Radial Passive Stability Analysis of Magnetically Levitated Impeller Position for the Development of Flow Rate Estimations of a Ventricular Assist Device

被引:1
|
作者
Shida, Shuya [1 ,3 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
机构
[1] Toyo Univ, Fac Sci & Engn, Kawagoe, Saitama, Japan
[2] Ibaraki Univ, Grad Sch Sci & Engn, Hitachi, Ibaraki, Japan
[3] 2100 Kujirai, Kawagoe, Saitama 3508585, Japan
关键词
ventricular assist device; magnetic suspension; flow rate estimation; centrifugal blood pump; computational fluid dynamics; CENTRIFUGAL BLOOD PUMP;
D O I
10.14326/abe.11.194
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable ventricular assist devices (iVADs) are used for severe heart failure therapy. A flow rate estimation method is required for adequate control of the blood pump flow rate without using a flow meter in iVAD therapy. We developed a flow rate estimation method using the eccentric position of a magnetically levitated (maglev) impeller determined by radial passive stability. This estimation method meets clinical re-quirements because of high estimation accuracy even under varying blood viscosity (pump Reynolds number, Rep). In this study, a computational fluid dynamics (CFD) analysis was performed to clarify the basis of the maglev impeller passive stability, aiming at clinical application of this method. First, no significant variation in the pump pressure distribution, which determined the impeller's passively stabilized position, was observed un-der typical blood pump operating conditions (Rep = 35000-63000). In contrast, the working fluid density change affected the estimation accuracy by changing the pump characteristics. However, its effect on the esti-mation accuracy was less than 1% because the blood density can only change within a narrow range. These re-sults indicate that the estimation method can be applied to other devices with a passively stabilized impeller. Additionally, CFD analysis indicates that the variation in the pump flow path design based on radial clearance can change the impeller radial position, the radial hydraulic force exerted on the impeller, and its magnitude with respect to the pump flow rate. The CFD results suggest that the resolution of the method can be adjusted by changing the flow path design of the radial clearance. However, the radial hydraulic fluid direction was al-most constant in the radial clearance range of 1.0-3.0 mm. Therefore, the variation in the impeller radial posi-tion direction with respect to the pump flow rate did not change significantly with the radial clearance design under the present analysis conditions. However, the rate of variation in the impeller radial position with respect to the pump flow rate, which determines the accuracy of the estimation method, can be adjusted with the radi-al clearance design. The estimation method, which has remarkable characteristics, evaluated in this research has the potential for use in clinical practice.
引用
收藏
页码:194 / 202
页数:9
相关论文
共 29 条
  • [1] Dynamic motion analysis of impeller for the development of real-time flow rate estimations of a ventricular assist device
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (01): : 52 - 59
  • [2] Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    Sato, Ryota
    ADVANCED BIOMEDICAL ENGINEERING, 2018, 7 : 63 - 71
  • [3] Sound analysis of the magnetically levitated left ventricular assist device HeartMate 3™
    Sundbom, Per
    Roth, Michael
    Granfeldt, Hans
    Karlsson, Daniel M.
    Ahn, Henrik
    Gustafsson, Fredrik
    Dellgren, Goran
    Hubbert, Laila
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (12): : 717 - 724
  • [4] Fully Magnetically Levitated Continuous Flow Left Ventricular Assist Device: Are We There Yet?
    Frahm, T.
    Kfoury, A. G.
    Caine, W. T.
    Ehlert, R.
    Hebl, V.
    Reid, B.
    House, K.
    Butschek, R. A.
    Mohebali, D.
    Alharethi, R.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2023, 42 (04): : S340 - S340
  • [5] Effects of gravity on flow rate estimations of a centrifugal blood pump using the eccentric position of a levitated impeller
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12): : 774 - 781
  • [6] Design and Implementation of a Novel Passive Magnetically Levitated Nutation Blood Pump for the Ventricular-Assist Device
    Chen, Gang
    Yao, Ligang
    Zheng, Rongye
    Zhang, Long
    Ding, Jiaxin
    IEEE ACCESS, 2019, 7 : 169327 - 169337
  • [7] Hemodynamic effects of a fully magnetically levitated centrifugal-flow left ventricular assist device
    Estep, Jerry D.
    Ventura, Hector O.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2017, 36 (01): : 19 - 21
  • [8] Effects of Eccentric Impeller Position on Radial Passive Stability in a Magnetically Levitated Centrifugal Blood Pump with a Double Volute (vol 7, pg 63, 2018)
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    Sato, Ryota
    ADVANCED BIOMEDICAL ENGINEERING, 2018, 7 : 117 - 117
  • [9] Design and analysis of a magnetically levitated axial flux BLDG motor for a ventricular assist device (VAD)
    Fenercioglu, Ahmet
    TURKISH JOURNAL OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCES, 2016, 24 (04) : 2881 - 2892
  • [10] Flow rate estimation of a centrifugal blood pump using the passively stabilized eccentric position of a magnetically levitated impeller
    Shidai, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (06): : 291 - 298