Radial Passive Stability Analysis of Magnetically Levitated Impeller Position for the Development of Flow Rate Estimations of a Ventricular Assist Device

被引:1
|
作者
Shida, Shuya [1 ,3 ]
Masuzawa, Toru [2 ]
Osa, Masahiro [2 ]
机构
[1] Toyo Univ, Fac Sci & Engn, Kawagoe, Saitama, Japan
[2] Ibaraki Univ, Grad Sch Sci & Engn, Hitachi, Ibaraki, Japan
[3] 2100 Kujirai, Kawagoe, Saitama 3508585, Japan
关键词
ventricular assist device; magnetic suspension; flow rate estimation; centrifugal blood pump; computational fluid dynamics; CENTRIFUGAL BLOOD PUMP;
D O I
10.14326/abe.11.194
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable ventricular assist devices (iVADs) are used for severe heart failure therapy. A flow rate estimation method is required for adequate control of the blood pump flow rate without using a flow meter in iVAD therapy. We developed a flow rate estimation method using the eccentric position of a magnetically levitated (maglev) impeller determined by radial passive stability. This estimation method meets clinical re-quirements because of high estimation accuracy even under varying blood viscosity (pump Reynolds number, Rep). In this study, a computational fluid dynamics (CFD) analysis was performed to clarify the basis of the maglev impeller passive stability, aiming at clinical application of this method. First, no significant variation in the pump pressure distribution, which determined the impeller's passively stabilized position, was observed un-der typical blood pump operating conditions (Rep = 35000-63000). In contrast, the working fluid density change affected the estimation accuracy by changing the pump characteristics. However, its effect on the esti-mation accuracy was less than 1% because the blood density can only change within a narrow range. These re-sults indicate that the estimation method can be applied to other devices with a passively stabilized impeller. Additionally, CFD analysis indicates that the variation in the pump flow path design based on radial clearance can change the impeller radial position, the radial hydraulic force exerted on the impeller, and its magnitude with respect to the pump flow rate. The CFD results suggest that the resolution of the method can be adjusted by changing the flow path design of the radial clearance. However, the radial hydraulic fluid direction was al-most constant in the radial clearance range of 1.0-3.0 mm. Therefore, the variation in the impeller radial posi-tion direction with respect to the pump flow rate did not change significantly with the radial clearance design under the present analysis conditions. However, the rate of variation in the impeller radial position with respect to the pump flow rate, which determines the accuracy of the estimation method, can be adjusted with the radi-al clearance design. The estimation method, which has remarkable characteristics, evaluated in this research has the potential for use in clinical practice.
引用
收藏
页码:194 / 202
页数:9
相关论文
共 29 条
  • [21] Impact of Race on Clinical Outcomes After Implantation With a Fully Magnetically Levitated Left Ventricular Assist Device An Analysis From the MOMENTUM 3 Trial
    Sheikh, Farooq H.
    Ravichandran, Ashwin K.
    Goldstein, Daniel J.
    Agarwal, Richa
    Ransom, John
    Bansal, Aditya
    Kim, Gene
    Cleveland, Joseph C.
    Uriel, Nir
    Sheridan, Brett C.
    Chomsky, Don
    Patel, Snehal R.
    Dirckx, Nick
    Franke, Abi
    Mehra, Mandeep R.
    CIRCULATION-HEART FAILURE, 2021, 14 (10) : 1083 - 1094
  • [22] Predictors of Physical Capacity 6 Months After Implantation of a Full Magnetically Levitated Left Ventricular Assist Device: An Analysis From the ELEVATE Registry
    Gustafsson, Finn
    Mirza, Kiran K.
    Pya, Yuri
    Shaw, Steven
    Diegeler, Anno
    Netuka, Ivan
    Lavee, Jacob
    Garbade, Jens
    Morshuis, Michiel
    Heatley, Jerry
    Saeed, Diyar
    Potapov, Evgenij
    Schmitto, Jan D.
