Primary stability of fixation methods for periprosthetic fractures of the humerus: a biomechanical investigation

被引:2
|
作者
Ott, Nadine [1 ]
Harland, Arne [1 ]
Hackl, Michael [1 ]
Prescher, Andreas [2 ]
Leschinger, Tim [1 ]
Mu, Lars Peter [1 ]
Wegmann, Kilian [1 ]
机构
[1] Univ Hosp Cologne, Dept Orthoped & Trauma Surg, Kerpener Str 62, D-50937 Cologne, Germany
[2] Rhein Westfal TH Aachen, Inst Mol & Cellular Anat Prosektur, Aachen, Germany
关键词
Periprosthetic fracture; revision total shoulder arthroplasty; long stem; biomechanical; locking plates; strain gauges; TOTAL SHOULDER ARTHROPLASTY; BONE-MINERAL DENSITY; HOUNSFIELD UNITS; COMPLICATIONS; MANAGEMENT; REVISION; ADJACENT;
D O I
10.1016/j.jse.2020.12.027
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The incidence of periprosthetic fractures of the proximal humerus is gradually increasing, following an increase in reverse shoulder arthroplasties in recent years. Locking plate fixation and revision arthroplasty are both valuable treatment methods. However, the primary stability of fixation methods for periprosthetic fractures has not been investigated in detail. The aim of this study was to analyze and compare the primary stability of the common treatment measures. Materials and methods: Cemented reverse total shoulder arthroplasty (Delta Xtend; DePuy Synthes, Warsaw, IN, USA) was performed in 5 shoulders, and a distal, mid-diaphysis humeral fracture (Wright and Cofield type B) was induced. The implant was left in place, and 3 distinct fixation scenarios were tested: osteosynthesis using 4.5-mm locking plate fixation (subgroup A), 4.5-mm locking plate fixation with an additional 3.5-mm locking plate (subgroup B), and 4.5-mm locking plate fixation with an additional K-wire cerclage (subgroup C). The specimens were tested in a biomechanical setup simulating activities of daily living including rotation. Strain gauges (4-wire strain at 120 U; Vishay Measurements Group, Chartres, France) mounted on the 4.5-mm locking plates were used to evaluate the strain of the fixation and to give an estimate of primary stability. Results: Regarding the simulation of activities of daily living, no statistically significant differences were found in the measured strains on the locking plate between subgroups A, B, and C. A maximum measured strain of 216.85 mm/m in subgroup A resulted in bending of the locking plate (length, 134 mm) of 0.03 mm. In subgroup B (277.01 mm/m), the plate strained 0.04 mm compared with a strain measurement of 0.01 mm in subgroup C (75.93 mm/m). Conclusion: Additional K-wire cerclages or additional 3.5-mm locked plating did not increase primary stability. With a stable prosthetic implant in place, 4.5-mm locked plating is sufficient to address periprosthetic humeral shaft fractures in the present in vitro setup. Level of evidence: Basic Science Study; Biomechanics (c) 2021 Journal of Shoulder and Elbow Surgery Board of Trustees. All rights reserved.
引用
收藏
页码:2184 / 2190
页数:7
相关论文
共 50 条
  • [1] The biomechanical assessment of fixation methods in periprosthetic femur fractures
    Gulsen, Mehmet
    Karatosun, Vasfi
    Uyulgan, Bahadir
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2011, 45 (04) : 266 - 269
  • [2] Intraprosthetic screw fixation increases primary fixation stability in periprosthetic fractures of the femur-A biomechanical study
    Brand, Stephan
    Klotz, Johannes
    Hassel, Thomas
    Petri, Maximilian
    Ettinger, Max
    Bach, Friedrich-Wilhelm
    Krettek, Christian
    Goesling, Thomas
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (02) : 239 - 243
  • [3] Biomechanical Assessment of Three Osteosynthesis Constructs by Periprosthetic Humerus Fractures
    Harb, Afif
    Welke, Bastian
    Liodakis, Emmanouil
    Razaeian, Sam
    Zhang, Dafang
    Krettek, Christian
    Hurschler, Christof
    Hawi, Nael
    ADVANCES IN ORTHOPEDICS, 2020, 2020
  • [4] Treatment of Periprosthetic Humerus Fractures With Open Reduction and Internal Fixation
    Kurowicki, Jennifer
    Momoh, Enesi
    Levy, Jonathan C.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 (11) : e369 - e374
  • [5] Periprosthetic fractures of the humerus
    Mavrogenis, Andreas F.
    Angelini, Andrea
    Guerra, Enrico
    Papagelopoulos, Panayiotis J.
    Ruggieri, Pietro
    Rotini, Roberto
    Journal of Long-Term Effects of Medical Implants, 2009, 19 (04) : 305 - 311
  • [6] Intraprosthetic screw fixation increases primary fixation stability in periprosthetic fractures of the femur-A biomechanical study (vol 36, pg 239, 2014)
    Brand, Stephan
    Klotz, Johannes
    Hassei, Thomas
    Petri, Maximilian
    Ettinger, Max
    Bach, Friedrich-Wilhelm
    Krettek, Christian
    Goesling, Thomas
    MEDICAL ENGINEERING & PHYSICS, 2016, 38 (06) : 586 - 586
  • [7] Intraprosthetic fixation techniques in the treatment of periprosthetic fractures: A biomechanical study
    Brand, Stephan
    Klotz, Johannes
    Hassel, Thomas
    Petri, Maximilian
    Haasper, Carl
    Bach, Friedrich-Wilhelm
    Krettek, Christian
    Goesling, Thomas
    WORLD JOURNAL OF ORTHOPEDICS, 2012, 3 (10): : 162 - 166
  • [8] Periprosthetic supracondylar femoral fractures following knee arthroplasty: a biomechanical comparison of four methods of fixation
    Tatu J. Mäkinen
    Herman S. Dhotar
    Simcha G. Fichman
    Matthew J. Gunton
    Mitchell Woodside
    Oleg Safir
    David Backstein
    Thomas L. Willett
    Paul R. T. Kuzyk
    International Orthopaedics, 2015, 39 : 1737 - 1742
  • [9] Periprosthetic supracondylar femoral fractures following knee arthroplasty: a biomechanical comparison of four methods of fixation
    Maekinen, Tatu J.
    Dhotar, Herman S.
    Fichman, Simcha G.
    Gunton, Matthew J.
    Woodside, Mitchell
    Safir, Oleg
    Backstein, David
    Willett, Thomas L.
    Kuzyk, Paul R. T.
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (09) : 1737 - 1742
  • [10] Biomechanical study of two different fixation methods for the treatment of Neer III proximal humerus fractures
    Cheng, Bangjun
    Jiang, Xiping
    Zhang, Xiaofeng
    Guo, Shixin
    Chen, Qi
    Du, Shengchao
    Luo, Yi
    He, Yaohua
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)