EVALUATING THE MAXIMUM PATIENT RADIATION DOSE IN CARDIAC INTERVENTIONAL PROCEDURES

被引:16
|
作者
Kato, Mamoru [1 ,2 ]
Chida, Koichi [2 ]
Sato, Tadaya [3 ]
Oosaka, Hajime [1 ]
Tosa, Tetsuo [1 ]
Kadowaki, Ken [3 ]
机构
[1] Akita Med Ctr, Dept Radiol, Akita 0100874, Japan
[2] Tohoku Univ, Grad Sch Med, Course Radiol Technol, Aoba Ku, Sendai, Miyagi 9808575, Japan
[3] Akita Med Ctr, Dept Cardiol, Akita 0100874, Japan
关键词
FLUOROSCOPICALLY GUIDED PROCEDURES; SKIN INJURIES; CATHETERIZATION; PHYSICIANS; CARDIOLOGY; EXPOSURE; TIME;
D O I
10.1093/rpd/ncq286
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Many of the X-ray systems that are used for cardiac interventional radiology provide no way to evaluate the patient maximum skin dose (MSD). The authors report a new method for evaluating the MSD by using the cumulative patient entrance skin dose (ESD), which includes a back-scatter factor and the number of cineangiography frames during percutaneous coronary intervention (PCI). Four hundred consecutive PCI patients (315 men and 85 women) were studied. The correlation between the cumulative ESD and number of cineangiography frames was investigated. The irradiation and overlapping fields were verified using dose-mapping software. A good correlation was found between the cumulative ESD and the number of cineangiography frames. The MSD could be estimated using the proportion of cineangiography frames used for the main angle of view relative to the total number of cineangiography frames and multiplying this by the cumulative ESD. The average MSD (3.0 +/- 1.9 Gy) was lower than the average cumulative ESD (4.6 +/- 2.6 Gy). This method is an easy way to estimate the MSD during PCI.
引用
收藏
页码:69 / 73
页数:5
相关论文
共 50 条
  • [41] Assessment of maximum entrance skin dose of patients undergoing cardiac interventional procedures and its correlation with other dosimetric parameters
    Toossi M.T.B.
    Khosroabadi M.
    Mehrpouyan M.
    Moghaddam H.R.
    Iranian Journal of Medical Physics, 2020, 17 (04) : 235 - 246
  • [42] Patient radiation exposure related to complexity of coronary interventional procedures
    Federman, J
    Cusma, JT
    Holmes, DR
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 54A - 54A
  • [43] Patient peak skin doses from cardiac interventional procedures
    Zontar, D.
    Kuhelj, D.
    Skrk, D.
    Zdesar, U.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 262 - 265
  • [44] Evaluating operator’s organ dose and effective dose according to the lead equivalent of radiation protection devices in performing angiography and interventional procedures
    Dong-Hee Han
    Man-Seok Han
    Seung-Jae Lee
    Byung-Seok Shin
    Jang-Oh Kim
    Da-Eun Kwon
    Kyung-Hwan Jung
    Seung-Hwan Lee
    Cheol-Ha Baek
    Journal of the Korean Physical Society, 2023, 82 : 315 - 320
  • [45] Patient radiation doses in paediatric interventional cardiology procedures: a review
    Harbron, R. W.
    Dreuil, S.
    Bernier, M-O
    Pearce, M. S.
    Thierry-Chef, I.
    Chapple, C-L
    Baysson, H.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (04) : R131 - R144
  • [46] Evaluating operator's organ dose and effective dose according to the lead equivalent of radiation protection devices in performing angiography and interventional procedures
    Han, Dong-Hee
    Han, Man-Seok
    Lee, Seung-Jae
    Shin, Byung-Seok
    Kim, Jang-Oh
    Kwon, Da-Eun
    Jung, Kyung-Hwan
    Lee, Seung-Hwan
    Baek, Cheol-Ha
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2023, 82 (03) : 315 - 320
  • [47] Optimizing patient radiation dose in intervention procedures
    Chida, Koichi
    Kato, Mamoru
    Saito, Haruo
    Ishibashi, Tadashi
    Takahashi, Shoki
    Kohzuki, Masahiro
    Zuguchi, Masayuki
    ACTA RADIOLOGICA, 2010, 51 (01) : 33 - 39
  • [48] Using Radiation Dose Structured Reports in the Patient Skin Dose Tracking System (DTS) for Fluoroscopically-Guided Interventional Procedures
    Troville, J.
    Collins, J.
    Islam, M.
    Nosal, J.
    Woods, J.
    Syeda, A.
    Orji, M.
    Rudin, S.
    Ionita, C.
    Bednarek, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E217 - E218
  • [49] Radiation Dose in CT-guided Interventional Procedures: Establishing a Benchmark
    Leng, Shuai
    RADIOLOGY, 2018, 289 (01) : 158 - 159
  • [50] Medical staff dose estimation during pediatric cardiac interventional procedures
    Roberto, Maryanna Regina
    Leite, Luiz M. A.
    Cruz, Aricia R. P.
    Khoury, Helen
    de Barros, Vinicius S. M.
    Neves, Juliana R.
    Cartaxo, Jessica M.
    Ubeda, Carlos
    Asfora, Viviane K.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (02)