantray, Franck
    Touisni, Nadia
    Hecquet, Laurence
    Noguer, Thierry
    Mousty, Christine
    ELECTROANALYSIS, 2015, 27 (06) : 1341 - 1344
  • [29] COMPARISON OF AMPEROMETRIC MEASURING PRINCIPLES FOR DETERMINATIONS OF GLUCOSE WITH ELECTRODES BASED ON GLUCOSE-OXIDASE
    ASPERGER, L
    GEPPERT, G
    KRABISCH, C
    ANALYTICA CHIMICA ACTA, 1987, 201 : 281 - 287
  • [30] Novel Glucose Oxidase Interlocked Prussian Blue/Polysulfone Stereo-Structure and its Application in Amperometric Glucose Biosensor
    Wang, Cang
    Pan, Min
    Chen, Hang
    Chen, Dajing
    Chen, Yuquan
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2019, 14 (08): : 8014 - 8027