THE MEAN MACHINE; ACCURATE NON-INVASIVE BLOOD PRESSURE MEASUREMENT IN THE CRITICALLY ILL PATIENT

被引:13
|
作者
Muecke, Sandy [1 ]
Bersten, Andrew [1 ]
Plummer, John [2 ]
机构
[1] Flinders Univ S Australia, Dept Crit Care Med, Adelaide, SA 5042, Australia
[2] Flinders Med Ctr, Pain Management Unit, Adelaide, SA 5042, Australia
关键词
critical illness; blood pressure monitoring; validation; non-invasive; indirect; ASCENDING AORTIC PRESSURE; ARTERY PRESSURE; OSCILLOMETRIC METHOD; MEASURING DEVICES; SYSTEM; RELIABILITY; VALIDATION; AGREEMENT; WAVES; HYPERTENSION;
D O I
10.1007/s10877-009-9195-7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective. Accurate indirect prehospital blood pressure measurement in the critically ill patient remains an important challenge to both patient management and prehospital research. Ambulatory blood pressure measuring devices have not been trialled for prehospital use in critically ill patients. Prior to prehospital validation where conditions are suboptimal, we aimed to test under favourable conditions in the Intensive Care Unit, a selection of ambulatory devices that may be suitable for use in the field. Methods. Systolic, diastolic and mean pressures of three ambulatory devices were compared to the average of 1 min of independently recorded, high fidelity intra-arterial reference pressures. Eighteen critically ill patients were recruited. Device performance was required to fulfil the Association for the Advancement of Medical Instrumentation (AAMI) protocol. Additionally, agreement between measurement methods was examined using Bland-Altman plots. Two-level linear mixed model analyses were undertaken. Results. For each device, 150 paired measurements (arterial reference and device) were analysed. According to the AAMI protocol, no device measured systolic pressures accurately. One device measured diastolic pressures accurately. Integrated mean pressures were accurately measured by all devices. Overall, SunTech Medical's Oscar 2 performed best with mean pressure error not exceeding 17 mmHg. For this device, Bland-Altman plots showed uniform agreement across a wide range of mean pressures. Two-level linear mixed effects analyses showed that Oscar 2 mean error reduced during vasopressor use by (-) 3.9 mmHg (95% CI -5.9, -1.9; P < 0.001), and clinically, performance was little affected during mechanical ventilation. For the Oscar 2, there was up to (-) 7.0 mmHg (95% CI -10.3, -3.5; P < 0.001) l ess mean error during hypotension compared to normotension. Conclusions. In the Intensive Care Unit, the performance of one device, the Oscar 2, surpassed the others and fulfilled the AAMI protocol criteria for mean pressure measurement. This device is suitable for prehospital validation.
引用
收藏
页码:283 / 297
页数:15
相关论文
共 50 条
  • [31] Non-invasive blood pressure measurement displays - a reply
    Bailey, C. R.
    Nathanson, M. H.
    ANAESTHESIA, 2018, 73 (10) : 1300 - 1300
  • [32] ACCURATE MEASUREMENT OF BLOOD-PRESSURE IN NEONATE AND YOUNG CHILD BY A NEW NON-INVASIVE METHOD
    SHAW, A
    MAXTED, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1974, 19 (04): : 565 - 566
  • [33] STUDY OF AN AUTOMATIC NON-INVASIVE TECHNIQUE FOR THE MEASUREMENT OF SYSTOLIC, DIASTOLIC AND MEAN BLOOD-PRESSURE
    MOREL, D
    SUTER, P
    ANNALES DE L ANESTHESIOLOGIE FRANCAISE, 1981, 22 (01): : 61 - 66
  • [34] Non-invasive measurement of blood pressure and prevalence of blood pressure anomalies in dogs
    Gaur, T
    Varshney, JP
    INDIAN JOURNAL OF ANIMAL SCIENCES, 2002, 72 (01): : 68 - 69
  • [35] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Jiang, Ziqing
    Li, Shaoying
    Wang, Lin
    Yu, Feng
    Zeng, Yanping
    Li, Hongbo
    Li, Jun
    Zhang, Zhanfeng
    Zuo, Junling
    JOURNAL OF ANESTHESIA, 2024, 38 (02) : 222 - 231
  • [36] A comparison of invasive arterial blood pressure measurement with oscillometric non-invasive blood pressure measurement in patients with sepsis
    Ziqing Jiang
    Shaoying Li
    Lin Wang
    Feng Yu
    Yanping Zeng
    Hongbo Li
    Jun Li
    Zhanfeng Zhang
    Junling Zuo
    Journal of Anesthesia, 2024, 38 : 222 - 231
  • [37] Non-invasive assessment of the microcirculation in critically ill patients
    Nanas, S.
    Gerovasili, V.
    Renieris, P.
    Angelopoulos, E.
    Poriazi, M.
    Kritikos, K.
    Siafaka, A.
    Baraboutis, I.
    Zervakis, D.
    Markaki, V.
    Routsi, C.
    Roussos, C.
    ANAESTHESIA AND INTENSIVE CARE, 2009, 37 (05) : 733 - 739
  • [38] Non-invasive hemodynamic assessment in critically ill children
    Durand, P.
    Salin, J. Bailly
    Roulleau, P.
    REANIMATION, 2013, 22 (02): : 164 - 172
  • [39] The non-invasive Blood Pressure Measurement - A Measurement of a special Kind Comment
    Zweiker, D.
    JOURNAL FUR HYPERTONIE, 2021, 25 (02): : 38 - 39
  • [40] Non-Invasive Blood Pressure Sensing via Machine Learning
    Attivissimo, Filippo
    D'Alessandro, Vito Ivano
    De Palma, Luisa
    Lanzolla, Anna Maria Lucia
    Di Nisio, Attilio
    SENSORS, 2023, 23 (19)