Repeatability of central corneal thickness measurements of donor corneas in preservation chamber using Fourier-domain anterior segment optical coherence tomography

被引:5
|
作者
Golla, Abhinav [1 ]
Wang, Li [1 ]
Morris, Caleb [1 ]
Weikert, Mitchell P. [1 ,2 ]
Khandelwal, Sumitra S. [1 ,2 ]
机构
[1] Baylor Coll Med, Cullen Eye Inst, 6565 Fannin St NC505B, Houston, TX 77030 USA
[2] Lions Eye Bank Texas, 1977 Butler Blvd, Houston, TX 77030 USA
关键词
Central corneal thickness; Optical coherence tomography; Repeatability; Donor cornea; Preservation;
D O I
10.1007/s10561-018-9724-z
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
To evaluate the repeatability of central corneal thickness (CCT) measurements in donor corneas using optical coherence tomography (OCT, RTVueOptovue, Inc., Fremont, CA). Consecutive corneas were measured by a single observer using the RTVue. All corneas were preserved in the Transend chamber and Life4 degrees C media (Numedis, Inc., Isanti, MN/USA) and stored at 4 degrees C. The repeatability was evaluated using a pooled within-subject standard deviation (SD), coefficient of variation (CoV), and intraclass correlation coefficient (ICC). To investigate inter-observer repeatability, a second observer independently measured the CCT for each image scan. CCT was measured in 32 eyes from 18 donors. Measurements were independently repeated by a second observer. The corneas had a mean CCT of 490.99m +/- 65.95 (381-642) as measured by Observer 1. For observer 1, the SD value for the CCT was 2.94m, the CoV value was 0.597%, and the ICC value was 0.998 (95% CI 0.996, 0.999). For observer 2, the SD value was 5.91m, the CoV value was 1.21%, and the ICC value was 0.992 (95% CI 0.985, 0.996). The Kappa statistic 21.88% with a p value<0.001. The Bland-Altman plot shows that the average CCT measurements between the two observers were within 20m of each other. The CCT measurements of donor corneas in the preservation chamber using Fourier domain OCT is highly repeatable.
引用
收藏
页码:717 / 720
页数:4
相关论文
共 50 条
  • [31] Normal corneal thickness measurements in pigmented rabbits using spectral-domain anterior segment optical coherence tomography
    Wang, Xiaogang
    Wu, Qiang
    VETERINARY OPHTHALMOLOGY, 2013, 16 (02) : 130 - 134
  • [32] Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography
    Randleman, J. Bradley
    Lynn, Michael J.
    Perez-Straziota, Claudia E.
    Weissman, Heather M.
    Kim, Sang Woo
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2015, 99 (09) : 1176 - 1181
  • [33] Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography
    Muscat, S
    McKay, N
    Parks, S
    Kemp, E
    Keating, D
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 (06) : 1791 - 1795
  • [34] Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography: Repeatability, Reproducibility, and Comparison With Scheimpflug and Automated Keratometry Measurements
    Wang, Qinmei
    Hua, Yanjun
    Savini, Giacomo
    Chen, Hao
    Bao, Fangjun
    Lin, Shishi
    Lu, Weicong
    Huang, Jinhai
    CORNEA, 2015, 34 (10) : 1266 - 1271
  • [35] Corneal Epithelial Thickness Mapping by Fourier-Domain Optical Coherence Tomography in Normal and Keratoconic Eyes
    Li, Yan
    Tan, Ou
    Brass, Robert
    Weiss, Jack L.
    Huang, David
    OPHTHALMOLOGY, 2012, 119 (12) : 2425 - 2433
  • [36] Corneal Epithelial Thickness Mapping in Normal and Keratoconic Eyes With Fourier-Domain Optical Coherence Tomography
    Li, Y.
    Tan, O.
    Huang, D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [37] Anterior Chamber Angle and Trabecular Meshwork Measurements Made by Fourier-domain Optical Coherence Tomography in Healthy White Children
    Fernandez-Vigo, Jose I.
    De-Pablo-Gomez-de-Liano, Lucia
    Fernandez-Vigo, Cristina
    Arcos-Villegas, Gabriel
    Fernandez-Perez, Cristina
    Garcia-Feijoo, Julian
    Fernandez-Vigo, Jose A.
    JOURNAL OF GLAUCOMA, 2017, 26 (09) : 810 - 815
  • [38] Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus
    Temstet, Cyril
    Sandali, Otman
    Bouheraoua, Nacim
    Hamiche, Taous
    Galan, Alice
    El Sanharawi, Mohamed
    Bashi, Elena
    Laroche, Laurent
    Borderie, Vincent
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (04): : 812 - 820
  • [39] Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera
    Gonul, Saban
    Koktekir, Bengu Ekinci
    Bakbak, Berker
    Gedik, Sansal
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2014, 77 (06) : 345 - 350
  • [40] MACULAR THICKNESS MEASUREMENTS IN NORMAL EYES WITH TIME-DOMAIN AND FOURIER-DOMAIN OPTICAL COHERENCE TOMOGRAPHY
    Huang, Jingjing
    Liu, Xing
    Wu, Ziqiang
    Xiao, Hui
    Dustin, Laurie
    Sadda, Srinivas
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (07): : 980 - 987