Bent Laue analyser crystal for Rayleigh-to-Compton computed tomography

被引:1
|
作者
Schulze, C [1 ]
Kleuker, U
机构
[1] Paul Scherrer Inst, Swiss Light Source, CH-5232 Villigen, Switzerland
[2] European Synchrotron Radiat Facil, F-38043 Grenoble, France
关键词
Rayleigh-to-Compton method; computed tomography; conical bending;
D O I
10.1107/S0909049597017251
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new optical system to perform tomography based on the Rayleigh-to-Compton (RC) method with high spatial and spectral resolution is presented. The RC technique allows the effective atomic number of a sample to be measured and finds application in bone mineral densitometry in medicine. It is particularly useful for the characterization of the distribution of biological materials which do not exhibit distinctive diffraction peaks. The system is based on the separation of the elastic line from the spectrum that is scattered by the sample by means of a bent Laue analyser crystal, and the subsequent independent detection of the elastic and inelastic parts of the spectrum with two large-area scintillation counters. The high energy resolution permits operation at low momentum transfer, where the RC method has its best contrast-to-noise ratio for low-Z materials. The geometrical and spectral requirements in terms of the incident beam and the conical analyser crystal are discussed. A first-generation tomographic imaging system (pencil beam, scanned sample) as implemented at the ESRF Compton-Scattering Station ID15B is described. A high-resolution tomographic reconstruction of a bone sample is presented.
引用
收藏
页码:1085 / 1087
页数:3
相关论文
共 50 条
  • [41] A new water-cooled and doubly bent crystal monochromator for Compton scattering experiments
    Kawata, H
    Sato, M
    Higashi, Y
    Yamaoka, H
    JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 673 - 675
  • [42] Improvements on water-cooled and doubly bent crystal monochromator for Compton scattering experiments
    Kawata, H
    Sato, M
    Higashi, Y
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 : 404 - 408
  • [43] EXPERIMENTAL-OBSERVATION OF DYNAMIC FOCUSING OF X-RAYS BY A UNIFORMLY BENT CRYSTAL IN THE LAUE CASE
    KUSHNIR, VI
    SUVOROV, EV
    JETP LETTERS, 1980, 32 (09) : 534 - 536
  • [44] INVESTIGATION OF DYNAMIC DIFFRACTION FOCUSING OF X-RAYS BY A HOMOGENEOUSLY BENT CRYSTAL IN THE SYMMETRIC LAUE CASE
    KUSHNIR, VI
    SUVOROV, EV
    PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 1982, 69 (02): : 483 - 490
  • [45] CLINICAL APPLICATION OF COMPTON AND PHOTOELECTRIC RECONSTRUCTION IN COMPUTED TOMOGRAPHY - PRELIMINARY-RESULTS
    AVRIN, DE
    MACOVSKI, A
    ZATZ, LM
    INVESTIGATIVE RADIOLOGY, 1977, 12 (05) : 445 - 445
  • [46] X-ray Fluorescence Computed Tomography with a Compton Camera for a Clinical Application
    Vernekohl, Don
    Ahmad, Moiz
    Chinn, Garry
    Xing, Lei
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [47] CLINICAL APPLICATION OF COMPTON AND PHOTOELECTRIC RECONSTRUCTION IN COMPUTED TOMOGRAPHY - PRELIMINARY-RESULTS
    AVRIN, DE
    MACOVSKI, A
    ZATZ, LM
    INVESTIGATIVE RADIOLOGY, 1978, 13 (03) : 217 - 222
  • [48] Novel reference-free methods for the determination of the instrumental response of Laue-type bent crystal spectrometers
    Szlachetko, Monika
    Hoszowska, Joanna
    Szlachetko, Jakub
    Dousse, Jean-Claude
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (11) : 2325 - 2332
  • [49] Rayleigh to Compton ratio scatter tomography applied to breast cancer diagnosis: A preliminary computational study
    Antoniassi, M.
    Conceicao, A. L. C.
    Poletti, M. E.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 288 - 291
  • [50] SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY - CORRECTION FOR COMPTON SCATTERED PHOTONS
    LOWRY, CA
    TAYLOR, DN
    HOLT, RS
    COOPER, MJ
    MCINTOSH, JAR
    PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (02): : 284 - 284