A practical sparse-view ultra-low dose CT acquisition scheme for PET attenuation correction

被引:0
|
作者
Miao, J. [1 ]
Fan, J. [1 ]
机构
[1] GE Healthcare, Waukesha, WI 53188 USA
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For PET imaging, CT scans are often used for PET attenuation correction and can be acquired at greatly reduced CT radiation dose levels. Techniques as low tube voltage/current have been used to obtain adequate attenuation maps in medium size patients. These techniques usually employ a smooth filter before backprojcction to reduce CT image noise which can introduce bias in the conversion from HU to attenuation values. Due to the heaviness of the smooth filter and advancement of CT reconstruction algorithm, the CT dose can be further reduced while providing the same attenuation estimation. In this work, we propose an ultra-low dose CT technique for PET attenuation correction based on sparse-view acquisition. That is, instead of an acquisition of full amount of views, only a fraction of views are acquired. We tested this technique on a 256-slice GE Revolution CT scanner using a whole-body anthropomorphic phantom. An FBP reconstruction with Q.AC on 492 views (10 mA, 120kV, 0.5s) and an FBP reconstruction with standard filter on 984 views (150 mA, 120 kV, 0.5s) produced similar attenuation uniformity. We also simulated sparse-view acquisition by skipping views in an interleaved manner from the fully-acquired data. FBP reconstruction with Q.AC filter on simulated 246 views (20mA, 120 kV, 0.5s) looks similarly to the reconstruction on acquired 984 views (20 mA, 120 kV, 0.5s), showing a further potential for dose reduction compared to the full acquisition. With the proposed sparse-view method, this work can bring at least 2x more CT dose reduction to the current Ultra-Low Dose (ULD) PET/CT protocol.
引用
收藏
页数:2
相关论文
共 50 条
  • [31] Effect of the dose reduction applied low dose for PET/CT according to CT attenuation correction method
    Jeong, Seung Woo
    Kim, Hong Gyun
    Gwon, Jae Beom
    Shin, Young Man
    Kim, Yeong Heon
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [32] Effect of the dose reduction applied low dose for PET/CT according to CT attenuation correction method
    Jeong, Seung Woo
    Kim, Hong Gyun
    Gwon, Jae Beom
    Shin, Young Man
    Kim, Yeong Heon
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [33] Low-dose CT perfusion with sparse-view filtered back projection in acute ischemic stroke
    Chung, Kevin
    MEDICAL PHYSICS, 2021, 48 (08) : 4699 - 4700
  • [34] Survey of CT dose for PET/CT localization and attenuation correction
    Jallow, Ngoneh
    Nye, Jonathon
    Christian, Paul
    Sunderland, John
    Graham, Michael
    Hoffman, John
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [35] Ultra Low Dose CT Attenuation Correction Maps for Emission Computed Tomography
    Kadhem, Haval
    Rodriguez, Daniel
    Tena, Jose-Rafael
    Wells, Kevin
    Lewis, Emma
    Guy, Matthew
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2123 - 2127
  • [36] Synchrotron based sparse-view CT artifact correction with STC-UNet
    Zhang, Renmu
    Wang, Jun
    Wang, Lihua
    Deng, Biao
    Hu, Jun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2025, 1073
  • [37] Optimization and evaluation of low dose CT protocol for attenuation correction of PET data
    Mojabi, Z.
    Ghafarian, P.
    Ghadiri, H.
    Bakhshayeshkaram, M.
    Jamaati, H.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S486 - S487
  • [38] Low-dose CT Perfusion with Sparse-view Filtered Back Projection in Acute Ischemic Stroke
    Chung, Kevin J.
    Khaw, Alexander V.
    Lee, Donald H.
    Pandey, Sachin
    Mandzia, Jennifer
    Lee, Ting-Yim
    ACADEMIC RADIOLOGY, 2022, 29 (10) : 1502 - 1511
  • [39] Long axial field of view PET enables ultra-low-dose PET/CT
    Krarup, M.
    d'Este, S. Honore
    Rexhepi, L.
    Andersen, F. Littrup
    Hojgaard, L.
    Fischer, B. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S76 - S76
  • [40] Sparse-view image reconstruction in prospectively gated micro-CT for fast and low-dose imaging
    Jonghwan Min
    Gyuseong Cho
    Seungryong Cho
    Kyoungwoo Kim
    Journal of the Korean Physical Society, 2012, 60 : 1157 - 1160