Dynamic motion analysis of impeller for the development of real-time flow rate estimations of a ventricular assist device

被引:3
|
作者
Shida, Shuya [1 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
机构
[1] Morinomiya Univ Med Sci, Fac Hlth Sci, Osaka, Japan
[2] Ibaraki Univ, Grad Sch Sci & Engn, Hitachi, Ibaraki, Japan
来源
关键词
Flow rate estimation; impeller whirling; rotor-dynamic force; ventricular assist device; centrifugal blood pump; CENTRIFUGAL BLOOD PUMP; POSITION; LOAD;
D O I
10.1177/0391398820984485
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable ventricular assist devices are used in heart failure therapy. These devices require real-time flow rate estimation for effective mechanical circulatory support. We previously developed a flow rate estimation method using the eccentric position of a magnetically levitated impeller to achieve real-time estimation. However, dynamic motion of the levitated impeller can compromise the method's performance. Therefore, in this study, we investigated the effects of dynamic motion of the levitated impeller on the time resolution and estimation accuracy of the proposed method. The magnetically levitated impeller was axially suspended and radially restricted by the passive stability in a centrifugal blood pump that we developed. The dynamic motions of impeller rotation and whirling were analyzed at various operating conditions to evaluate the reliability of estimation. The vibration response curves of the impeller revealed that the resonant rotational speed was 1300-1400 revolutions per minute (rpm). The blood pump was used as a ventricular assist device with rotational speed (over 1800 rpm) sufficiently higher than the resonant speed. The rotor-dynamic forces on the impeller (0.03-0.14 N) suppressed the whirling motion of the impeller, indicating that the dynamic motion could be stable. Although the temporal responsiveness should be determined based on the trade-offs among the estimation accuracy and time resolution, the real-time estimation capability of the proposed method was confirmed.
引用
收藏
页码:52 / 59
页数:8
相关论文
共 50 条
  • [1] Radial Passive Stability Analysis of Magnetically Levitated Impeller Position for the Development of Flow Rate Estimations of a Ventricular Assist Device
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    ADVANCED BIOMEDICAL ENGINEERING, 2022, 11 : 194 - 202
  • [2] Measuring real-time blood viscosity with a ventricular assist device
    Hijikata, Wataru
    Maruyama, Takuro
    Suzumori, Yuki
    Shinshi, Tadahiko
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2019, 233 (05) : 562 - 569
  • [3] Flow study on a newly developed impeller for a left ventricular assist device
    Cheung-Hwa Hsu
    Journal of Artificial Organs, 2003, 6 (2) : 92 - 100
  • [4] Development of a flow rate monitoring method for the wearable ventricular assist device driver
    Kentaro Ohnuma
    Akihiko Homma
    Hirohito Sumikura
    Tomonori Tsukiya
    Yoshiaki Takewa
    Toshihide Mizuno
    Hiroshi Mukaibayashi
    Koichi Kojima
    Kazuo Katano
    Yoshiyuki Taenaka
    Eisuke Tatsumi
    Journal of Artificial Organs, 2015, 18 : 106 - 113
  • [5] Development of a flow rate monitoring method for the wearable ventricular assist device driver
    Ohnuma, Kentaro
    Homma, Akihiko
    Sumikura, Hirohito
    Tsukiya, Tomonori
    Takewa, Yoshiaki
    Mizuno, Toshihide
    Mukaibayashi, Hiroshi
    Kojima, Koichi
    Katano, Kazuo
    Taenaka, Yoshiyuki
    Tatsumi, Eisuke
    JOURNAL OF ARTIFICIAL ORGANS, 2015, 18 (02) : 106 - 113
  • [6] Dynamic modeling and identification of an axial flow ventricular assist device
    Moscato, Francesco
    Danieli, Guido A.
    Schima, Heinrich
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (06): : 336 - 343
  • [7] Flow rate prediction of a percutaneous ventricular assist device using nonlinear circuit analysis
    Mushi, S
    Yu, YC
    Simaan, MA
    Zorn, NV
    2005 IEEE 31ST ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2005, : 61 - 62
  • [8] Physiologic analysis of cardiac cycle in an implantable impeller centrifugal left ventricular assist device
    Chou, NK
    Wang, SS
    Chu, SH
    Chen, YS
    Lin, YH
    Chang, CJ
    Shyu, JJ
    Jan, GJ
    ARTIFICIAL ORGANS, 2001, 25 (08) : 613 - 616
  • [9] Development and analysis of a real-time human motion tracking system
    Luck, JP
    Debrunner, C
    Hoff, W
    He, QA
    Small, DE
    SIXTH IEEE WORKSHOP ON APPLICATIONS OF COMPUTER VISION, PROCEEDINGS, 2002, : 196 - 202
  • [10] Computational fluid dynamic analysis of the flow around the pivot bearing of the centrifugal ventricular assist device
    Nishida, Masahiro
    Yamane, Takashi
    Maruyama, Osamu
    Sankai, Yoshiyuki
    Tsutsui, Tatsuo
    JSME INTERNATIONAL JOURNAL SERIES C-MECHANICAL SYSTEMS MACHINE ELEMENTS AND MANUFACTURING, 2006, 49 (03) : 837 - 851