Adaptive noise reduction toward low-dose computed tomography

被引:17
|
作者
Lu, HB [1 ]
Li, X [1 ]
Li, LH [1 ]
Chen, DQ [1 ]
Xing, YX [1 ]
Hsieh, J [1 ]
Liang, ZR [1 ]
机构
[1] SUNY Stony Brook, Dept Radiol, Stony Brook, NY 11794 USA
关键词
noise reduction; nonstationary noise; dose reduction; computed tomography; adaptive filtering; edge-preserving smoothing; streak artifacts;
D O I
10.1117/12.480374
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An efficient noise treatment, scheme has been developed to achieve low-dose CT diagnosis based on currently available CT hardware and image reconstruction technologies. The scheme proposed includes two main parts: filtering in sinogram domain and smoothing in image domain. The acquired projection sinograms were first treated by our previously proposed Karhunen-Loeve (K-L) domain penalized weighted least-square (PWLS) filtering, which fully utilizes the prior statistical noise property and three-dimensional (3D) spatial information for an accurate restoration of the low-dose projections. To treat the streak artifacts due to photon starvation, we also incorporated an adaptive filtering into our PWLS framework, which selectively smoothes those channels contributing most to the streak artifacts. After the sinogram filtering, the image was reconstructed by the conventional filtered backprojection (FBP) method. The image is assumed as piecewise regions each has a unique texture. Therefore, an edge-preserving. smoothing (EPS) with locally-adaptive parameters to the noise variation was applied for further noise reduction in image domain. Experimental phantom projections acquired by A GE spiral computed tomography (CT) scanner under 10 mAs tube current were used to evaluate the proposed smoothing scheme. The reconstructed imaged demonstrated that the smoothing scheme with appropriate control parameters provides a significant improvement on noise suppression without sacrificing the spatial resolution.
引用
收藏
页码:759 / 766
页数:8
相关论文
共 50 条
  • [21] Rotational Augmented Noise2Inverse for Low-Dose Computed Tomography Reconstruction
    Xu, Hang
    Perelli, Alessandro
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2024, 8 (02) : 208 - 221
  • [22] Radiation Dose Reduction in Cervical Spine Computed Tomography in Cervical Spine Trauma: Comparison of Low-dose with Standard-dose Multidetector Computed Tomography
    Man, K. Y.
    Mo, K. M.
    Sung, H. T.
    Cho, K. Y.
    Tam, C. W.
    HONG KONG JOURNAL OF RADIOLOGY, 2019, 22 (03): : 180 - 187
  • [23] Low-dose computed tomography via spatially adaptive Monte-Carlo reconstruction
    Lui, Dorothy
    Cameron, Andrew
    Modhafar, Amen
    Cho, Daniel S.
    Wong, Alexander
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2013, 37 (7-8) : 438 - 449
  • [24] Low-dose abdominopelvic computed tomography in patients with lymphoma: An image quality and radiation dose reduction study
    Yoon, Sungjin
    Yoo, Kwai Han
    Park, So Hyun
    Kim, Hawk
    Lee, Jae Hoon
    Park, Jinny
    Park, Seong Ho
    Kim, Hwa Jung
    PLOS ONE, 2022, 17 (08):
  • [25] Low-Dose Computed Tomographic Urography Using Adaptive Iterative Dose Reduction 3-Dimensional: Comparison With Routine-Dose Computed Tomography With Filtered Back Projection
    Juri, Hiroshi
    Matsuki, Mitsuru
    Inada, Yuki
    Tsuboyama, Takahiro
    Kumano, Seishi
    Azuma, Haruhito
    Narumi, Yoshifumi
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (03) : 426 - 431
  • [26] Screening for lung cancer with low-dose computed tomography
    Diederich, S
    Wormanns, D
    Heindel, W
    RADIOLOGE, 2001, 41 (03): : 256 - 260
  • [27] Image Fusion for Low-Dose Computed Tomography Reconstruction
    Ma, Jianhua
    Huang, Jing
    Liang, Zhengrong
    Zhang, Hua
    Fan, Yi
    Feng, Qianjin
    Chen, Wufan
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4239 - 4243
  • [28] Subsolid Pulmonary Nodules in Low-dose Computed Tomography
    Hekmat, K.
    Bruns, C. J.
    CHIRURG, 2019, 90 (Suppl 2): : S113 - S113
  • [29] VARIATIONAL DEEP LEARNING FOR LOW-DOSE COMPUTED TOMOGRAPHY
    Kobler, Erich
    Muckley, Matthew
    Chen, Baiyu
    Knoll, Florian
    Hammernik, Kerstin
    Pock, Thomas
    Sodickson, Daniel
    Otazo, Ricardo
    2018 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2018, : 6687 - 6691
  • [30] Lung Cancer Screening with Low-Dose Computed Tomography
    Chiles, Caroline
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2014, 52 (01) : 27 - +