Long-term performance of screw-in atrial pacing leads: A Randomized comparison of J-shaped and straight leads

被引:12
|
作者
Luria, D [1 ]
Bar-Lev, D
Gurevitz, S
Granit, H
Rotstein, Z
Eldar, M
Glikson, M
机构
[1] Chaim Sheba Med Ctr, Inst Heart, Cardiac Pacing & Electrophysiol Unit, IL-52621 Tel Hashomer, Israel
[2] Tel Aviv Univ, Sackler Fac Med, IL-69978 Tel Aviv, Israel
来源
关键词
atrial pacing; J-shaped; straight shaped leads;
D O I
10.1111/j.1540-8159.2005.00204.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Long-Term Performance of Screw-In Atrial Pacing Leads: A Randomized Comparison of J-Shaped and Straight Leads.Objective: In this prospective, randomized, control study we compared performance of J-shaped (JL) and straight atrial leads (SL) with active (screw-in)fixation mechanism, over a 5-year follow-up period. Subjects and Methods: A total of 208 consecutive patients were randomized for implantation with a Medtronic 4568 JL model (n = 103) versus a Medtronic 4068 SL model (n = 105), and all lead-related measurements and complications were recorded. Results: Lead-related complications occurred in JL and SL with similar frequency (6% and 9%, respectively, P = 0.45). Lead dislodgment occurred in 8 (7.8%) cases in SL versus none in JL (P = 0.004). Lead malfunction and excessive pacing thresholds without macrodislodgment occurred more frequently in JL (11; 10.7%) than in SL (4; 3.8%), P = 0.055. Lead material failure (insulation break) occurred only once in SL at the 5-year checkup. Most complications occurred during the first year after implantation. Other events occurred in both leads with similar frequency and included: death in 39 and 35, loss of follow-up in 8 and 6, and development of atrial fibrillation in 9 and 15 patients, respectively. Of those who completed follow-up, lead impedance and pacing threshold increased significantly in both groups. P wave decreased in SL but not in JL. Conclusion: More dislocations occurred in SL, but were counterbalanced by more lead malfunctions and excessive pacing thresholds in the JL group. Both leads demonstrated minor deterioration in electrical measurements during follow-up, with only 1 case of lead material failure in SL. Overall, both groups demonstrated favorable lead performance throughout follow-up.
引用
收藏
页码:898 / 902
页数:5
相关论文
共 50 条
  • [21] Safety and efficacy of regular screw-in leads implanted within the coronary sinus as a mean of chronic left atrial pacing
    Anselme, F
    Hidden, F
    Cribier, A
    Frank, R
    Saoudi, N
    EUROPEAN HEART JOURNAL, 2000, 21 : 193 - 193
  • [22] LONG-TERM FUNCTIONAL INTEGRITY OF ATRIAL LEADS
    PARSONNET, V
    HESSELSON, AB
    HARARI, DC
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (04): : 517 - 521
  • [23] Long-term performance of active-fixation pacing leads: A prospective study
    Kistler, PM
    Liew, G
    Mond, HG
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (03): : 226 - 230
  • [24] A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads
    Lie Liu
    Jiaojiao Tang
    Hu Peng
    Shulin Wu
    Chunying Lin
    Dongli Chen
    Qianhuan Zhang
    Yuanhong Liang
    Silin Chen
    Yan Chen
    Huiqiang Wei
    Scientific Reports, 5
  • [25] A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads
    Liu, Lie
    Tang, Jiaojiao
    Peng, Hu
    Wu, Shulin
    Lin, Chunying
    Chen, Dongli
    Zhang, Qianhuan
    Liang, Yuanhong
    Chen, Silin
    Chen, Yan
    Wei, Huiqiang
    SCIENTIFIC REPORTS, 2015, 5
  • [26] Long-term experience with bipolar mixed and polyurethane pacing leads
    Mantovan, R
    Verlato, R
    Baccillieri, S
    Turrini, P
    DeConti, F
    Piovesana, P
    EUROPACE '97 - THE OFFICIAL MEETING OF THE WORKING GROUPS ON CARDIAC PACING AND ARRHYTHMIAS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1997, : 583 - 587
  • [27] LONG-TERM PERFORMANCE OF POLYURETHANE PACING LEADS - MECHANISMS OF DESIGN-RELATED FAILURES
    PHILLIPS, R
    FREY, M
    MARTIN, RO
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06): : 1166 - 1172
  • [28] Comparison between latest generation steroid-eluting screw-in and tined leads. Long term follow-up
    Menozzi, C
    PACEMAKER LEADS 1997, 1997, : 389 - 393
  • [29] Rapid Improvement and Long-term Stability of Pacing Threshold with Active-fixation Screw-in Lead
    Sekita, Gaku
    Nakazato, Yuji
    Hayashi, Hidemori
    Hirano, Keiko
    Sugihara, Masami
    Yamase, Miki
    Komatsu, Kaoru
    Suzuki, Takeshi
    Kawano, Yasunobu
    Tokano, Takashi
    Sumiyoshi, Masataka
    Daida, Hiroyuki
    JOURNAL OF ARRHYTHMIA, 2010, 26 (04) : 244 - 249
  • [30] LONG-TERM FOLLOW-UP OF VENTRICULAR ENDOCARDIAL PACING LEADS - COMPLICATIONS, ELECTRICAL PERFORMANCE, AND LONGEVITY OF 561 RIGHT VENTRICULAR LEADS
    KAZAMA, S
    NISHIYAMA, K
    MACHII, M
    TANAKA, K
    AMANO, T
    NOMURA, T
    OHUCHI, M
    KASAHARA, S
    NIE, M
    ISHIHARA, A
    JAPANESE HEART JOURNAL, 1993, 34 (02): : 193 - 200