A practical sparse-view ultra-low dose CT acquisition scheme for PET attenuation correction

被引:0
|
作者
Miao, J. [1 ]
Fan, J. [1 ]
机构
[1] GE Healthcare, Waukesha, WI 53188 USA
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For PET imaging, CT scans are often used for PET attenuation correction and can be acquired at greatly reduced CT radiation dose levels. Techniques as low tube voltage/current have been used to obtain adequate attenuation maps in medium size patients. These techniques usually employ a smooth filter before backprojcction to reduce CT image noise which can introduce bias in the conversion from HU to attenuation values. Due to the heaviness of the smooth filter and advancement of CT reconstruction algorithm, the CT dose can be further reduced while providing the same attenuation estimation. In this work, we propose an ultra-low dose CT technique for PET attenuation correction based on sparse-view acquisition. That is, instead of an acquisition of full amount of views, only a fraction of views are acquired. We tested this technique on a 256-slice GE Revolution CT scanner using a whole-body anthropomorphic phantom. An FBP reconstruction with Q.AC on 492 views (10 mA, 120kV, 0.5s) and an FBP reconstruction with standard filter on 984 views (150 mA, 120 kV, 0.5s) produced similar attenuation uniformity. We also simulated sparse-view acquisition by skipping views in an interleaved manner from the fully-acquired data. FBP reconstruction with Q.AC filter on simulated 246 views (20mA, 120 kV, 0.5s) looks similarly to the reconstruction on acquired 984 views (20 mA, 120 kV, 0.5s), showing a further potential for dose reduction compared to the full acquisition. With the proposed sparse-view method, this work can bring at least 2x more CT dose reduction to the current Ultra-Low Dose (ULD) PET/CT protocol.
引用
收藏
页数:2
相关论文
共 50 条
  • [11] Attenuation correction in PET/CT from ultra-low dose CT: Photon starvation and iterative CT reconstruction
    Palmer, Matthew
    Fahey, Frederic
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [12] Ultra-Low-Dose Sparse-View Quantitative CT Liver Perfusion Imaging
    Enjilela, Esmaeil
    Lee, Ting-Yim
    Hsieh, Jiang
    Murjoomdar, Amol
    Stewart, Errol
    Dekaban, Mark
    Su, Feng
    So, Aaron
    TOMOGRAPHY, 2017, 3 (04) : 175 - 179
  • [13] Ultra low-dose CT attenuation correction in PET SPM
    Wang, Shyh-Jen
    Yang, Bang-Hung
    Tsai, Chia-Jung
    Yang, Ching-Ching
    Lee, Jason J. S.
    Wu, Tung-Hsin
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3): : 397 - 400
  • [14] Ultra-low dose CT scanning for PET/CT
    Mostafapour, Samaneh
    Greuter, Marcel
    van Snick, Johannes H.
    Brouwers, Adrienne H.
    Dierckx, Rudi A. J. O.
    van Sluis, Joyce
    Lammertsma, Adriaan A.
    Tsoumpas, Charalampos
    MEDICAL PHYSICS, 2024, 51 (01) : 139 - 155
  • [15] Ultra-low dose whole-body CT for attenuation correction in a dual tracer PET/CT protocol for multiple myeloma
    Prieto, Elena
    Jose Garcia-Velloso, Maria
    Damaso Aquerreta, Jesus
    Jose Rosales, Juan
    Fernando Bastidas, Juan
    Soriano, Ignacio
    Irazola, Leticia
    Rodriguez-Otero, Paula
    Quincoces, Gemma
    Maria Marti-Climent, Josep
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 1 - 9
  • [16] The implementation of (ultra-low dose) CT protocols in PET/CT
    Wojtylak, Patrick
    Sher, Andrew
    Vercher-Conejero, Jose
    Kohan, Andres
    O'Donnell, James
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [17] Iterative Image Reconstruction for Ultra-low-dose CT with a Combined Low-mAs and Sparse-view Protocol
    Zhang, Yunwan
    Huang, Jing
    Ma, Jianhua
    Zhang, Hua
    Bian, Zhaoying
    Zeng, Dong
    Gao, Yang
    Chen, Wufan
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 5107 - 5110
  • [18] Sparse-View Reconstruction from Restored Low-Dose CT Projections
    Liu, Wenlei
    Rong, Junyan
    Gao, Peng
    Liao, Qimei
    Jiao, Chun
    Lu, Hongbing
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [19] Beam Hardening Correction for Sparse-View CT Reconstruction
    Liu, Wenlei
    Rong, Junyan
    Gao, Peng
    Liao, Qimei
    Lu, HongBing
    MEDICAL IMAGING 2015: IMAGE PROCESSING, 2015, 9413
  • [20] Ultra-Low Dose Computed Tomography Attenuation Correction for Pediatric PET CT Using Adaptive Statistical Iterative Reconstruction (ASiR™)
    Brady, S.
    Shulkin, B.
    MEDICAL PHYSICS, 2015, 42 (06) : 3262 - 3262