Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT

被引:61
|
作者
De Zordo, Tobias [1 ]
von Lutterotti, Klemens [1 ]
Dejaco, Christian [2 ]
Soegner, Peter F. [1 ]
Frank, Renate [1 ]
Aigner, Friedrich [1 ]
Klauser, Andrea S. [1 ]
Pechlaner, Christoph [3 ]
Schoepf, U. Joseph [4 ]
Jaschke, Werner R. [1 ]
Feuchtner, Gudrun M. [1 ]
机构
[1] Med Univ Innsbruck, Dept Radiol, A-6020 Innsbruck, Austria
[2] Med Univ Graz, Dept Rheumatol, A-8036 Graz, Austria
[3] Med Univ Innsbruck, Dept Internal Med, A-6020 Innsbruck, Austria
[4] Med Univ S Carolina, Heart & Vasc Ctr, Charleston, SC 29425 USA
关键词
Dual energy; Dual source; Computer tomography; Pulmonary embolism; Radiation; MULTIDETECTOR-ROW CT; COMPUTED-TOMOGRAPHY; LUNG PERFUSION; EMBOLISM; ANGIOGRAPHY; RECONSTRUCTION; FEASIBILITY; ARTERIES;
D O I
10.1007/s00330-011-2251-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To compare image quality and radiation dose of high-pitch dual-source computed tomography (DSCT), dual energy CT (DECT) and conventional single-source spiral CT (SCT) for pulmonary CT angiography (CTA) on a 128-slice CT system. Methods Pulmonary CTA was performed with five protocols: high-pitch DSCT (100 kV), high-pitch DSCT (120 kV), DECT (100/140 kV), SCT (100 kV), and SCT (120 kV). For each protocol, 30 sex, age, and body-mass-index (mean 25.3 kg/m(2)) matched patients were identified. Retrospectively, two observers subjectively assessed image quality, measured CT attenuation (HU +/- SD) at seven central and peripheral levels, and calculated signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR). Radiation exposure parameters (CTDIvol and DLP) were compared. Results Subjective image quality was rated good to excellent in > 92% (> 138/150) with an interobserver agreement of 91.4%. The five protocols did not significantly differ in image quality, neither by subjective, nor by objective measures (SNR, CNR). By contrast, radiation exposure differed between protocols: significant lower radiation was achieved by using high-pitch DSCT at 100 kV (p < 0.01 in all). Radiation exposure of DECT was in between SCT at 100 kV and 120 kV. Conclusions SCT, high-pitch DSCT, and DECT protocols techniques result in similar subjective and objective image quality, but radiation exposure was significantly lower with high-pitch DSCT at 100 kV. Key Points New CT protocols show promising results in pulmonary embolism assessment. High-pitch dual-source CT (DSCT) at 100 kV provides radiation dose savings for pulmonary CTA. High-pitch DSCT at 100 kV maintains diagnostic image quality for pulmonary CTA. Dual energy CT uses more radiation but also provides lung perfusion evaluation. Whether the additional perfusion data is worth the extra radiation remains undetermined.
引用
收藏
页码:279 / 286
页数:8
相关论文
共 50 条
  • [1] Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT
    Tobias De Zordo
    Klemens von Lutterotti
    Christian Dejaco
    Peter F. Soegner
    Renate Frank
    Friedrich Aigner
    Andrea S. Klauser
    Christoph Pechlaner
    U. Joseph Schoepf
    Werner R. Jaschke
    Gudrun M. Feuchtner
    European Radiology, 2012, 22 : 279 - 286
  • [2] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Pei Nie
    Ximing Wang
    Zhaoping Cheng
    Xiaopeng Ji
    Yanhua Duan
    Jiuhong Chen
    European Radiology, 2012, 22 : 2057 - 2066
  • [3] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Nie, Pei
    Wang, Ximing
    Cheng, Zhaoping
    Ji, Xiaopeng
    Duan, Yanhua
    Chen, Jiuhong
    EUROPEAN RADIOLOGY, 2012, 22 (10) : 2057 - 2066
  • [4] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents
    Yang, X.
    Chen, C. Y. D.
    Yang, Y. J. J.
    He, H. B.
    Jing, Wang
    EUROPEAN HEART JOURNAL, 2013, 34 : 844 - 844
  • [5] 128-slice dual-source CT coronary angiography with prospectively electrocardiography-triggered high-pitch spiral mode: radiation dose, image quality, and diagnostic acceptability
    Matsubara, Kosuke
    Sakuda, Keita
    Nunome, Haruka
    Takata, Tadanori
    Koshida, Kichiro
    Gabata, Toshifumi
    ACTA RADIOLOGICA, 2016, 57 (01) : 25 - 32
  • [6] High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model
    Donati, Olivio F.
    Burg, Matthias C.
    Desbiolles, Lotus
    Karlo, Christoph
    Stolzmann, Paul
    Bunck, Alexander
    Baumueller, Stephan
    Marincek, Borut
    Seifarth, Harald
    Alkadhi, Hatem
    Leschka, Sebastian
    Maintz, David
    ACADEMIC RADIOLOGY, 2010, 17 (11) : 1366 - 1374
  • [7] Diagnostic performance and radiation dose of lower extremity CT angiography using a 128-slice dual source CT at 80 kVp and high pitch
    Kim, Jin Woo
    Choo, Ki Seok
    Jeon, Ung Bae
    Kim, Tae Un
    Hwang, Jae Yeon
    Yeom, Jeong A.
    Jeong, Hee Seok
    Choi, Yoon Young
    Nam, Kyung Jin
    Kim, Chang Won
    Jeong, Dong Wook
    Lim, Soo Jin
    ACTA RADIOLOGICA, 2016, 57 (07) : 822 - 828
  • [8] High-Pitch Spiral Coronary Angiography by 128-Slice Dual-Source Computed Tomography (CT) Realizes a Sensational Low Radiation Exposure Without Reduced Image Quality
    Utanohara, Yuko
    Iguchi, Nobuo
    Yamashita, Mitsumi
    Umemura, Jun
    Sumiyoshi, Tetsuya
    CIRCULATION, 2010, 122 (21)
  • [9] Image quality and radiation dose of 128-slice dual-source CT venography using low kilovoltage combined with high-pitch scanning and automatic tube current modulation
    Park, Chan Kue
    Choo, Ki Seok
    Jeon, Ung Bae
    Baik, Seung Kug
    Kim, Yong Woo
    Kim, Tae Un
    Kim, Chang Won
    Jeong, Yeon Ju
    Jeong, Dong Wook
    Lim, Soo Jin
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 : 47 - 51
  • [10] Image quality and radiation dose of 128-slice dual-source CT venography using low kilovoltage combined with high-pitch scanning and automatic tube current modulation
    Chan Kue Park
    Ki Seok Choo
    Ung Bae Jeon
    Seung Kug Baik
    Yong Woo Kim
    Tae Un Kim
    Chang Won Kim
    Yeon Ju Jeong
    Dong Wook Jeong
    Soo Jin Lim
    The International Journal of Cardiovascular Imaging, 2013, 29 : 47 - 51