    Zimpfer, Daniel
    JOURNAL OF CARDIAC FAILURE, 2020, 26 (07) : 580 - 587
  • [23] Diagnosis and Treatment Strategies of Outflow Graft Obstruction in the Fully Magnetically Levitated Continuous-Flow centrifugal Left Ventricular Assist Device: A Multicenter Case Series
    Wert, Leonhard
    Kaufmann, Friedrich
    Solowjowa, Natalia
    Dreysse, Stephan
    Zimpfer, Daniel
    Falk, Volkmar
    Potapov, Evgenij, V
    Mulzer, Johanna
    ASAIO JOURNAL, 2021, 67 (01) : E52 - E54
  • [24] Diagnosis and Treatment Strategies of Outflow Graft Obstruction in the Fully Magnetically Levitated Continuous-Flow Centrifugal Left Ventricular Assist Device - A Multicenter Case Series
    Wert, L.
    Kaufmann, F.
    Solowjowa, N.
    Dreysse, S.
    Zimpfer, D.
    Falk, V.
    Potapov, E. V.
    Mulzer, J.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2020, 39 (04): : S21 - S21
  • [25] Atrial Fibrillation is a Predictor of Poor Physical Capacity 6 Months after Implantation of a Full Magnetically Levitated Left Ventricular Assist Device: An Analysis from ELEVATE
    Mirza, K. K.
    Gustafsson, F.
    Pya, Y.
    Shaw, S.
    Diegeler, A.
    Netuka, I.
    Lavee, J.
    Garbade, J.
    Morshuis, M.
    Heatley, J.
    Saeed, D.
    Potapov, V.
    Schmitto, J.
    Zimpfer, D.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04): : S45 - S45
  • [26] Incidence and clinical correlates of de-novo aortic regurgitation with a fully magnetically levitated left ventricular assist device: a MOMENTUM 3 trial portfolio analysis
    Uriel, Nir
    Milano, Carmelo
    Agarwal, Richa
    Lee, Sangjin
    Cleveland, Joseph
    Goldstein, Daniel
    Wang, AiJia
    Crandall, Daniel
    Mehra, Mandeep R.
    EUROPEAN JOURNAL OF HEART FAILURE, 2023, 25 (02) : 286 - 294
  • [27] Strategies in Diagnosis and Therapy of External Outflow Graft Obstruction in Patients with a Fully Magnetically Levitated Left Ventricular Assist Device: A Meta-Analysis and Systematic Review
    Huang, Anna
    von Mackensen, Johanna K. R.
    Zwaans, Vanessa I. T.
    Schuering, Carla L.
    Iske, Jasper
    Stein, Julia
    Ott, Sascha
    Heck, Roland
    Starck, Christoph T.
    Kempfert, Joerg
    Jacobs, Stephan
    Falk, Volkmar
    Potapov, Evgenij V.
    Wert, Leonhard
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (01)
  • [28] Six-month outcomes after treatment of advanced heart failure with a full magnetically levitated continuous flow left ventricular assist device: report from the ELEVATE registry
    Gustafsson, Finn
    Shaw, Steven
    Lavee, Jacob
    Saeed, Diyar
    Pya, Yuriy
    Krabatsch, Thomas
    Schmitto, Jan
    Morshuis, Michiet
    Chuang, Joyce
    Damme, Laura
    Zimpfer, Daniel
    Garbade, Jens
    EUROPEAN HEART JOURNAL, 2018, 39 (37) : 3454 - 3460
  • [29] Comparing Patients Bridged to Transplant with a Fully Magnetically Levitated Left Ventricular Assist Device Before and after the UNOS/OPTN Allocation Change: An STS-Intermacs Registry Analysis
    Phan, D. K.
    Forest, S. J.
    Skendelas, J. P.
    Cantor, R.
    Deng, L.
    Kirklin, J.
    Weiss, A. J.
    Milena, J.
    Jacobs, J. P.
    Atluri, P.
    Goldstein, D.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2023, 42 (04): : S135 - S135