Reliability of EUCLIDIAN: An autonomous robotic system for image-guided prostate brachytherapy

被引:6
|
作者
Podder, Tarun K. [1 ]
Buzurovic, Ivan [1 ]
Huang, Ke [1 ]
Showalter, Timothy [1 ]
Dicker, Adam P. [1 ]
Yu, Yan [1 ]
机构
[1] Thomas Jefferson Univ, Dept Radiat Oncol, Kimmel Canc Ctr NCI Designated, Philadelphia, PA 19107 USA
关键词
brachytherapy robot; reliability of robot; medical robot; image-guided brachytherapy; prostate brachytherapy; prostate seed implant; GROWTH;
D O I
10.1118/1.3523097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Recently, several robotic systems have been developed to perform accurate and consistent image-guided brachytherapy. Before introducing a new device into clinical operations, it is important to assess the reliability and mean time before failure (MTBF) of the system. In this article, the authors present the preclinical evaluation and analysis of the reliability and MTBF of an autonomous robotic system, which is developed for prostate seed implantation. Methods: The authors have considered three steps that are important in reliability growth analysis. These steps are: Identification and isolation of failures, classification of failures, and trend analysis. For any one-of-a-kind product, the reliability enhancement is accomplished through test-fix-test. The authors have used failure mode and effect analysis for collection and analysis of reliability data by identifying and categorizing the failure modes. Failures were classified according to severity. Failures that occurred during the operation of this robotic system were considered as nonhomogenous Poisson process. The failure occurrence trend was analyzed using Laplace test. For analyzing and predicting reliability growth, commonly used and widely accepted models, Duane's model and the Army Material Systems Analysis Activity, i.e., Crow's model, were applied. The MTBF was used as an important measure for assessing the system's reliability. Results: During preclinical testing, 3196 seeds (in 53 test cases) were deposited autonomously by the robot and 14 critical failures were encountered. The majority of the failures occurred during the first few cases. The distribution of failures followed Duane's postulation as well as Crow's postulation of reliability growth. The Laplace test index was 3.82 (< 0), indicating a significant trend in failure data, and the failure intervals lengthened gradually. The continuous increase in the failure occurrence interval suggested a trend toward improved reliability. The MTBF was 592 seeds, which implied that several prostate seed implantation cases would be possible without encountering any critical failure. The shape parameter for the MTBF was 0.3859 (< 1), suggesting a positive reliability growth of this robotic system. At 95% confidence, the reliability for deposition of 65 seeds was more than 90%. Conclusions: Analyses of failure mode strongly indicated a gradual improvement of reliability of this autonomous robotic system. High MTBF implied that several prostate seed implant cases would be possible without encountering any critical failure. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3523097]
引用
收藏
页码:96 / 106
页数:11
相关论文
共 50 条
  • [21] An image-guided robotic surgery system for spinal fusion
    Chung, GB
    Kim, S
    Lee, SG
    Yi, BJ
    Kim, W
    Oh, SM
    Kim, YS
    So, BR
    Park, JI
    Oh, SH
    INTERNATIONAL JOURNAL OF CONTROL AUTOMATION AND SYSTEMS, 2006, 4 (01) : 30 - 41
  • [22] Image-guided Brachytherapy for Gynecologic Surgeons
    Chino, Junzo
    Secord, Angeles Alvarez
    SURGICAL ONCOLOGY CLINICS OF NORTH AMERICA, 2013, 22 (03) : 495 - +
  • [23] Magnetic resonance image-guided focused ultrasound robotic system for transrectal prostate cancer therapy
    Giannakou, Marinos
    Drakos, Theocharis
    Menikou, Georgios
    Evripidou, Nikolas
    Filippou, Antria
    Spanoudes, Kyriakos
    Ioannou, Leonidas
    Damianou, Christakis
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2021, 17 (03):
  • [24] An ultrasound probe holder for image-guided robot-assisted prostate brachytherapy
    Yousef, Basem
    Patel, Rajni
    Moallem, Mehrdad
    PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, 2007, : 232 - 237
  • [25] The benefit of intraoperative image-guided adaptative prostate brachytherapy:: Toxicity and outcome.
    Vigneault, E.
    Beaulieu, L.
    Harel, F.
    Martin, A-G.
    Aubin, S.
    Chretien, M.
    Varlfalvy, N.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S41 - S41
  • [26] AAPMGEC-ESTRO Task Group 342: An Update on Image-Guided Robotic Brachytherapy
    Podder, T. K.
    Buzurovic, I.
    Cunha, J. A. M.
    Langmack, C.
    Merzouki, R.
    Moreland, M.
    Rivard, M. J.
    Schlaefer, A.
    Song, D.
    Traughber, B. J., Jr.
    Yan, K.
    Yu, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : S110 - S111
  • [27] Image-guided Prostate biopsy
    Hadaschik, B.
    AKTUELLE UROLOGIE, 2012, 43 (05) : 279 - 280
  • [28] Stereo-fluoroscopic image-guided robotic biopsy system
    Shi, MY
    Liu, H
    Tao, G
    Fajardo, LL
    BIOMEDICAL DIAGNOSTIC, GUIDANCE, AND SURGICAL-ASSIST SYSTEMS, 1999, 3595 : 193 - 200
  • [29] A robotic system for automated image-guided transcranial magnetic stimulation
    Lebosse, Cyrille
    Renaud, Pierre
    Bayle, Bernard
    de Mathelin, Michel
    Piccin, Olivier
    Foucher, Jack
    2007 IEEE/NIH LIFE SCIENCE SYSTEMS AND APPLICATIONS WORKSHOP, 2007, : 55 - +
  • [30] Development of an image-guided robotic system for surgical positioning and drilling
    Tseng, Ching-Shiow
    Huang, Chiao-Chi
    Chen, Chen-San
    ROBOTICA, 2007, 25 (03) : 375 - 383